Surgical stabilization is indicated for the treatment of a variety of vertebral conditions in dogs, including fracture or luxation,1,2 lumbosacral instability, diskospondylitis,1 and congenital malformation.3 In addition, some authors recommend vertebral stabilization following spinal cord decompression via hemilaminectomy or corpectomy.4–8
Several techniques of vertebral stabilization have been described in dogs. Pins or screws and polymethyl-methacrylate,1 vertebral body plates,9 and most other implants are inserted into the ventral aspect of the vertebral body, although screw or pin loosening and migration, infection, and cement breakage are problems reported with this technique as well as with techniques for dorsal insertion.10–12 Furthermore, major vessels, peripheral nerves, and the thoracic cavity are located ventral to the thoracolumbar vertebral bodies and are at risk of injury. Fluoroscopy has been used to minimize iatrogenic soft tissue damage and to provide safe insertion of implants.13–15 In contrast, dorsal vertebral plating, dorsal vertebral stapling,16 and tension-band fixation17 involve use of the dorsal part of the vertebrae to stabilize the vertebral column, which thus averts potential damage to ventrally located soft tissues. Although biomechanics of injured lumbar vertebrae may differ, the predominate motion in intact vertebral columns is flexion, which can be counteracted by dorsal fixation techniques that act as a tension band.18,19
Dorsal stabilization involves the use of the spinous process for implant insertion, but bone purchase is limited. Laminar stabilization is a novel technique that has the advantages of dorsal fixation with additional cortical layers providing stability and involves a surgical approach that is easy to perform. Nevertheless, laminar stabilization involves the risk of spinal cord damage, and care must be taken to avoid such damage.
To our knowledge, guidelines for safe and optimal implantation have not been established for laminar fixation in dogs but are necessary prior to recommending its use in clinical practice. Implantation angles for other vertebral fixation techniques have been established by use of CT.20–22,a The purpose of the study reported here was to determine via CT the point and angle of insertion and safety of laminar stabilization of L1 and L2 in dogs by use of a locking plate for preoperative and postoperative control.
Materials and Methods
Sample—Vertebral columns, including the paraspinous musculature and the supporting ligaments, were harvested from cadavers of 8 dogs (6 Beagles and 2 mixed-breed dogs) euthanized for reasons unrelated to the study. Body weight of each dog at the time of euthanasia was recorded, and cadavers were stored at −20°C and thawed at 21°C for 12 hours prior to the study. Pathological changes were detected via CT and via gross examination during implantation.
Study design—Computed tomography was performed before and after unilateral laminar plating of L1 and L2. The desired angle of insertion of the screws was measured before plating and evaluated in the postoperative CT of the vertebral column. All measurements and laminar fixations were performed by the same investigator (SCK).
CT evaluation—A multislice helical CT scannerb was used to obtain images of T13 through L3. Technique settings were peak voltage of 120 kV, 250 mA (rotation time, 1 second) with 2-mm collimation, and pitch of 0.5 (slice thickness, 2 mm; bone window) in an extended HU scale.c Images were reconstructed in an extended HU scale in 0.6-mm slices in a high-contrast algorithm and evaluated in a bone window (window level, 300 HUs; window width, 1,500 HUs). Analysis was performed by use of imaging software.d
Implants—Laminar fixation was provided by use of a 5-hole locking platee with 2 locking screws (3.0 mm) inserted on the lateral surface of L1 and L2, respectively, and 1 nonlocking screw (2.4 mm) inserted in the facet joint (Figure 1). Laminar screws (locking screws inserted in the dorsal lamina of L1 and L2) and the facet screw (a nonlocking screw inserted into the facet joint between L1 and L2) were examined separately.
Implant measurements—Measurements were determined on CT images obtained for 3 insertion points. To measure the insertion angle, a POI was defined corresponding to the position of the screws. These were the middle of L1 (insertion of the cranial laminar screw), the L1–2 facet joint (insertion of the facet screw), and the middle of L2 (insertion of the caudal laminar screw). The points for L1 and L2 were defined in the center of the vertebra in the transverse section of the CT image. The center was considered the middle of the long axis of the vertebra measured in the sagittal section. For the laminar screws, the center was located in the preoperative image as the most cranial part of the base of the articular facet (the transition from the dorsal lamina to the pedicle surrounding the vertebral canal on the lateral side in a horizontal plane; Figure 2). The POI for the facet joint was located dorsal to the intervertebral disk in the transverse section. It was at a point 0.6 cm ventral to a horizontal line drawn between the dorsal rims of the contralateral 2 facet joints (Figure 3). This distance was based on the position of the plate because the plate was placed with its dorsal rim attached tangential to the dorsal border of the facet joint. The distance of 0.6 cm was half the width of the 5-hole locking plate and the point at which the center of the screw was located.
