Effects of sterilization with hydrogen peroxide gas plasma, ethylene oxide, and steam on bioadhesive properties of nylon and polyethylene lines used for stabilization of canine stifle joints

Matthieu Gatineau Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Montreal, St-Hyacinthe, QC J2S 2M2, Canada.

Search for other papers by Matthieu Gatineau in
Current site
Google Scholar
PubMed
Close
 DMV
,
Alexander O El-Warrak Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Montreal, St-Hyacinthe, QC J2S 2M2, Canada.

Search for other papers by Alexander O El-Warrak in
Current site
Google Scholar
PubMed
Close
 DMV, PhD
,
Christian Bolliger Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Montreal, St-Hyacinthe, QC J2S 2M2, Canada.

Search for other papers by Christian Bolliger in
Current site
Google Scholar
PubMed
Close
 Dr med vet
,
Michael Mourez Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Montreal, St-Hyacinthe, QC J2S 2M2, Canada.

Search for other papers by Michael Mourez in
Current site
Google Scholar
PubMed
Close
 PhD
, and
Frederic Berthiaume Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Montreal, St-Hyacinthe, QC J2S 2M2, Canada.

Search for other papers by Frederic Berthiaume in
Current site
Google Scholar
PubMed
Close
 MSc

Abstract

Objective—To compare effects of sterilization with hydrogen peroxide gas plasma (HPGP), ethylene oxide, and steam on bioadhesive properties of nylon and polyethylene lines used for stabilization of canine stifle joints.

Sample—Samples of a 36.3-kg test nylon leader line, 57.8-kg test nylon fishing line, and 2-mm ultrahigh–molecular weight polyethylene (UHMWPE) were used.

Procedures—In this in vitro study, samples of nylon leader line, fishing line, and UHMWPE sterilized by use of HPGP, ethylene oxide, and steam or unsterilized samples were used. Bacterial adherence on unsterilized and sterilized samples was tested with Staphylococcus epidermidis and Escherichia coli. Five samples were examined for each line type and sterilization condition, and final colony counts were obtained.

Results—Bacterial adherence was significantly affected by method of sterilization for all 3 line types. For most of the samples, bacterial adherence was similar or lower when HPGP sterilization was used, compared with results for sterilization via ethylene oxide and steam, respectively. Bacterial adherence was significantly higher for UHMWPE, compared with adherence for the nylon line, regardless of the sterilization method used. Bacterial adherence was higher for nylon fishing line than for nylon leader line for S epidermidis after ethylene oxide sterilization and for E coli after HPGP and ethylene oxide sterilization.

Conclusions and Clinical Relevance—Effects of HPGP sterilization on bioadhesive properties of nylon and polyethylene lines compared favorably with those for ethylene oxide and steam sterilization. Also, nylon line may be a more suitable material than UHMWPE for suture prostheses on the basis of bacterial adherence properties.

Abstract

Objective—To compare effects of sterilization with hydrogen peroxide gas plasma (HPGP), ethylene oxide, and steam on bioadhesive properties of nylon and polyethylene lines used for stabilization of canine stifle joints.

Sample—Samples of a 36.3-kg test nylon leader line, 57.8-kg test nylon fishing line, and 2-mm ultrahigh–molecular weight polyethylene (UHMWPE) were used.

Procedures—In this in vitro study, samples of nylon leader line, fishing line, and UHMWPE sterilized by use of HPGP, ethylene oxide, and steam or unsterilized samples were used. Bacterial adherence on unsterilized and sterilized samples was tested with Staphylococcus epidermidis and Escherichia coli. Five samples were examined for each line type and sterilization condition, and final colony counts were obtained.

Results—Bacterial adherence was significantly affected by method of sterilization for all 3 line types. For most of the samples, bacterial adherence was similar or lower when HPGP sterilization was used, compared with results for sterilization via ethylene oxide and steam, respectively. Bacterial adherence was significantly higher for UHMWPE, compared with adherence for the nylon line, regardless of the sterilization method used. Bacterial adherence was higher for nylon fishing line than for nylon leader line for S epidermidis after ethylene oxide sterilization and for E coli after HPGP and ethylene oxide sterilization.

Conclusions and Clinical Relevance—Effects of HPGP sterilization on bioadhesive properties of nylon and polyethylene lines compared favorably with those for ethylene oxide and steam sterilization. Also, nylon line may be a more suitable material than UHMWPE for suture prostheses on the basis of bacterial adherence properties.

Bacterial contamination of biomaterial surfaces during surgery is the primary cause of implant-related infections.1 Results of several studies2–5 have clearly indicated that sterilization can affect surface properties of polymers and, consequently, their adhesive properties. Rupture of the cranial cruciate ligament is one of the most commonly diagnosed orthopedic conditions in dogs and leads to patient morbidity, reduced activity, and progressive degenerative joint disease.6,7 Surgical stabilization of the stifle joint is recommended, especially in large-breed dogs.8 Extra-articular stabilization with a lateral suture is a well-accepted and commonly performed technique,9 and many materials have been used for the lateral suture technique.10–17 However, to the authors knowledge, no studies have been conducted to evaluate the effect of various sterilization methods on bioadhesive properties of those materials.

