Monitoring of arterial BP is important during anesthesia because trends in BP changes can be tracked and provide a basis for decisions on when to treat hypotensive or hypertensive states.1,2 Hypotension during anesthesia, the most common complication associated with anesthesia in dogs,3 is important to recognize because it would indicate poor circulatory function and could affect tissue perfusion.4 Tissues particularly sensitive to poor perfusion include kidney, brain, and heart.
Blood pressure monitoring has become more common with the wide availability of NIBPM techniques.5–8 Noninvasive techniques offer less invasive ways to diagnose normo-, hyper-, and hypotension in dogs and cats, and clinicians should be aware of the accuracies of the NIBPM techniques they use to allow optimal interpretation of the results (eg, the probability of hypertension or hypotension in patients with given NIBPM results).9 Guidelines from the ACVIM7 highly recommend measurement and monitoring of BP to recognize hypertension in conscious dogs and cats, as do the American College of Veterinary Anesthesia and Analgesia10 and American Animal Hospital Association11 to ensure adequacy of circulatory function.
Blood pressure monitoring techniques can be classified as noninvasive (eg, use of a Doppler ultrasonic flow detector or oscillometric measurement device) or invasive (eg, use of an arterial catheter connected to an arterial BP transducer). In general, NIBPM monitoring techniques are technically easier, more cost-effective, and involve less risk to patients than do IBPM techniques.12 Although IBPM is the gold standard, it requires more time and technical skill, has a higher cost, and poses more risk (eg, infection, embolism, and hemorrhage) to patients.1,13–17 A DBPM in dogs and cats is usually obtained at a metatarsal or metacarpal artery but can also be obtained at the base of the tail.
Studies have compared NIBPM and IBPM in dogs1,5,6,13,16–21 and cats,2,22,23 but mainly were performed under controlled, experimental conditions with healthy animals in which hypo-, normo-, and hypertensive states were induced. Similarly, most studies1,13,17,21,22 of the oscillometric method for BP measurement were performed in controlled conditions. Such studies have been helpful in advancing our understanding of trends (increasing or decreasing BP measurements in a patient) and accuracies associated with NIBPM, compared with those associated with IBPM; however, they typically involved average-sized, healthy animals with little variation in conformation that were not undergoing clinical procedures. Those studies also looked explicitly at the agreement among BP measurement methods to establish the validity and reliability of certain methods. To the authors' knowledge, no study has been reported in which epidemiological methods were used to assess the overall bias, sensitivity, and specificity of DBPM as a diagnostic tool in dogs.
The objectives of the study reported here were to determine the bias, sensitivity, and specificity of DBPM, compared with IBPM (criterion standard), as a diagnostic tool to detect hypotension in a series of dogs with various disease states and to determine whether certain patient characteristics (eg, limb conformation) could affect the accuracy of DBPM in dogs. We anticipated that DBPM would not be as useful as IBPM in detection of hypotension, but would be useful in obtaining systolic BP measurements. We also hypothesized that measurements in chondrodystrophic dog breeds with disproportionally short and curved limbs would be less accurate than those in dogs with no such limb conformation because of cuff placement difficulties.
Materials and Methods
Animals
Dogs that underwent procedures requiring general anesthesia and arterial catheterization at the University of Georgia Veterinary Teaching Hospital between April 2007 and August 2010 were eligible for the study. Dogs were included in the study whether they underwent emergency or elective procedures. Dogs with data missing after review were excluded. The Clinical Research Committee at the University of Georgia approved the study and determined that client consent was not required because the data used for the study were already being collected and related procedures were already being performed as part of clinical management.
Data collection
Data collected for each dog were breed, limb conformation, sex, American Society of Anesthesiologists physical status classification,24 anesthetic protocol, surgical procedure, arterial catheter size and location, and DBPM location. Altered limb conformation was defined as disproportionally short and curved limbs (eg, such as typical in Dachshunds and Pembroke Welsh Corgis). Anesthetic premedication and induction protocols were determined by senior veterinary students assigned to the anesthesia service and were approved by the attending faculty anesthesiologist. Each protocol was customized for the individual patient in accordance with the planned procedure. Locoregional anesthesia was performed in appropriate circumstances and recorded.