From these points, a minimum and a maximum angle of insertion was determined in relation to a horizontal line that passed through the transition from the dorsal spinous process and the dorsal lamina of the vertebral canal. The maximum angle was defined as the angle for which the screw would be placed as close as possible to the vertebral canal without penetrating it. The minimum angle was the angle that included 2 cortices of the lamina below the base of the spinous process. This angle was set as a negative value such that screw placement at a more negative angle would penetrate only the spinous process, thereby providing inadequate stability (Figure 4). Minimum and maximum angles were similarly defined for the nonlocking facet screw (Figure 3). For the nonlocking facet screw, the maximum angle was the most ventral angle that would not result in damage of the vertebral canal. The minimum angle was the most dorsal angle that included cortices of the facet joint; this angle was set as a negative value.
After laminar stabilization, the position of the screws was analyzed on postoperative CT images. Images were examined for evidence of penetration of screws into the vertebral canal, insertion angles, comparison of the POI with preoperative measurements, cortices penetrated by the screw, and placement into the spinous process or dorsal to the facet joint.
Points of insertion determined before implantation were compared with position of the screws determined by evaluation of postoperative CT images. The distance between these 2 points along a tangential line representing the cortical layer was measured (Figure 5). Mean deviation and range of the values were recorded.
Surgical stabilization technique—A dorsolateral approach was performed on the right side of each vertebra; each vertebra was positioned on its ventral surface, which corresponded to a patient in sternal position. The lumbodorsal fascia was incised over 4 vertebrae lateral to each spinous process. Muscles were retracted, and the muscle insertion was transected at the facet joint. Gelpi retractors were placed to facilitate exposure. The locking plate was applied unilaterally on the right side of L1 and L2. Knowledge of the angles measured on CT for each individual vertebral column was used to guide implant insertion, with the surgeons taking into account an estimation error of 4°.23 Because locking plates were used, bore holes for the screws were drilled perpendicular to the plate to facilitate locking of the screw heads. The two outermost screw holes on both ends of the plate were torqued at an angle of 5° toward ventral to achieve correct screw placement. Furthermore, the plate was contoured to assist in positioning around the facet joint (Figure 6); bending and adjustment of the plate on the vertebrae was verified visually. The plate was positioned such that the middle hole was located laterally on the facet joint; the second and fourth holes remained empty, and the first and fifth holes were seated on laminar bone (Figure 1).
An adjustable drill guidef was used to determine the point of exit of screws on the opposite side of the base of the dorsal spinous process. The previously measured insertion angle for drilling of the screw hole was used, and an attempt was made to position the screws close to the maximum angle without entering the vertebral canal. The plate was placed on the facet joint and moved ventrally until the cranial and caudal ends of the ventral border of the plate were seated on the lateral part of the dorsal lamina. The dorsal border of the plate did not touch the lamina. The dorsal border of the plate at the facet joint was at the same height as the dorsal tip of the facet joint. Accordingly, the POI was related to the position of the plate. The POI for the laminar screw could be located during implantation by following 3 lines that originated from the accessory process and extended cranially, from the mammillary process and extended caudally, and from the caudal articular process and extended cranially. The point at which these lines intersected was the middle of the vertebra and represented the laminar insertion point (Figure 7).
Drilling was performed first for the cranial and caudal laminar screws, which was followed by drilling for the facet screw. Holes were drilled with a 2.4-mm drill bit for the 2 locking screws (3.0-mm locking screwe) and with a 2.0-mm drill bit for the standard facet screw (2.4-mm nonlocking screwe). Length of each screw was measured, and the self-tapping screws were inserted at the same angle as for the drilled holes.
Data analysis—Descriptive analysis was performed by use of statistical software.g Minimum and maximum angles measured on preoperative images were recorded as mean ± SD. Ranges were also reported. Postoperative angles were evaluated with regard to the preoperative corridor (the value between the minimum and maximum preoperative angles) for each vertebral column.