Compared with steam sterilization, ethylene oxide sterilization was found to have the least detrimental effects on mechanical characteristics of lateral suture materials.9,18,19 However, studies20–22 in humans suggest that ethylene oxide could be carcinogenic and associated with breast and other cancers in exposed individuals. The use of ethylene oxide sterilization in North America has diminished because of strict regulations.23 Sterilization with HPGP is an attractive alternative to low-temperature sterilization with ethylene oxide. The HPGP sterilization is accomplished through synergism between hydrogen peroxide and low-temperature gas plasma. Toxic residues or emissions have not been reported; thus, aeration, ventilation, and other special procedures are not required.24 System by-products are oxygen and water vapor, and operators never come in contact with hazardous materials because the system involves easy-to-insert cassettes. Investigators of a recent study25 reported that HPGP sterilization was an attractive alternative to ethylene oxide sterilization with regard to mechanical properties of nylon lines used for lateral suture techniques.

The objective of the study reported here was to compare the effect of HPGP, ethylene oxide, and steam sterilization on the bacterial adherence properties of materials currently used for extra-articular stabilization of cranial cruciate ligament–deficient canine stifle joints. We hypothesized that HPGP sterilization would perform as well as ethylene oxide or steam sterilization.

Materials and Methods

Sample—Nylon leader line, nylon fishing line, and UHMWPE were selected for evaluation because of their wide application in small animal orthopedic procedures. Samples of 36.3-kg test monofilament nylon leader line,a 57.8-kg test monofilament nylon fishing line,b and 2-mm UHMWPEc were examined.

Sterilization procedures—Samples (length, 10 cm) of each line type were sterilized. The HPGP sterilization was performed as recommended by the manufacturer for 28 minutes.d Ethylene oxide sterilization was performed as recommended by the manufacturer at 55°C and 56 kPa.e Steam sterilization was performed as recommended by the manufacturer at 121°C and 110 kPa with an autoclave. After ethylene oxide sterilization, aeration was allowed for 12 to 24 hours as recommended by the manufacturer to remove any traces of toxic chemical.26 An unsterilized piece of each line type was used as a control sample. Thus, 12 groups of samples were evaluated (nylon leader line, unsterilized and sterilized with HPGP, ethylene oxide, and steam; nylon fishing line, unsterilized and sterilized with HPGP, ethylene oxide, and steam; and UHMWPE, unsterilized and sterilized with HPGP, ethylene oxide, and steam).

Evaluation of bacterial adherence—Gram-positive and -negative bacteria were used to evaluate bacterial adherence on samples. A well-characterized biofilm-producing Staphylococcus epidermidis strain (ATCC 35984)f was used as the gram-positive bacteria. Escherichia coli K12 strain C600 containing the pAgH or pTrc99a vectors described in another study27 was used as the gram-negative bacteria. The pAgH vector codes for the AIDA-I autotransporter, which allows the production of a biofilm; pTrc99a is a control empty vector.

Staphylococcus epidermidis strains were inoculated into brain-heart infusion brothg containing 0.25% glucose with a turbidity of 0.5 McFarland units. Escherichia coli strains were grown without agitation for 24 hours at 30°C in M9 medium containing 0.005% proline, leucine, and threonine and 100 μg of ampicillin/mL. One milliliter of the normalized cultures was placed in the wells of a 24-well plate that contained a section of the various nylon and polyethylene lines (each section was 1 cm in length; 5 replicates for each line type and sterilization condition). Plates then were incubated overnight at 37°C. At the end of the incubation period, each of the sections was removed from the medium and washed 3 times with 1 mL of PBS solution to remove nonadherent microorganisms.

Adhered S epidermidis or E coli biofilms were separated from the surface of the nylon and polyethylene lines by the addition of 1 mL of PBS solution (for S epidermidis) or 1 mL of PBS solution containing 0.5M NaCl and 1% Triton X-100 (for E coli), which was followed by vortexing for 10 minutes at room temperature (approx 27°C). The bacteria were diluted in PBS solution, and 100 μL of each diluted bacterial suspension was plated on Luria-Bertani agar (for S epidermidis) or Luria-Bertani agar containing 100 μg of ampicillin/mL (for E coli). After incubation for 24 hours, the number of CFUs were counted and adjusted on the basis of the dilution factor.

Statistical analysis—A 2-way ANOVA, with line type and sterilization condition as main factors and the interaction between the 2 main factors, was used to examine the effect of sterilization method on bacterial adherence for each material. A post hoc Tukey test was used to determine differences between pairs of means. A statistical programh was used for all analyses. Values of P < 0.05 were considered significant.

Results

The mean ± SD log10 number of CFUs of S epidermidis and E coli was determined for each line type and sterilization condition.

Number of CFUs of S epidermidis—The mean number of CFUs was significantly higher (P < 0.001) after sterilization of UHMWPE with HPGP, ethylene oxide, and steam, compared with the number of CFUs for unsterilized UHMWPE (Table 1). There was no significant difference in the number of CFUs among sterilization methods.

Table 1—

Mean ± SD log10 number of CFUs* of Staphylococcus epidermidis on nylon and polyethylene lines that were not sterilized or that were sterilized by use of HPGP, ethylene oxide, and steam.

Line typeUnsterilizedHPGPEthylene oxideSteam
UHMWPE5.77 ± 0.99a6.63 ± 0.09b6.63 ± 0.07b6.78 ± 0.08b
Nylon leader line5.27 ± 0.21c5.27 ± 0.16c4.75 ± 0.18d5.64 ± 0.07a
Nylon fishing line4.98 ± 0.08c,d5.07 ± 0.18c5.10 ± 0.19c5.73 ± 0.15a

Values reported represent results for 5 replicates for each line type and sterilization condition.

