Accuracy and precision of a point-of-care hemoglobinometer for measuring hemoglobin concentration and estimating packed cell volume in horses

Helene Chevalier Equine Performance Testing Clinic, Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853.

Search for other papers by Helene Chevalier in
Current site
Google Scholar
PubMed
Close
 DMV
,
Dr. Lysa P. Posner Equine Performance Testing Clinic, Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853.

Search for other papers by Dr. Lysa P. Posner in
Current site
Google Scholar
PubMed
Close
 DVM
,
John W. Ludders Equine Performance Testing Clinic, Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853.

Search for other papers by John W. Ludders in
Current site
Google Scholar
PubMed
Close
 DVM, DACVA
,
Tracy W. French Department of Population Medicine and Diagnostic Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853.

Search for other papers by Tracy W. French in
Current site
Google Scholar
PubMed
Close
 DVM, DACVP
,
Hollis N. Erb Department of Population Medicine and Diagnostic Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853.

Search for other papers by Hollis N. Erb in
Current site
Google Scholar
PubMed
Close
 DVM, PhD
, and
Robin D. Gleed Equine Performance Testing Clinic, Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853.

Search for other papers by Robin D. Gleed in
Current site
Google Scholar
PubMed
Close
 BVSc, DACVA

Abstract

Objective—To determine accuracy and precision of a point-of-care hemoglobinometer for measuring hemoglobin concentration and estimating PCV in horses.

Design—Prospective trial.

Animals—55 horses.

Procedure—Blood samples were obtained from 43 horses examined at a veterinary teaching hospital. Hemoglobin concentration was measured with the hemoglobinometer and by means of the standard cyanmethemoglobin method; PCV was measured by centrifugation. Blood samples were also obtained from 12 healthy horses, and PCV of aliquots of these samples was altered to approximately 5 to 80% by removing or adding plasma. Hemoglobin concentration and PCV were then measured.

Results—For samples from the clinic patients, hemoglobin concentrations obtained with the hemoglobinometer were less than concentrations obtained with the cyanmethemoglobin method; however, there was a linear relationship between concentrations obtained with the 2 methods. Breed, sex, body weight, and duration of sample storage did not significantly affect the difference between hemoglobin concentrations obtained with the 2 methods. There was a significant linear relationship between PCV and hemoglobinometer hemoglobin concentration (PCV = [2.83 X hemoglobin concentration] − 0.62). For samples from the healthy horses, a substantial negative bias was evident with the hemoglobinometer when hemoglobin concentration exceeded 16 g/dL.

Conclusions and Clinical Relevance—Results suggest that this hemoglobinometer is reasonably accurate and precise when used to measure hemoglobin concentration in blood samples from horses with a hemoglobin concentration < 16 g/dL. (J Am Vet Med Assoc 2003;223:78–83)

Abstract

Objective—To determine accuracy and precision of a point-of-care hemoglobinometer for measuring hemoglobin concentration and estimating PCV in horses.

Design—Prospective trial.

Animals—55 horses.

Procedure—Blood samples were obtained from 43 horses examined at a veterinary teaching hospital. Hemoglobin concentration was measured with the hemoglobinometer and by means of the standard cyanmethemoglobin method; PCV was measured by centrifugation. Blood samples were also obtained from 12 healthy horses, and PCV of aliquots of these samples was altered to approximately 5 to 80% by removing or adding plasma. Hemoglobin concentration and PCV were then measured.

Results—For samples from the clinic patients, hemoglobin concentrations obtained with the hemoglobinometer were less than concentrations obtained with the cyanmethemoglobin method; however, there was a linear relationship between concentrations obtained with the 2 methods. Breed, sex, body weight, and duration of sample storage did not significantly affect the difference between hemoglobin concentrations obtained with the 2 methods. There was a significant linear relationship between PCV and hemoglobinometer hemoglobin concentration (PCV = [2.83 X hemoglobin concentration] − 0.62). For samples from the healthy horses, a substantial negative bias was evident with the hemoglobinometer when hemoglobin concentration exceeded 16 g/dL.

Conclusions and Clinical Relevance—Results suggest that this hemoglobinometer is reasonably accurate and precise when used to measure hemoglobin concentration in blood samples from horses with a hemoglobin concentration < 16 g/dL. (J Am Vet Med Assoc 2003;223:78–83)

All Time Past Year Past 30 Days
Abstract Views 133 0 0
Full Text Views 813 726 12
PDF Downloads 140 82 4
Advertisement