Results
Preoperative CT measurements—Mean ± SD maximum and minimum insertion angles for L1 were 18 ± 4.0° (range, 12° to 25°) and −11 ± 4.4° (range, −7° to −21°), respectively (Table 1). Mean maximum and minimum insertion angles for L2 were 21 ± 5° (range, 9° to 26°) and −10 ± 3° (range, −8° to −16°), respectively. Mean maximum and minimum insertion angles of the facet screws were 24 ± 4° (range, 18° to 29°) and −12 ± 2° (range, −8° to −14°), respectively.
Preoperative minimum and maximum angles for screw insertion and insertion angles measured after implantation for 24 screws inserted in the vertebral columns obtained from 8 canine cadavers.
L1 | L2 | Facet joint | Measured postoperative angles (°) | |||||||
---|---|---|---|---|---|---|---|---|---|---|
Variable | Body weight (kg) | Minimum angle (°) | Maximum angle (°) | Minimum angle (°) | Maximum angle (°) | Minimum angle (°) | Maximum angle (°) | L1* | L2 | Facet |
Mean ± SD | 22.9 ± 5.25 | −11 ± 4.4 | 18 ± 4.0 | −10 ± 3.0 | 21 ± 5.0 | −12 ± 2.0 | 24 ± 4.0 | 8 ± 7.0 | 11 ± 8.0 | 3 ± 11.0 |
Range | 17 to 30 | −21 to −7 | 12 to 25 | −16 to −8 | 9 to 26 | −14 to −8 | 18 to 29 | −7 to 17 | −2 to 24 | −14 to 17 |
The screw did not follow the drill hole and was not within the planned insertion corridor (value between the minimum and maximum preoperative angles) in 1 vertebra, which resulted in that screw penetrating the vertebral canal.
Postoperative CT measurements—Mean ± SD insertion angle for L1 was 7.9 ± 7.0° (range, −7° to 17°). Two cortices were included with the laminar screws, and the screws were placed in the planned area of L1. Mean length of purchased bone was 9.9 ± 1.9 mm (range, 8 to 13 mm). Inadvertent insertion into the vertebral canal was detected in 1 screw inserted into L1 (Figure 8). In this case, the drill hole appeared to be at the correct angle but the self-tapping screw was inserted and traversed ventral to the drill hole, which created a new screw tract that penetrated the vertebral canal.
Mean ± SD insertion angle for L2 was 11 ± 8° (range, −2° to 24°). Two cortices were included with all the laminar screws, and the screws were placed dor-sally in L2. Mean length of purchased bone was 14.5 ± 2.6 mm (range, 11 to 18 mm). The preoperatively measured angle of insertion was maintained in all cases for L2.
Mean ± SD insertion angle of the facet screw was 3.0 ± 11.0° (range, −14° to 17°; Figure 9). Mean number of cortices included for the facet screws was 6.4 ± 0.7 (range, 5 to 7), and none of the screws was placed dorsal to the facets or into the vertebral canal. Mean length of purchased bone was 13.8 ± 5.2 mm (range, 8 to 23 mm). The preoperatively measured insertion angle was maintained for all of the facet screws.
Implant assessment—Inadvertent insertion into the vertebral canal was detected for 1 of 24 (4%) screws; that screw was inserted outside the drill hole. The remaining inserted screws were dorsal to the vertebral canal. There was no breakage of cortices, and screws were satisfactorily placed with regard to position and bone purchase.
Mean measured deviation from POI and achieved position of the laminar screws was 0.80 mm for L1 and 1.75 mm for L2. Mean measured deviation for the facet screw was 0.85 mm (Table 2).
Mean and range values for the association between the planned POI and the point at which a screw was inserted as determined by use of measurements obtained after implantation.
Distance (mm) | |||
---|---|---|---|
Structure | Direction of deviation | Mean | Range |
L1 | 4 dorsal, 2 ventral, and 2 no deviation | 0.80 | 0–1.60 |
L2 | 6 dorsal and 2 ventral | 1.75 | 0.60–2.60 |
Facetjoint | — | 0.85 | 0.45–3.00 |
Direction of deviation (dorsal or ventral) is indicated for laminar screws; for screws with no deviation, the distance was 0 mm.
— = Not applicable.
Discussion
In the study reported here, the angles of insertion for the placement of screws into the lamina of L1 and L2 and the adjacent facetjoint were evaluated to determine the optimum insertion angles for laminar vertebral stabilization in dogs. Defining these landmarks is important to ensure adequate stability of the implant and to avert damage to adjacent structures. Therefore, we determined a maximum angle, above which the vertebral canal would be penetrated, and a minimum angle, below which stability would be questionable.