Values with different superscript letters differ significantly (P < 0.05).

The mean number of CFUs did not differ significantly between unsterilized nylon leader line and nylon leader line sterilized with HPGP (Table 1). However, the mean number of CFUs was significantly lower for nylon leader line sterilized with ethylene oxide (P < 0.001) or steam (P = 0.009), compared with results for unsterilized nylon leader line. Steam-sterilized nylon leader line had a significantly (P = 0.008) higher mean number of CFUs, compared with results for HPGP-sterilized nylon leader line, which had a significantly (P < 0.001) higher mean number of CFUs than the number for ethylene oxide–sterilized nylon leader line.

The mean number of CFUs did not differ significantly between unsterilized nylon fishing line and nylon fishing line sterilized with HPGP or ethylene oxide (Table 1). The mean number of CFUs was significantly (P < 0.001) higher after steam sterilization of nylon fishing line, compared with results for unsterilized nylon fishing line. Steam-sterilized nylon fishing line had a significantly (P < 0.001) higher mean number of CFUs, compared with the mean number of CFUS for HPGP- and ethylene oxide–sterilized nylon fishing line. There was no significant difference in the mean number of CFUs between HPGP-sterilized and ethylene oxide-sterilized nylon fishing line.

The mean number of CFUs for UHMWPE was significantly (P < 0.001) higher than that for both nylon lines for each of the sterilization conditions (including unsterilized). Nylon fishing line had a significantly (P = 0.021) higher mean number of CFUs, compared with results for nylon fishing line, after ethylene oxide sterilization. There were no significant differences between nylon fishing line and nylon leader line when they were unsterilized or after HPGP and steam sterilization.

Number of CFUs of E coli—The mean number of CFUs did not differ significantly between unsterilized UHMWPE and UHMWPE sterilized with HPGP or ethylene oxide (Table 2). However, UHMWPE sterilized with steam had a significantly higher mean number of CFUs, compared with the number for UHMWPE sterilized with HPGP (P < 0.001) or unsterilized UHMWPE (P = 0.004). There was no significant difference in the mean number of CFUs between UHMWPE sterilized with ethylene oxide or steam.

Table 2—

Mean ± SD log10 number of CFUs* of Escherichia coli on nylon and polyethylene lines that were not sterilized or that were sterilized by use of HPGP, ethylene oxide, and steam.

Line typeUnsterilizedHPGPEthylene oxideSteam
UHMWPE5.27 ± 0.05a,b5.13 ± 0.09a5.47 ± 0.04a,c5.56 ± 0.04c
Nylon leader line4.50 ± 0.14d4.20 ± 0.09e4.08 ± 0.11e4.74 ± 0.09d,f
Nylon fishing line4.18 ± 0.14e4.63 ± 0.17d,f4.75 ± 0.12f4.75 ± 0.08f

Values with different superscript letters differ significantly (P < 0.05).

The mean number of CFUs did not differ significantly between unsterilized nylon leader line and nylon leader line sterilized with steam (Table 2). The mean number of CFUs was significantly decreased after nylon leader line was sterilized with HPGP (P = 0.002) or ethylene oxide (P < 0.001), compared with results for unsterilized nylon leader line. Steam-sterilized nylon leader line had a significantly (P < 0.001) higher mean number of CFUs, compared with the mean number of CFUs for HPGP- and ethylene oxide–sterilized nylon leader line. There was no significant difference in the mean number of CFUs between HPGP- and ethylene oxide–sterilized nylon leader line.

The mean number of CFUs was significantly (P < 0.001) higher for nylon fishing line after sterilization with HPGP, ethylene oxide, and steam, compared with results for unsterilized nylon fishing line (Table 2). There were no significant differences in the mean number of CFUs between nylon fishing line sterilized with HPGP, ethylene oxide, or steam.

The mean number of CFUs for UHMWPE was significantly (P < 0.001) higher than that for both nylon lines for each of the sterilization conditions (including unsterilized). Unsterilized nylon leader line had a significantly (P = 0.001) higher mean number of CFUs, compared with the mean number of CFUs for unsterilized nylon fishing line. Nylon fishing line had a significantly (P < 0.001) higher mean number of CFUs, compared with results for nylon leader line, after sterilization with HPGP and ethylene oxide. There was no significant difference in the mean number of CFUs between nylon fishing line and nylon leader line after steam sterilization.

Discussion

Sterilization methods used for polymers in the medical field should be considered when developing new products because they may adversely affect or enhance the properties of surgical implants.5,10,28,29 The HPGP sterilization method was compared with ethylene oxide and steam sterilization of nylon and polyethylene lines used to stabilize canine stifle joints. Analysis of results of the present study indicated that HPGP sterilization compared favorably with ethylene oxide or steam sterilization of nylon and polyethylene lines with regard to the bacterial adherence properties of S epidermidis and E coli. Furthermore, UHMWPE had higher bacterial adherence, compared with bacterial adherence for nylon leader line and nylon fishing line.