BP measurements
After induction of anesthesia, an arterial catheter for IBPM was placed into a femoral, radial, dorsal pedal, or coccygeal artery. Prior to placement, the hair over the catheter site was clipped, and the skin was aseptically prepared. A 22-, 20-, or 18-gauge catheter was placed percutaneously into the artery, and a transducer was connected to the arterial catheter with a 48-inch noncompliant pressure tubing line filled with saline (0.9% NaCl) solution. The transducer was positioned as close to the level of the right atrium as possible. Afterward, a fast-flush test was performed: the transducer was zero calibrated to atmospheric pressure, the catheter was flushed with 1 to 2 mL of heparinized saline solution, and the arterial pressure waveform was observed for 1 complete oscillation. Blood pressure measurements (by IBPM [SAP, diastolic arterial BP, and MAP] and NIBPM [SAP by DBPM]) were obtained every 5 minutes along with heart rate, respiratory rate, partial pressures of carbon dioxide, and end-tidal partial pressure of carbon dioxide.
For NIBPM, a median artery or plantar artery was used. Hair at the site (palmar or plantar aspect, respectively) of the selected foot was clipped. Coupling gel was applied to the probe of the Doppler ultrasonic flow detector, then the probe was placed on the skin over the respective artery and secured in place with adhesive tape. An inflatable cuff of appropriate size (cuff width, 40% of the circumference of the limb) was placed around the limb midway between the carpus and elbow when the DPB was to be monitored from the median artery or midway between the tibiotarsal joint and stifle joint when the DPB was to be monitored from the plantar artery.8,25 The sphygmomanometer was then attached to the cuff and used to inflate the cuff until a pulse was no longer audible through the Doppler ultrasonic flow detector probe. The pressure in the cuff was then slowly released, and the pressure at which the dog's pulse became audible again was recorded as its SAP.
Statistical analysis
Agreement between IBPMs and DBPMs of SAP was determined with the Bland-Altman plot technique for repeated measures. Bias was defined as the mean difference in values between SAP obtained by IBPM and peripheral systolic pressure obtained by DBPM, and the 95% LOA were calculated as the bias ± 1.96 SD. To be included in this analysis, dogs were required to have received at least the median number of paired measurements (IBPM and DBPM) for all dogs in the study.
The sensitivity and specificity of DBPM to detect hypotension, defined as MAP < 60 mm Hg on IBPM (criterion standard) and SAP < 90 mm Hg on DBPM, was determined for each patient by 2 methods. Method 1 involved random selection with a random number generator programa of 1 time point/dog for which paired BP recordings (IBPM and DBPM) were obtained during anesthesia and used in the calculations. Method 2 involved use of mode values for IBPM and DBPM that had been recorded for each dog. An unpaired t test was used to compare sensitivity and specificity between patients grouped by the size of catheter used for IBPM (20 vs 22 gauge), breed type (those with disproportionately short limb conformation vs proportional limb length conformation), and whether IBPMs and DBPM involved ipsilateral or contralateral limbs (excluding those paired with coccygeal arterial DBPM). Significance was set at P < 0.05. Statistical analyses were performed with available software.b–d
Results
Animals
One hundred forty-seven dogs were reviewed for inclusion; however, 1 dog was excluded because some IBPMs and DBPMs had not been recorded simultaneously. Therefore, 146 dogs were included in the study and were classified (by breed as reported by the owners) as Labrador Retriever (n = 33 [22.6%]); mixed-breed dog (12 [8.2%]); German Shepherd Dog (9 [6.2%]); Dachshund (including 2 miniature Dachshund) and Golden Retriever (8 [5.5%] each); Beagle (6 [4.1%]); Boxer, Shih Tzu (including 1 miniature Shih Tzu), and Jack Russell Terrier (5 [3.4%] each); Yorkshire Terrier [4 [2.7%]); Australian Shepherd, Cocker Spaniel, Pekingese, and Staffordshire Bull Terrier (3 [2.1%] each); Bullmastiff, corgi-type dog, Doberman, English Bulldog, Great Pyrenees, Greyhound, Maltese, and Pug (2 [1.4%] each); and 23 other breeds (1 [0.7%] each). Altered limb conformation was noted in 18 of the 146 (12.3%) dogs (8 Dachshunds, 3 Pekingese, 2 English Bulldogs, 2 corgi-type dogs, 1 Cairn Terrier, 1 Bassett Hound, and 1 Shih Tzu). Fiftyone of the 146 (34.9%) dogs were neutered males, 48 (32.9%) were spayed females, 28 (19.2%) were sexually intact males, and 19 (13.0%) were sexually intact females. Mean age was 6.7 years (range, 2 months to 15.9 years).