Analysis of results of this study indicated that maintaining these angles was feasible. If a common insertion angle, calculated as the smallest angle for all specimens tested, is considered, some of the postoperative angles for screw insertion would not have maintained this guideline regarding damage to the vertebral canal. Steep angles were chosen on the basis of examination of preoperative CT images and because we were attempting to achieve an angle for the drill hole close to the measured maximum angle. Given the suggestion that in clinical patients, a surgeon would attempt to make a horizontal drill hole, we considered that the maximum angle was safe and could be easily achieved, which would avoid the need for preoperative CT for the determination of angles. A horizontal drill hole would provide enough bone purchase and can be considered safe, compared with the minimum angle calculated in the present study. Nonetheless, use of an adjustable drill guide is recommended to drill at precise planned angles and estimate the point of exit.
In the present study, the main implant complication was insertion of a screw into the vertebral canal. Other major structures at risk of damage during vertebral body plating, such as blood vessels or, in the case of stabilization of the thoracic vertebrae, the thorax, are not within the range of the drill corridor for laminar plating.13 Because the drill angles were based on the estimation of the surgeon, a certain margin of error must be considered, and an additional safety margin of ± 4° to allow operator error in estimation has been suggested.24 Results of the present study indicated a sufficiently wide margin (ie, corridor) for safe drilling because none of the drill holes led to penetration of the vertebral canal. In addition, none of the screws was placed too dorsally, which would have resulted in insufficient bony purchase. However, one of the screws was placed outside of the drill hole and penetrated the vertebral canal, which emphasizes the potential danger with the use of self-tapping screws and underscores the importance of maintaining the drill holes during insertion of these screws.
The angle recommended for vertebral body fixation of lumbar vertebrae ranges from 50° to 60° in relation to a vertical axis.20,23,a This is easy to achieve because no soft tissue structures obstruct the drill angle. The almost horizontal placement necessary for laminar fixation can be more difficult because the lateral paraspinal musculature is located in the line of drilling. However, the use of a drill guide was found helpful to facilitate correct placement of the screw. A small-incision approach through the muscles could also be considered to facilitate drilling. In contrast, placement of a dorsal plate requires a relatively superficial approach with little soft tissue dissection, compared with that necessary for vertebral body plates.
In bones with thick cortices, monocortical insertion of locking screws may be acceptable, although bicortical insertion is recommended when a cortex is weak, such as in osteoporotic bones in humans.25 We could confirm the purchase of 2 cortices for the laminar screws and between 5 and 7 cortices for the facet screws, which suggested good stability of the implant. Purchase of the cranial cortex is believed to provide a 25% increase in pullout strength of transpedicular vertebral screws used for vertebral alignment, stabilization, or fusion in humans.26 Compared with vertebral fixation, the number of cortices with laminar screws that were able to obtain purchase was high because it can be dangerous to penetrate the transcortex during vertebral body stabilization. Indeed, the distance from the transcortex of the vertebral body to the caudal vena cava and the aorta can be as little as 3.2 mm in the lumbar region,20 and inadvertent drilling through the far cortex must be confined to a short distance. The risk of damage to underlying structures is even higher for thoracic vertebrae.22 For the facet screws, biomechanical testing in the canine vertebral column should be performed to confirm the contribution of the facet screw to the implant stability, although satisfactory biomechanical stability of transfacet screws has been reported in humans.27,28
The locking plates used in the present study have been used in a number of veterinary applications, including ventral stabilization of the cervical vertebrae in dogs and cats.29–31 To ensure screws remain locked in a plate, the screws must be applied perpendicular to that plate. Bending of plates determines most of the insertion angle; thus, precontouring of plates is extremely important. We found that the ends of the plate must be torqued approximately 5° in a ventral direction to direct the screw toward the insertion angle and to affix the implant close to the bone. The process of plate bending requires some experience, and a considerable degree of bending is required to fit the plate around the facet joint (Figure 6).
The insertion corridors in the study reported here were relevant only to L1 and L2, and additional studies are required to determine insertion corridors for other areas of the vertebral column. It may be more difficult to apply plates to the thoracic vertebrae because of the deeply located, variably shaped articular surfaces, lack of mammillary processes cranial to T11, and short vertebral bodies. This may make it challenging to apply plates, especially in small dogs, but it also is challenging to perform vertebral body plating. The range of the vertebral column in which plating techniques can be used needs to be determined, but in our experience, preliminary clinical results are promising for the caudal portion of the thoracic vertebrae as well as the caudal portion of the lumbar region.