In 1 study,10 ethylene oxide sterilization was considered generally more efficient than HPGP sterilization in preventing bacterial adherence to UHMWPE. In the present study, effects of HPGP on the bacterial adhesive properties of UHMWPE, nylon leader line, and nylon fishing line compared favorably with the effects of ethylene oxide, which would indicate that HPGP is an effective alternative method for these materials. Moreover, bacterial adherence was higher after steam sterilization for almost all samples tested with S epidermidis and E coli. On the basis of these findings, steam sterilization could represent a less suitable sterilization method than HPGP or ethylene oxide sterilization for polyethylene and nylon lines. It has been clearly determined that the presence of suture material in host tissue increases the susceptibility to infection.11 Bacterial adherence to biomaterial surfaces is an important step in the pathogenesis of prosthetic-related infection.12–14 Numerous studies have been conducted to examine bacterial adhesion and colonization on a variety of biomaterials. In general, factors such as type of organism, concentration and growth phase of the organism, and surface properties of the material will affect the amount of colonization and the formation of biofilm.15–17,30–34 The method of sterilization may influence bacterial adherence, as indicated in the study reported here, and ethylene oxide and HPGP sterilization may alter the surface and biomechanical properties of polymers, at least temporarily.10,25,35

In the present study, UHMWPE had increased adhesion of bacteria, compared with results for nylon leader line and nylon fishing line. Another variable that considerably influences bacterial adhesion is the roughness of the substrate's surface.11,14,36,37 In 1 study,25 the roughness and hydrophobicity of the surface of the UHMWPE were increased, compared with roughness and hydrophobicity of the surface of nylon leader line and nylon fishing line, regardless of the sterilization method used (HPGP, ethylene oxide, or steam). A rougher surface provides a larger area for bacterial adherence and multiplication and biofilm formation.13

Bacteria adhering to irregular surfaces are protected against mechanical forces, even during the initial phase of reversible adherence. Bacterial adhesion as well as colonization is increased on biomaterials with a rougher surface.10,13,38–42 In addition, UHMWPE is considered to be hydrophobic, which facilitates bacterial contact and leads to increased early adherence.43–45 These findings could potentially explain the differences in bacterial adhesion properties between polyethylene and nylon lines in the present study. Moreover, we found that nylon fishing line had increased bacterial adhesion, compared with that for nylon leader line, for S epidermidis after ethylene oxide sterilization and for E coli after HPGP and ethylene oxide sterilization. These findings suggested that when ethylene oxide is used for sterilization, nylon leader line may be a more suitable material than nylon fishing line on the basis of bacterial adherence properties. However, in vivo factors (ie, the presence of adhesive molecules, dissolved proteins, and type of host tissue) also influence initial bacterial adhesion.46 In the present study, we examined in vitro some specific bacterial strains that may be involved in the pathogenesis of implant-related infection in a clinical setting.

Gram-positive and gram-negative bacteria were used in the present study. Even though it was not specifically examined, S epidermidis typically adheres to nylon and polyethylene lines more than does E coli. Differences in the bacterial wall may result in variation in affinity to a surface11,13 and should be considered when the bioadhesive properties of a polymer are evaluated. The cell envelope of a gram-positive bacterium such as S epidermidis consists of a single layer of peptidoglycans. The cell envelope of a gram-negative bacterium such as E coli is a multilayered, complex structure comprising an outer lipidic membrane with lipopolysaccharides and proteins. In another study,14 bacteria that were more hydrophobic had increased adherence to the test material. Factors that may influence early bacterial adhesion to polymer surfaces in vitro include the type of culture medium, culture conditions, incubation time, and growth phase of the bacteria.33,47,48 In the present study, both strains were identically prepared to eliminate the effect of difference in growth conditions.

Although many methods for counting bacteria have been described, indirect counting methods, such as the number of CFUs on a plate, are widely used, and a CFU plate count is one of the most popular methods.10,12–17,39,42,47,49,50 This technique is sensitive and has the advantage of providing counts for only live bacteria. However, investigators should consider that there are several disadvantages of this technique (only live bacteria develop colonies that are counted, and clumps or chains of bacteria that develop into a single colony cause a gross underestimation of the true bacterial population). In addition, experimental manipulations (especially during the dilution and plating steps) could potentially increase the risk of errors.

Analysis of the results of the present study indicated that the effects of HPGP sterilization on the bacterial adhesive properties of nylon and polyethylene lines compared favorably with those for ethylene oxide or steam sterilization, which makes HPGP an attractive alternative method for sterilization. In addition, bacterial adhesion of S epidermidis and E coli was more extensive on UHMWPE than on nylon leader line or nylon fishing line. Therefore, UHMWPE may be a less suitable material than nylon line for use as prostheses on the basis of bacterial adherence properties. Studies are needed to confirm the clinical relevance of our findings.

ABBREVIATIONS

HPGP

Hydrogen peroxide gas plasma

UHMWPE

Ultrahigh–molecular weight polyethylene

a.

Mason hard-type leader material, Mason Tackle Co, Otisville, Mich.

b.

Tortue La Soie-Neyme SA, Boulogne, France.

c.

Fibertape 2 mm, Arthrex SAS, Lezennes, France.

d.

Sterrad, ASP J&J Co Medical Products, Irvine, Calif.

e.

Steri-Vac, 3M, Saint Paul, Minn.

f.

provided by Dr. Marie Archambaut, Faculty of Veterinary Medicine, University of Montreal, St-Hyacinthe, QC, Canada.

g.

Difco, BD, Sparks, Md.

h.

SAS, version 9.1, SAS Institute Inc, Cary, NC.

References

  • 1.

    Arciola CRAlvi FIAn YH, et al. Implant infection and infection resistant materials: a mini review. Int J Artif Organs 2005; 28:11191125.