Anesthesia
Reported physical status of dogs according to the American Society of Anesthesiologists scale (1 = healthy to 6 = brain-dead) ranged from 1 to 5. All premedications included an opioid, and inductions were performed with various drugs, including propofol, ketamine, diazepam, thiopental, or etomidate. Anesthesia was maintained with inhalant anesthetic (isoflurane, n = 127/146 [87.0%]; sevoflurane, 17 [11.6%]; or desflurane, 2 [1.4%]) delivered in oxygen. Mean ± SD duration of anesthesia was 229.2 ± 96.4 minutes (range, 52 to 560 minutes). Three dogs were euthanized during their procedure.
BP measurements
Arterial catheters were placed (dorsal pedal artery, n = 112/146 [76.7%]; coccygeal artery, 26 [17.8%]; femoral artery, 4 [2.7%]; or radial artery, 4 [2.7%]) for IBPM monitoring. The catheter sizes used were 22 gauge (n = 106/146 [72.6%]), 20 gauge (37 [25.3%]), and 18 gauge (3 [2.1%]). Location combinations for paired IBPMs and DBPMs, respectively, were dorsal pedal and median arteries of 48 contralateral and 45 ipsilateral limbs (n = 93/146 [63.7%]), coccygeal and median arteries (20 [13.7%]), dorsal pedal and plantar arteries of contralateral limbs (19 [13.0%]), coccygeal and plantar arteries (6 [4.1%]), radial and median arteries of contralateral limbs (4 [2.7%]), and femoral and median arteries of contralateral limbs (4 [2.7%]).
Statistical analysis
Median number of paired measurements per dog was 24, and 85 of the 146 (58.2%) dogs met this criterion for inclusion in the Bland-Altman analysis. Mean bias for agreement in SAP values between the 2 techniques for these 85 dogs was 2.8 mm Hg (95% LOA, −46.4 to 51.9 mm Hg; Figure 1).

Repeated-measures Bland-Altman plot of agreement of paired DBPMs and IBPMs of SAP for 85 client-owned dogs that underwent general anesthesia for emergency or elective procedures at a veterinary teaching hospital and had arterial catheters placed. The solid line indicates the mean difference (bias, 2.8 mm Hg) in SAP values obtained by DBPM minus those obtained by IBPM for ≥ 24 paired measurements/dog. The dashed lines indicate the 95% LOA (−46.4 to 51.9 mm Hg). Each circle represents a single patient.
Citation: Journal of the American Veterinary Medical Association 253, 11; 10.2460/javma.253.11.1433

Repeated-measures Bland-Altman plot of agreement of paired DBPMs and IBPMs of SAP for 85 client-owned dogs that underwent general anesthesia for emergency or elective procedures at a veterinary teaching hospital and had arterial catheters placed. The solid line indicates the mean difference (bias, 2.8 mm Hg) in SAP values obtained by DBPM minus those obtained by IBPM for ≥ 24 paired measurements/dog. The dashed lines indicate the 95% LOA (−46.4 to 51.9 mm Hg). Each circle represents a single patient.