One main consideration was the association between the POI used to measure the preoperative angle and the point at which the screw was actually inserted. If these points do not correspond, the measured angles are of no use. This appears to be even more important because the anatomic location of the POI is difficult to describe. The point for the laminar screw in L1 and the facet screw corresponded with a mean difference of 0.80 and 0.85 mm between the tested specimen, respectively, which was well within the limits for the POI. The position of the screw for L2 differed with a mean of 1.75 mm, but the authors still considered this to be within acceptable limits. Furthermore, the screws could be placed satisfactorily even though the measured angle was not 100% accurate. The difference in the association may have been related to the fact that the implantation proceeded by insertion of laminar screws from cranial to caudal, so it was sometimes difficult to ensure the position of the plate coincided with the cranial and caudal laminar points of insertion because the caudal articular process interfered with the plate and necessitated a more dorsal position for the plate. This may explain the greater deviation in the median association of the screw in L2. However, the results still indicated a satisfactory position in these segments and a similar range of angles for the screw insertions.
Finally, the mean angles were calculated from only a small number of vertebral columns obtained from dogs with a limited range of body weights. To determine more specific angles, a larger number of cadavers of various sizes should be examined.
Safe insertion angles for the facet joint and lamina of L1 and L2 for laminar stabilization were confirmed in the study reported here. If appropriate angles are known, then screws can be placed safely. The intraoperative feasibility of maintaining the measured angles was confirmed by evaluation of CT images obtained during implant placement. Purchase into several cortices, especially for the facet screws, was verified. Superior implant stability for this technique, compared with that for other stabilization techniques, has been reported in an in vitro biomechanical study.32
Abbreviations
CT | Computed tomography |
HU | Hounsfield unit |
POI | Point of insertion |
Walker TM, Tucker R, Welsh RD, et al. Fluoroscopic placement of transfixation pins for the external skeleton fixation of the canine spine: an anatomical study (abstr). Vet Comp Orthop Traumatol 2001;14:A13.
Siemens SOMATOM Sensation Open 40, Siemens, Erlangen, Germany.
Syngo CT 2007 Software, Siemens, Erlangen, Germany.
OsiriX Imaging Software, version 3.6.1 (32-bit), OsiriX Foundation, Geneva, Switzerland.
UniLock 2.4 System, provided by Synthes, Solothurn, Switzerland.
Veterinary Instrumentation, Sheffield, South Yorkshire, England.
NCSS Number Cruncher Statistical Systems, Kaysville, Utah.
References
- 1.↑
Sharp NJH, Wheeler SJ. Trauma. Small animal spinal disorders. 2nd ed. London, England: Elsevier Mosby, 2005;281–318.
- 2.
Weh JM, Kraus KH. Use of a four pin and methylmethacrylate fixation in L7 and the iliac body to stabilize lumbosacral fracture-luxations: a clinical and anatomic study. Vet Surg 2007; 36: 775–782.
- 3.↑
Aikawa T, Kanazono S, Yoshigae Y, et al. Vertebral stabilization using positively threaded profile pins and polymethylmethacrylate, with or without laminectomy, for spinal canal stenosis and vertebral instability caused by congenital thoracic vertebral anomalies. Vet Surg 2007; 36: 432–441.
- 4.
Macias C, McKee WM, May C, et al. Thoracolumbar disc disease in large dogs: a study of 99 cases. J Small Anim Pract 2002; 43: 439–446.
- 5.
McKee WM, Downes CJ. Vertebral stabilisation and selective decompression for the management of triple thoracolumbar disc protrusions. J Small Anim Pract 2008; 49: 536–539.
- 6.
McKee WM. Intervertebral disc disease in the dog: part 2. In Pract 2000; 22: 458–471.
- 7.
Creed J, Yturraspe DJ. Thoracolumbar spine. In: Bojrab M, ed. Current techniques in small animal surgery. Baltimore, Md: Williams & Wilkins, 1998;835–858.
- 8.
McKee WM. Intervertebral disc disease in the dog: part 1. In Pract 2000; 22: 355–339.
- 9.↑
Swaim SF. Vertebral body plating for spinal immobilization. J Am Vet Med Assoc 1971; 158: 1683–1695.
- 10.
Blass C, Waldron D, Van Ee R. Cervical stabilisation in three dogs using Steinmann pins and methylacrylate. J Am Anim Hosp Assoc 1988; 24: 61–68.