  • 2.

    Nair PD. Currently practised sterilization methods–some inadvertent consequences. J Biomater Appl 1995; 10:121135.

  • 3.

    Deng MShalaby SW. Effects of γ irradiation, gas environments, and postirradiation aging on ultrahigh molecular weight polyethylene. J Appl Polymer Sci 1995; 58:21112119.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 4.

    Hirata NMatsumoto K-IInishita T, et al. γ-ray irradiation, autoclave and ethylene oxide sterilization to thermosetting polyurethane: sterilization to polyurethane. Radiat Phys Chem 1995; 46:377381.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 5.

    Costa LLuda MPTrossarelli L, et al. Oxidation in orthopaedic UHMWPE sterilized by gamma-radiation and ethylene oxide. Biomaterials 1998; 19:659668.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 6.

    Arnoczky SPMarshall JL. The cruciate ligaments of the canine stifle. An anatomical and functional analysis. Am J Vet Res 1977; 38:18071814.

    • Search Google Scholar
    • Export Citation
  • 7.

    Arnoczky SP The cruciate ligaments: the enigma of the canine stifle. J Small Anim Pract 1988; 29:7190.

  • 8.

    Vasseur PB. Clinical results following nonoperative management for rupture of the cranial cruciate ligament in dogs. Vet Surg 1984; 13:243246.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 9.

    Caporn TMRoe SC. Biomechanical evaluation of the suitability of monofilament nylon fishing and leader line for extra-articular stabilization of the canine cruciate ligament deficient stifle. Vet Comp Orthop Traumatol 1996; 9:126133.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 10.

    Kinnari TJEsteban JZamora N, et al. Effect of surface roughness and sterilization on bacterial adherence to ultra-high molecular weight polyethylene. Clin Microbiol Infect 2010; 16:10361041.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 11.

    Chu CCWilliams DF. Effects of physical configuration and chemical structure of suture materials on bacterial adhesion. Am J Surg 1984; 147:197204.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 12.

    An HYFriedman RJDraughn RA, et al. Rapid quantification of staphylococci adhered to titanium surfaces using image analyzed epifluorescence microscopy. J Microbiol Methods 1995; 24:2940.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 13.

    Ha KYChung YGRyoo SJ. Adherence and biofilm formation of Staphylococcus epidermidis and Mycobacterium tuberculosis on various spinal implants. Spine 2005; 30:3843.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 14.

    Verheyen CCPMDhert WJABlieck-Hogervorst JMA, et al. Adherence to a metal, polymer and composite by Staphylococcus aureus and Staphylococcus epidermidis. Biomaterials 1993; 14:383391.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 15.

    Sardin SMorrier JJBenay G, et al. In vitro streptococcal adherence on prosthetic and implant. Interactions with physicochemical surface properties. J Oral Rehabil 2004; 31:140148.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 16.

    O'kte ESultan NDogan B, et al. Bacterial adhesion of Actinobacillus actinomycetemcomitans serotypes to titanium implants: SEM evaluation. A preliminary report. J Periodontol 1999; 70:13761382.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 17.

    Zhao QLiu YWang C, et al. Reduction of bacterial adhesion on ion-implanted stainless steel surfaces. Med Eng Phys 2008; 30:341349.

  • 18.

    Sicard GKMeinen JPhillips T, et al. Comparison of fishing lines for repair of the cruciate deficient stifle. Vet Comp Orthop Traumatol 1999; 44:138141.

    • Search Google Scholar
    • Export Citation
  • 19.

    Sicard GKHayashi KManley PA. Evaluation of 5 types of fishing materials, 2 sterilization methods, and a crimp-clamp system for extra-articular stabilization of the canine stifle joint. Vet Surg 2002; 31:7894.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 20.

    Steenland KStayner LDeddens J. Mortality analyses in a cohort of 18 235 ethylene oxide exposed workers: follow up extended from 1987 to 1998. Occup Environ Med 2004; 61:27.

    • Search Google Scholar
    • Export Citation
  • 21.

    Steenland KWhelan EDeddens J, et al. Ethylene oxide and breast cancer incidence in a cohort study of 7576 women (United States). Cancer Causes Control 2003; 14:531539.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 22.

    Steenland KStayner LGreife A, et al. Mortality among workers exposed to ethylene oxide. N Engl J Med 1991; 324:14021407.

  • 23.

    Environment Canada. Guidelines for the reduction of ethylene oxide releases from sterilization applications. Part I. Canada Gazette 2004. Available at: canadagazette.gc.ca/partI/2004/20040403/html/notice-e.html#i2. Accessed Aug 21, 2012.

    • Search Google Scholar
    • Export Citation
  • 24.

    Mitchell SLBerg J. Sterilization. In: Slatter D, ed. Textbook of small animal surgery. 3rd ed. Philadelphia: WB Saunders Co, 2003;160161.

    • Search Google Scholar
    • Export Citation
  • 25.

    Gatineau MHuneault LLussier B, et al. Mechanical evaluation of hydrogen peroxyde gas plasma sterilization of nylon lines used for extra-articular stabilization of the canine stifle joint. Vet Surg 2010; 39:4853.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 26.

    Widmer AFFrei R. Decontamination, disinfection, and sterilization. In: Murray PRBaron EJJorgensen J, et al, eds. Manual of clinical microbiology. 9th ed. Washington, DC: ASM Press, 2007;6596.