Citation: Journal of the American Veterinary Medical Association 253, 11; 10.2460/javma.253.11.1433
Repeated-measures Bland-Altman plot of agreement of paired DBPMs and IBPMs of SAP for 85 client-owned dogs that underwent general anesthesia for emergency or elective procedures at a veterinary teaching hospital and had arterial catheters placed. The solid line indicates the mean difference (bias, 2.8 mm Hg) in SAP values obtained by DBPM minus those obtained by IBPM for ≥ 24 paired measurements/dog. The dashed lines indicate the 95% LOA (−46.4 to 51.9 mm Hg). Each circle represents a single patient.
Citation: Journal of the American Veterinary Medical Association 253, 11; 10.2460/javma.253.11.1433
Sensitivity and specificity of DBPM to detect hypotension (on the basis of data from all 146 dogs) were 69.2% and 82.2%, respectively, for method 1 and 66.7% and 86.8%, respectively, for method 2. No differences in sensitivity or specificity were found on the basis of breed type, catheter size (20 gauge vs 22 gauge), or whether paired measurements involved ipsilateral or contralateral limbs.
Discussion
Studies15,17,26–28 have compared BP values obtained by DBPM with those obtained by the criterion standard of IBPM by arterial catheterization, and most findings indicate that DBPM generally underestimates IBPMs of SAP. Two studies26,27 in dogs have shown DBPMs to closely reflect IBPMs of SAP, and DBPM reportedly has good sensitivity and specificity in rabbits.28 A more recent comparison of BP values obtained in dogs with an oscillometric device method, DBPM, and IBPM showed that oscillometric measurements had a better agreement with IBPMs than DBPMs.25
In a recent survey,29 diplomates of the American College of Veterinary Anesthesia and Analgesia and the European College of Veterinary Anesthesia and Analgesia defined hypotension as SAP < 87 mm Hg for dogs undergoing surgical procedures and MAP < 62 mm Hg for dogs undergoing diagnostic procedures. For survey respondents, the lowest patient BP that prompted treatment intervention during surgical procedures was SAP < 85 mm Hg and MAP < 61 mm Hg, and the lowest during diagnostic procedures was SAP < 84 mm Hg and MAP < 63 mm Hg. These BP values closely resembled the values chosen in the present study to define hypotension.
Mean bias of DBPM as calculated in the present study was low, indicating that, on a population basis, DBPM was an accurate method for SAP monitoring. However, the LOA were fairly wide, indicating a low precision of this bias estimate and, therefore, wide variability in agreement between the 2 techniques (or accuracy of DBPM) among individual dogs. Although DBPMs may not have accurately reflected true SAP values (per IBPM) in individual dogs and accurate BP measurement may not have been clinically important in some circumstances, such wide departures from true values could result in errors in diagnosing hypotension on the basis of a certain cutoff value in clinical settings. Similar findings of wide LOA were reported by Vachon et al,25 who found that DBPM bias ± SD was −4.1 ± 24.7 mm Hg. In addition, Weiser et al30 also reported a wide range of BP values but indicated that values were not significantly different between IBPM and DBPM methods.
The low sensitivity of DBPM to detect hypotension in dogs of the present study indicated that DBPM was not reliable for screening these dogs for hypotension. However, specificity was relatively high, indicating that DBPM could be useful for confirming hypotension. These findings are clinically useful because they suggested that when DBPM is used as a screening test for dogs, cases of hypotension are likely being missed, whereas when DBPM is used for hypotensive dogs, the resulting measurements or trends in those measurements can likely be trusted.
In a previous study,6 DBPM overestimated SAP in 20 dogs and failed to meet the ACVIM validation criteria7 for measuring BP. Although the methods used in that study are not comparable to those of the present study, the finding coincided in that DBPM was unreliable.6
The BP cuffs used in veterinary medicine are adapted from use in humans, and the lack of validation for NIBPM devices in veterinary medicine may be in part because limbs in dogs and cats are more conical than in people. However, no differences were found in the present study for sensitivity or specificity of DBPM on the basis of limb conformation. A recent study31 compared BP measurements obtained from a traditionally used BP cuff (cylindrical) with measurements obtained from a specially made conical BP cuff for hound dogs. The specially made BP cuff met the ACVIM consensus guidelines for BP measurement; however, use of the conical cuff resulted in no better agreement between NIBPMs and IBPMs than use of the cylindrical cuff. Results of the present study were consistent with these findings and suggested that dog limb conformation may not have had a notable effect on the validity of NIBPM methods.
