Assessment of five portable blood glucose meters, a point-of-care analyzer, and color test strips for measuring blood glucose concentration in dogs

Leah A. Cohn Department of Veterinary Medicine and Surgery, College of Veterinary Medicine, University of Missouri, Columbia, MO 65211.

Search for other papers by Leah A. Cohn in
Current site
Google Scholar
PubMed
Close
 DVM, PhD, DACVIM
,
Dudley L. McCaw Department of Veterinary Medicine and Surgery, College of Veterinary Medicine, University of Missouri, Columbia, MO 65211.

Search for other papers by Dudley L. McCaw in
Current site
Google Scholar
PubMed
Close
 DVM, DACVIM
,
Deborah J. Tate Department of Veterinary Medicine and Surgery, College of Veterinary Medicine, University of Missouri, Columbia, MO 65211.

Search for other papers by Deborah J. Tate in
Current site
Google Scholar
PubMed
Close
, and
Jane C. Johnson Health Sciences Center Integrated Technology Services Biostatistics Group, University of Missouri, Columbia, MO 65211.
Present address: Kirksville College of Osteopathic Medicine, Kirksville, MO 63501.

Search for other papers by Jane C. Johnson in
Current site
Google Scholar
PubMed
Close
 MA

Abstract

Objective—To compare blood glucose concentrations obtained using a point-of-care (POC) analyzer, 5 portable blood glucose meters (PBGM), and a color reagent test strip with concentrations obtained using a reference method, and to compare glucose concentrations obtained using fresh blood samples in the PBGM with concentrations obtained using blood anticoagulated with lithium heparin.

Design—Case series.

Sample Population—110 blood samples from 34 dogs; glucose concentration of the samples ranged from 41 to 596 mg/dl.

Procedure—Logistic regression was used to compare blood glucose concentrations obtained with the various devices with reference method concentrations. Ease of use was evaluated subjectively. Percentage of times a clinical decision would have been altered if results of each of these methods had been used, rather than results of the reference method, was calculated.

Results—For 3 of the PBGM, blood glucose concentrations obtained with fresh blood were not significantly different from concentrations obtained with blood samples anticoagulated with lithium heparin. None of the devices provided results statistically equivalent to results of the reference method, but the POC analyzer was more accurate than the others. For some samples, reliance on results of the PBGM or the color test strip would have resulted in erroneous clinical decisions.

Conclusions and Clinical Relevance—Although commercially available PBGM and color test strips provided blood glucose concentrations reasonably close to those obtained with reference methods, some devices were more accurate than others. Use of results from these devices could lead to erroneous clinical decisions in some cases. ( J Am Vet Med Assoc 2000;216:198–202)

Abstract

Objective—To compare blood glucose concentrations obtained using a point-of-care (POC) analyzer, 5 portable blood glucose meters (PBGM), and a color reagent test strip with concentrations obtained using a reference method, and to compare glucose concentrations obtained using fresh blood samples in the PBGM with concentrations obtained using blood anticoagulated with lithium heparin.

Design—Case series.

Sample Population—110 blood samples from 34 dogs; glucose concentration of the samples ranged from 41 to 596 mg/dl.

Procedure—Logistic regression was used to compare blood glucose concentrations obtained with the various devices with reference method concentrations. Ease of use was evaluated subjectively. Percentage of times a clinical decision would have been altered if results of each of these methods had been used, rather than results of the reference method, was calculated.

Results—For 3 of the PBGM, blood glucose concentrations obtained with fresh blood were not significantly different from concentrations obtained with blood samples anticoagulated with lithium heparin. None of the devices provided results statistically equivalent to results of the reference method, but the POC analyzer was more accurate than the others. For some samples, reliance on results of the PBGM or the color test strip would have resulted in erroneous clinical decisions.

Conclusions and Clinical Relevance—Although commercially available PBGM and color test strips provided blood glucose concentrations reasonably close to those obtained with reference methods, some devices were more accurate than others. Use of results from these devices could lead to erroneous clinical decisions in some cases. ( J Am Vet Med Assoc 2000;216:198–202)