    • Search Google Scholar
    • Export Citation
  • 27.

    Charbonneau MEBerthiaume FMourez M. Proteolytic processing is not essential for multiple functions of the Escherichia coli autotransporter adhesin involved in diffuse adherence (AIDA-I) J Bacteriol 2006; 188:85048512.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 28.

    Lerouge STabrizian MWertheimer MR, et al. Safety of plasma-based sterilization: surface modifications of polymeric medical devices induced by Sterrad and Plazlyte processes. BioMed Mat Eng 2002; 12:313.

    • Search Google Scholar
    • Export Citation
  • 29.

    Calvet JLGrafahrend DKlee D, et al. Sterilization effects on StarPEG coated polymer surfaces: characterization and cell viability. J Mater Sci Mater Med 2008; 19:16311636.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 30.

    Mcallister EWCarey LCBrady PG, et al. The role of polymeric surface smoothness of biliary stents in bacterial adherence, biofilm deposition, and stent occlusion. Gastrointest Endosc 1993; 39:422425.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 31.

    Wendel RN. Surface roughness and contact angle. J Phys Colloid Chem 1949; 53:14661467.

  • 32.

    Bayoudh SOthmane ABettaieb F, et al. Quantification of the adhesion free energy between bacteria and hydrophobic and hydrophilic substrata. Mater Sci Eng C Biomimetic Supramol Syst 2006; 26:300305.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 33.

    Bos Rvan der Mei HCBusscher HJ. Physico-chemistry of initial microbial adhesive interactions—its mechanisms and method for study. FEMS Microbiol Rev 1999; 23:179230.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 34.

    Roosjen ANorde Wvan der Mei HC, et al. Characterization of polymer surfaces and thin films. Prog Colloid Polym Sci 2006; 132:138144.

  • 35.

    Rhodes NPWilson DJWilliams RL. The effect of gas plasma modification on platelet and contact phase activation processes. Biomaterials 2007; 28:45614570.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 36.

    Quirynen MBollen CML. The influence of surface-roughness and surface-free energy on supragingival and subgingival plaque-formation in man—a review of the literature. J Clin Periodontol 1995; 22:114.

    • Search Google Scholar
    • Export Citation
  • 37.

    Katz SIzhar MMirelman D. Bacterial adherence to surgical sutures. A possible factor in suture induced infection. Ann Surg 1981; 184:3541.

    • Search Google Scholar
    • Export Citation
  • 38.

    Cordero JMunuera LFolgueira MD. Influence of metal implants on infection. J Bone Joint Surg Br 1994; 76:717720.

  • 39.

    Wu-Yuan CDEganhouse KJKeller JC, et al. Oral bacterial attachment to titanium surfaces: a scanning electron microscopy study. J Oral Implantol 1995; 21:207213.

    • Search Google Scholar
    • Export Citation
  • 40.

    Gallardo-Moreno AMGonzalez-Martin MLBruque JM, et al. The adhesion strength of Candida parapsilosis to glass and silicone as a function of hydrophobicity, roughness and cell morphology. Colloids Surf A Physicochem Eng Aspects 2004; 249:99103.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 41.

    Quirynen MVandermei HCBollen CML, et al. An in-vivo study of the influence of the surface roughness of implants on the microbiology of supragingival and subgingival plaque. J Dent Res 1993; 72:13041309.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 42.

    Tebbs SESawyer AElliott TSJ. Influence of surface-morphology on in-vitro bacterial adherence to central venous catheters. Br J Anaesth 1994; 72:587591.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 43.

    Wertheimer MRFozza ACHollander A. Industrial processing of polymers by low-pressure plasmas: the role of VUV radiation, Nuclear Instrum Methods Phys Res B 1999; 151:6575.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 44.

    MacDonald DEDeo NMarkovic B, et al. Adsorption and dissolution behavior of human plasma fibronectin on thermally and chemically modified titanium dioxide particles. Biomaterials 2002; 23:12691279.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 45.

    Hess HVogel V Molecular shuttles based on motor proteins: active transport in synthetic environments. J Biotech 2001; 82:6785.

  • 46.

    Karakecili AGGumusderelioglu M. Comparison of bacterial and tissue cell initial adhesion on hydrophilic/hydrophobic biomaterials. J Biomater Sci Polym Ed 2002; 13:185196.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 47.

    Hogt AHDankert JFeijen J. Adhesion of coagulase-negative staphylococci to methacrylate polymers and copolymers. J Biomed Mater Res 1986; 20:533545.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 48.

    Roosjen ANorde WMei HC, et al. The use of positively charged or low surface free energy coatings versus polymer brushes in controlling biofilm formation. Progr Colloid Polym Sci 2006; 132:138144.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 49.

    Li BKLogan BE. Bacterial adhesion to glass and metal-oxide surfaces. Colloids Surf B Biointerfaces 2004; 36:8190.

  • 50.

    Grivet MMorrier JJBenay G, et al. Effect of hydrophobicity on in vitro streptococcal adhesion to dental alloys. J Mater Sci Mater Med 2000; 11:637642.

    • Search Google Scholar
    • Export Citation

Contributor Notes

Dr. El-Warrack's present address is Animal Resources Center, University Health Network, 101 College St, Toronto, ON M5G 1L7, Canada.

Supported by the Fonds en Santé des Animaux de Compagnie (FSAC) de l'Université de Montréal.