Results of the present study indicated no difference in results obtained when DBPM and IBPM were performed ipsilaterally versus contralaterally. Obtaining NIBPMs and IBPMs on the same side of the body did not affect measurements, likely because of the physical separation cranially and caudally. In addition, oscillometric BP measurements in dogs are reportedly unaffected by whether hair over the site of cuff application has been clipped.22 Given these findings, other variables (eg, cuff size) may be more important in DBPM accuracy than limb conformation, body size, or hair presence.
One limitation of the present study was that DBPMs and IBPMs for each dog were obtained mainly by fourth-year veterinary students. The students may have been less experienced in performing DBPM or in troubleshooting DBPMs or IBPMs if problems arose. For instance, in a previous study,32 a student, a technician, a third-year cardiology student, and a board-certified veterinary cardiologist were compared on their ability to obtain DBPMs. All 4 operators successfully obtained systolic BP measurements in all their attempts; however, the highest variability in measurements was observed for the least experienced operator.32 Nevertheless, that investigation was designed to mimic the different training levels of individuals monitoring anesthesia and performing DBPM in any setting. Factors influencing DBPM accuracy include cuff fit (including size and tightness) and position in that a falsely high measurement could be obtained when a cuff is positioned below the level of the heart and a falsely low measurement could be obtained when a cuff is too large, too tight, or positioned above the level of the heart.33 Factors influencing IBPM accuracy include inaccurate zero calibration of the transducer, placement of the transducer at a level different from the level of the right atrium, length of the connection tubing, and air bubbles within the line.
Another limitation to the present study was that the fast-flush test was performed by hand with a syringe versus a bag of saline solution pressurized to 300 mm Hg, as described elsewhere.34 Performance by hand could have resulted in less accurate IBPMs; however, such practice is common in clinical settings.
Different sites were used for IBPM in the present study, depending on the individual circumstances (eg, dog body size and site accessibility). The differences in vessel sizes with the various sites could have resulted in small differences in the pulse-pressure waveform, which reflected stroke volume. Therefore, a small vessel could have collapsed and contributed to subsequent errors in IBPM, whereas this was less likely with a larger vessel. A study35 that compared various anatomic sites for arterial BP measurement showed that significant BP differences existed among sites and that differences were more pronounced with SAP measurement than with MAP measurement. Measurements obtained at the dorsal pedal artery were significantly higher than those obtained at the carpus.35 Considering this and that most of the catheters for IBPM in the present study were placed in a dorsal pedal artery, our IBPMs of SAP could have been falsely high because of the site of arterial catheterization. Nonetheless, this mimicked clinical settings because many clinicians at the time placed catheters for IBPM in a dorsal pedal artery. On a related note, different arterial catheter sizes were used in the present study, and the size chosen depended mainly on the vessel size of the patient. Ideally, a large arterial catheter should be used.36 This is relative, however, because a 22-gauge catheter could be considered large for a small dog, compared with a 20-gauge catheter that could be considered appropriate for a large dog.
Results indicated that for dogs requiring general anesthesia and arterial catheter placement in the present study, DBPM was less accurate for detecting hypotension and monitoring BP than was IBPM, the criterion standard. Therefore, we believe that decisions made regarding BP management in anesthetized dogs, particularly those regarding hypotension, should be made with IBPM instead of DBPM, despite the greater technical skill required to perform IBPM.
Acknowledgments
The authors declare that there were no conflicts of interest.
Presented in abstract form at the International Veterinary Emergency and Critical Care Symposium, Washington, DC, September 2015.