The authors thank Guy Beauchamp for statistical assistance.

Address correspondence to Dr. El-Warrak (aelwarra@uhnresearch.ca).
  • 1.

    Arciola CRAlvi FIAn YH, et al. Implant infection and infection resistant materials: a mini review. Int J Artif Organs 2005; 28:11191125.

  • 2.

    Nair PD. Currently practised sterilization methods–some inadvertent consequences. J Biomater Appl 1995; 10:121135.

  • 3.

    Deng MShalaby SW. Effects of γ irradiation, gas environments, and postirradiation aging on ultrahigh molecular weight polyethylene. J Appl Polymer Sci 1995; 58:21112119.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 4.

    Hirata NMatsumoto K-IInishita T, et al. γ-ray irradiation, autoclave and ethylene oxide sterilization to thermosetting polyurethane: sterilization to polyurethane. Radiat Phys Chem 1995; 46:377381.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 5.

    Costa LLuda MPTrossarelli L, et al. Oxidation in orthopaedic UHMWPE sterilized by gamma-radiation and ethylene oxide. Biomaterials 1998; 19:659668.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 6.

    Arnoczky SPMarshall JL. The cruciate ligaments of the canine stifle. An anatomical and functional analysis. Am J Vet Res 1977; 38:18071814.

    • Search Google Scholar
    • Export Citation
  • 7.

    Arnoczky SP The cruciate ligaments: the enigma of the canine stifle. J Small Anim Pract 1988; 29:7190.

  • 8.

    Vasseur PB. Clinical results following nonoperative management for rupture of the cranial cruciate ligament in dogs. Vet Surg 1984; 13:243246.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 9.

    Caporn TMRoe SC. Biomechanical evaluation of the suitability of monofilament nylon fishing and leader line for extra-articular stabilization of the canine cruciate ligament deficient stifle. Vet Comp Orthop Traumatol 1996; 9:126133.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 10.

    Kinnari TJEsteban JZamora N, et al. Effect of surface roughness and sterilization on bacterial adherence to ultra-high molecular weight polyethylene. Clin Microbiol Infect 2010; 16:10361041.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 11.

    Chu CCWilliams DF. Effects of physical configuration and chemical structure of suture materials on bacterial adhesion. Am J Surg 1984; 147:197204.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 12.

    An HYFriedman RJDraughn RA, et al. Rapid quantification of staphylococci adhered to titanium surfaces using image analyzed epifluorescence microscopy. J Microbiol Methods 1995; 24:2940.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 13.

    Ha KYChung YGRyoo SJ. Adherence and biofilm formation of Staphylococcus epidermidis and Mycobacterium tuberculosis on various spinal implants. Spine 2005; 30:3843.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 14.

    Verheyen CCPMDhert WJABlieck-Hogervorst JMA, et al. Adherence to a metal, polymer and composite by Staphylococcus aureus and Staphylococcus epidermidis. Biomaterials 1993; 14:383391.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 15.

    Sardin SMorrier JJBenay G, et al. In vitro streptococcal adherence on prosthetic and implant. Interactions with physicochemical surface properties. J Oral Rehabil 2004; 31:140148.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 16.

    O'kte ESultan NDogan B, et al. Bacterial adhesion of Actinobacillus actinomycetemcomitans serotypes to titanium implants: SEM evaluation. A preliminary report. J Periodontol 1999; 70:13761382.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 17.

    Zhao QLiu YWang C, et al. Reduction of bacterial adhesion on ion-implanted stainless steel surfaces. Med Eng Phys 2008; 30:341349.

  • 18.

    Sicard GKMeinen JPhillips T, et al. Comparison of fishing lines for repair of the cruciate deficient stifle. Vet Comp Orthop Traumatol 1999; 44:138141.

    • Search Google Scholar
    • Export Citation
  • 19.

    Sicard GKHayashi KManley PA. Evaluation of 5 types of fishing materials, 2 sterilization methods, and a crimp-clamp system for extra-articular stabilization of the canine stifle joint. Vet Surg 2002; 31:7894.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 20.

    Steenland KStayner LDeddens J. Mortality analyses in a cohort of 18 235 ethylene oxide exposed workers: follow up extended from 1987 to 1998. Occup Environ Med 2004; 61:27.

    • Search Google Scholar
    • Export Citation
  • 21.

    Steenland KWhelan EDeddens J, et al. Ethylene oxide and breast cancer incidence in a cohort study of 7576 women (United States). Cancer Causes Control 2003; 14:531539.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 22.

    Steenland KStayner LGreife A, et al. Mortality among workers exposed to ethylene oxide. N Engl J Med 1991; 324:14021407.

  • 23.

    Environment Canada. Guidelines for the reduction of ethylene oxide releases from sterilization applications. Part I. Canada Gazette 2004. Available at: canadagazette.gc.ca/partI/2004/20040403/html/notice-e.html#i2. Accessed Aug 21, 2012.

    • Search Google Scholar
    • Export Citation
  • 24.

    Mitchell SLBerg J. Sterilization. In: Slatter D, ed. Textbook of small animal surgery. 3rd ed. Philadelphia: WB Saunders Co, 2003;160161.

    • Search Google Scholar
    • Export Citation
  • 25.

    Gatineau MHuneault LLussier B, et al. Mechanical evaluation of hydrogen peroxyde gas plasma sterilization of nylon lines used for extra-articular stabilization of the canine stifle joint. Vet Surg 2010; 39:4853.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 26.