ABBREVIATIONS
ACVIM | American College of Veterinary Internal Medicine |
BP | Blood pressure |
DBPM | Doppler ultrasonic flow detector measurement of blood pressure |
IBPM | Invasive blood pressure measurement |
LOA | Limits of agreement |
MAP | Mean arterial blood pressure |
NIBPM | Noninvasive blood pressure measurement |
SAP | Systolic arterial blood pressure |
Footnotes
True random number generator, Randomness and Integrity Services Ltd, Dublin, Ireland.
GraphPad Prism, version 5, GraphPad Software Inc, La Jolla, Calif.
MedCalc, version 16.8.4, MedCalc Software, Ostend, Belgium.
Excel, version 14.6.0, Microsoft Corp, Redmond, Wash.
References
1. Sawyer DC, Brown M, Striler EL, et al. Comparison of direct and indirect blood pressure measurement in anesthetized dogs. Lab Anim Sci 1991;41:134–138.
2. Caulkett NA, Cantwell SL, Houston DM. A comparison of indirect blood pressure monitoring techniques in the anesthetized cat. Vet Surg 1998;27:370–377.
3. Gaynor JS, Dunlop CI, Wagner AE, et al. Complications and mortality associated with anesthesia in dogs and cats. J Am Anim Hosp Assoc 1999;35:13–17.
4. Tibby SM, Murdoch IA. Measurement of cardiac output and tissue perfusion. Curr Opin Pediatr 2002;14:303–309.
5. Garofalo NA, Teixeira Neto FJ, Alvaides RK, et al. Agreement between direct, oscillometric and Doppler ultrasound blood pressures using three different cuff positions in anesthetized dogs. Vet Anaesth Analg 2012;39:324–334.
6. Seliŝnkar A, Zrimŝnek P, Sredenŝnek J, et al. Comparison of high definition oscillometric and Doppler ultrasound devices with invasive blood pressure in anesthetized dogs. Vet Anaesth Analg 2013;40:21–27.
7. Brown S, Atkins C, Bagley R, et al. Guidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats: American College of Veterinary Internal Medicine consensus statement. J Vet Intern Med 2007;21:542–558.
8. Geddes LA, Combs W, Denton W, et al. Indirect mean arterial pressure in the anesthetized dog. Am J Physiol 1980;238:H664–H666.
9. Stepien RL, Rapoport GS, Henik RA, et al. Comparative diagnostic test characteristics of oscillometric and Doppler ultrasonographic methods in the detection of systolic hypertension in dogs. J Vet Intern Med 2003;17:65–72.
10. American College of Veterinary Anesthesiologists. Monitoring guidelines update 2009: recommendations for monitoring anesthetized veterinary patients. Available at: www.acvaa.org. Accessed Jun 15, 2015.
11. Bednarski R, Grimm K, Harvey R, et al. AAHA anesthesia guidelines for dogs and cats. J Am Anim Hosp Assoc 2011;47:377–385.
12. Waddell LS. Direct blood pressure monitoring. Clin Tech Small Anim Pract 2000;15:111–118.
13. Gains MJ, Grodecki KM, Jacobs RM, et al. Comparison of direct and indirect blood pressure measurements in anesthetized dogs. Can J Vet Res 1995;59:238–240.
14. Wagner AE, Brodbelt DC. Arterial blood pressure monitoring in anesthetized animals. J Am Vet Med Assoc 1997;210:1279–1285.
15. Stepien RL, Rapoport GS. Clinical comparison of three methods to measure blood pressure in nonsedated dogs. J Am Vet Med Assoc 1999;215:1623–1628.
16. Haberman CE, Kang CW, Morgan JD, et al. Evaluation of oscillometric and Doppler ultrasonic methods of indirect blood pressure estimation in conscious dogs. Can J Vet Res 2006;70:211–217.
17. Bosiack AP, Mann FA, Dodman JR, et al. Comparison of ultrasonic Doppler flow monitor, oscillometric, and direct arterial blood pressure measurements in ill dogs. J Vet Emerg Crit Care (San Antonio) 2010;20:207–215.
18. Hunter JS, McGrath CJ, Thatcher CD, et al. Adaptation of human oscillometric blood pressure monitors for use in dogs. Am J Vet Res 1990;51:1439–1442.