    Widmer AFFrei R. Decontamination, disinfection, and sterilization. In: Murray PRBaron EJJorgensen J, et al, eds. Manual of clinical microbiology. 9th ed. Washington, DC: ASM Press, 2007;6596.

    • Search Google Scholar
    • Export Citation
  • 27.

    Charbonneau MEBerthiaume FMourez M. Proteolytic processing is not essential for multiple functions of the Escherichia coli autotransporter adhesin involved in diffuse adherence (AIDA-I) J Bacteriol 2006; 188:85048512.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 28.

    Lerouge STabrizian MWertheimer MR, et al. Safety of plasma-based sterilization: surface modifications of polymeric medical devices induced by Sterrad and Plazlyte processes. BioMed Mat Eng 2002; 12:313.

    • Search Google Scholar
    • Export Citation
  • 29.

    Calvet JLGrafahrend DKlee D, et al. Sterilization effects on StarPEG coated polymer surfaces: characterization and cell viability. J Mater Sci Mater Med 2008; 19:16311636.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 30.

    Mcallister EWCarey LCBrady PG, et al. The role of polymeric surface smoothness of biliary stents in bacterial adherence, biofilm deposition, and stent occlusion. Gastrointest Endosc 1993; 39:422425.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 31.

    Wendel RN. Surface roughness and contact angle. J Phys Colloid Chem 1949; 53:14661467.

  • 32.

    Bayoudh SOthmane ABettaieb F, et al. Quantification of the adhesion free energy between bacteria and hydrophobic and hydrophilic substrata. Mater Sci Eng C Biomimetic Supramol Syst 2006; 26:300305.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 33.

    Bos Rvan der Mei HCBusscher HJ. Physico-chemistry of initial microbial adhesive interactions—its mechanisms and method for study. FEMS Microbiol Rev 1999; 23:179230.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 34.

    Roosjen ANorde Wvan der Mei HC, et al. Characterization of polymer surfaces and thin films. Prog Colloid Polym Sci 2006; 132:138144.

  • 35.

    Rhodes NPWilson DJWilliams RL. The effect of gas plasma modification on platelet and contact phase activation processes. Biomaterials 2007; 28:45614570.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 36.

    Quirynen MBollen CML. The influence of surface-roughness and surface-free energy on supragingival and subgingival plaque-formation in man—a review of the literature. J Clin Periodontol 1995; 22:114.

    • Search Google Scholar
    • Export Citation
  • 37.

    Katz SIzhar MMirelman D. Bacterial adherence to surgical sutures. A possible factor in suture induced infection. Ann Surg 1981; 184:3541.

    • Search Google Scholar
    • Export Citation
  • 38.

    Cordero JMunuera LFolgueira MD. Influence of metal implants on infection. J Bone Joint Surg Br 1994; 76:717720.

  • 39.

    Wu-Yuan CDEganhouse KJKeller JC, et al. Oral bacterial attachment to titanium surfaces: a scanning electron microscopy study. J Oral Implantol 1995; 21:207213.

    • Search Google Scholar
    • Export Citation
  • 40.

    Gallardo-Moreno AMGonzalez-Martin MLBruque JM, et al. The adhesion strength of Candida parapsilosis to glass and silicone as a function of hydrophobicity, roughness and cell morphology. Colloids Surf A Physicochem Eng Aspects 2004; 249:99103.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 41.

    Quirynen MVandermei HCBollen CML, et al. An in-vivo study of the influence of the surface roughness of implants on the microbiology of supragingival and subgingival plaque. J Dent Res 1993; 72:13041309.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 42.

    Tebbs SESawyer AElliott TSJ. Influence of surface-morphology on in-vitro bacterial adherence to central venous catheters. Br J Anaesth 1994; 72:587591.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 43.

    Wertheimer MRFozza ACHollander A. Industrial processing of polymers by low-pressure plasmas: the role of VUV radiation, Nuclear Instrum Methods Phys Res B 1999; 151:6575.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 44.

    MacDonald DEDeo NMarkovic B, et al. Adsorption and dissolution behavior of human plasma fibronectin on thermally and chemically modified titanium dioxide particles. Biomaterials 2002; 23:12691279.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 45.

    Hess HVogel V Molecular shuttles based on motor proteins: active transport in synthetic environments. J Biotech 2001; 82:6785.

  • 46.

    Karakecili AGGumusderelioglu M. Comparison of bacterial and tissue cell initial adhesion on hydrophilic/hydrophobic biomaterials. J Biomater Sci Polym Ed 2002; 13:185196.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 47.

    Hogt AHDankert JFeijen J. Adhesion of coagulase-negative staphylococci to methacrylate polymers and copolymers. J Biomed Mater Res 1986; 20:533545.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 48.

    Roosjen ANorde WMei HC, et al. The use of positively charged or low surface free energy coatings versus polymer brushes in controlling biofilm formation. Progr Colloid Polym Sci 2006; 132:138144.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 49.

    Li BKLogan BE. Bacterial adhesion to glass and metal-oxide surfaces. Colloids Surf B Biointerfaces 2004; 36:8190.

  • 50.

    Grivet MMorrier JJBenay G, et al. Effect of hydrophobicity on in vitro streptococcal adhesion to dental alloys. J Mater Sci Mater Med 2000; 11:637642.

    • Search Google Scholar
    • Export Citation

Advertisement