19. Bodey AR, Young LE, Bartram DN, et al. A comparison of direct and indirect (oscillometric) measurements of arterial blood pressure in anesthetized dogs, using tail and limb cuffs. Res Vet Sci 1994;57:265–269.
20. Deflandre CJA, Hellebrekers LJ. Clinical evaluation of the Surgivet V60046, a non-invasive blood pressure monitor in anesthetized dogs. Vet Anaesth Analg 2008;35:13–21.
21. Rysnik MK, Cripps P, Iff I. A clinical comparison between a non-invasive blood pressure monitor using high definition oscillometry (Memodiagnostic MD 15/90 Pro) and invasive arterial blood pressure measurement in anesthetized dogs. Vet Anaesth Analg 2013;40:503–511.
22. Branson KR, Wagner-Mann CC, Mann FA. Evaluation of an oscillometric blood pressure monitor on anesthetized cats and the effect of cuff placement and fur on accuracy. Vet Surg 1997;26:347–353.
23. Pedersen KM, Butler MA, Ersb⊘ll AK, et al. Evaluation of an oscillometric blood pressure monitor for use in anesthetized cats. J Am Vet Med Assoc 2002;221:646–650.
24. American Society of Anesthesiologists. ASA physical status classification system. Available at: www.asahq.org/resources/clinical-information/asa-physical-status-classification-system. Accessed Mar 28, 2018.
25. Vachon C, Belanger MC, Burns PM. Evaluation of oscillometric and Doppler ultrasonic devices for blood pressure measurements in anesthetized and conscious dogs. Res Vet Sci 2014;97:111–117.
26. Garner HE, Hahn AW, Hartley JW, et al. Indirect blood pressure measurement in the dog. Lab Anim Sci 1975;25:197–202.
27. Grandy JL, Dunlop CI, Hodgson DS, et al. Evaluation of the Doppler ultrasonic method of measuring systolic arterial blood pressure in cats. Am J Vet Res 1992;53:1166–1169.
28. Harvey L, Knowles T, Murison PJ. Comparison of direct and Doppler arterial blood pressure measurements in rabbits during isoflurane anaesthesia. Vet Anaesth Analg 2012;39:174–184.
29. Ruffato M, Lorenzo N, Clark L. What is the definition of intraoperative hypotension in dogs? Results from a survey of diplomats of the ACVAA and ECVAA. Vet Anaesth Analg 2015;42:55–64.
30. Weiser MG, Spangler WL, Gribble DH. Blood pressure measurement in the dog. J Am Vet Med Assoc 1977;171:364–368.
31. Ramos SJ, da Cunha AF, Domingues M, et al. Comparison of blood pressure measurements of anesthetized dogs obtained noninvasively with a cylindrical blood pressure cuff and an anatomically modified conical blood pressure cuff. Am J Vet Res 2016;77:59–64.
32. Gouni V, Tissier R, Misbach C, et al. Influence of the observer's level of experience on systolic and diastolic arterial blood pressure measurements using Doppler ultrasonography in healthy conscious cats. J Feline Med Surg 2015;17:94–100.
33. Reuss-Lamky H. Monitoring blood pressure and end-tidal CO2 in the anesthetized patient. In: Bryant S, ed. Anesthesia for veterinary technicians. Ames, Iowa: Blackwell Publishing, 2010;110–111.
34. Kleinman B, Powell S, Kumar P, et al. The fast flush test measures the dynamic response of the entire blood pressure monitoring system. Anesthesiology 1992;77:1215–1220.
35. Acierno MJ, Domingues ME, Ramos SJ, et al. Comparison of directly measured arterial blood pressure at various anatomic locations in anesthetized dogs. Am J Vet Res 2015;76:266–271.
36. Haskins S. Monitoring anesthetized patients. In: Tranquilli WJ, Thurmon JC, Grimm KA. Lumb and Jones’ veterinary anesthesia and analgesia. 4th ed. Ames, Iowa: Blackwell Publishing, 2007;544.