- 11.
Blass C, Seim H. Spinal stabilisation in dogs using Steinmann pins and methylacrylate. Vet Surg 1984; 13: 203–210.
- 12.
Wong W, Emms S. Use of pins and methylacrylate in stabilisation of spinal fractures and luxations. J Small Anim Pract 1992; 33: 415–422.
- 13.↑
Wheeler JL, Lewis DD, Cross AR, et al. Closed fluoroscopic-assisted spinal arch external skeletal fixation for the stabilization of vertebral column injuries in five dogs. Vet Surg 2007; 36: 442–448.
- 14.
Tomlinson JL, Cook JL, Payne JT, et al. Closed reduction and lag screw fixation of sacroiliac luxations and fractures. Vet Surg 1999; 28: 188–193.
- 15.
Cook J, Tomlinson JL, Reed AL. Fluoroscopically guided closed reduction and internal fixation of fractures of the lateral portion of the humeral condyle: prospective clinical study of the technique and results in ten dogs. Vet Surg 1999; 28: 315–321.
- 16.↑
Bruecker KA. Surgical treatment of spinal fractures, luxations and subluxations. In: Bojrab M, ed. Current techniques in small animal surgery. Baltimore, Md: Williams & Wilkins, 1998;988–996.
- 17.↑
Voss K, Montavon PM. Tension band stabilization of fractures and luxations of the thoracolumbar vertebrae in dogs and cats: 38 cases (1993–2002). J Am Vet Med Assoc 2004; 225: 78–83.
- 18.
Lang B, Loeffler K. Bewegungsmöglichkeiten der Wirbelsäule von Hund und Katze. Kleintierpraxis 1972; 17: 217–244.
- 19.
Chung SK, Kim YE, Wang KC. Biomechanical effect of constraint in lumbar total disc replacement: a study with finite element analysis. Spine (Phila Pa 1976) 2009; 34: 1281–1286.
- 20.↑
Watine S, Cabassu JP, Catheland S, et al. Computed tomography study of implantation corridors in canine vertebrae. J Small Anim Pract 2006; 47: 651–657.
- 21.
Corlazzoli D. Bicortical implant insertion in caudal cervical spondylomyelopathy: a computed tomography simulation in affected Doberman Pinschers. Vet Surg 2008; 37: 178–185.
- 22.↑
Wheeler S, Cross AR, Rapoff AJ. A comparison of the accuracy and safety of vertebral body pin placement using a fluoroscopically guided versus an open surgical approach: an in vitro study. Vet Surg 2002; 31: 468–474.
- 23.↑
Walker TM, Pierce WA, Welch RD. External fixation of the lumbar spine in a canine model. Vet Surg 2002; 31: 181–188.
- 24.↑
Shales CJ, Langley-Hobbs SJ. Canine sacroiliac luxation: anatomic study of dorsoventral articular surface angulation and safe corridor for placement of screws used for lag fixation. Vet Surg 2005; 34: 324–331.
- 26.↑
Weinstein JN, Rydevik BL, Rauschning W. Anatomic and technical considerations of pedicle screw fixation. Clin Orthop Relat Res1992: 34–46.
- 27.
Klekamp JW, Ugbo JL, Heller JG, et al. Cervical transfacet versus lateral mass screws: a biomechanical comparison. J Spinal Disord 2000; 13: 515–518.
- 28.
Liu GY, Xu RM, Ma WH, et al. Biomechanical comparison of cervical transfacet pedicle screws versus pedicle screws. Chin Med J (Engl) 2008; 121: 1390–1393.
- 29.
Voss K, Keller M, Montavon P. Internal splinting of dorsal intertarsal and tarsometatarsal instabilities in dogs and cats with the ComPact UniLock 2.0/2.4TM System*. Vet Comp Orthop Traumatol 2004; 17: 125–130.
- 30.
Voss K, Steffen F, Montavon PM. Use of the ComPact UniLock System for ventral stabilization procedures of the cervical spine: a retrospective study. Vet Comp Orthop Traumatol 2006; 19: 21–28.
- 31.
Keller MA, Voss K, Montavon PM. The ComPact UniLock 2.0/2.4 system and its clinical application in small animal orthopedics. Vet Comp Orthop Traumatol 2005; 18: 83–93.
- 32.↑
Knell SC, Bürki A, Hurter K, et al. Biomechanical comparison after in vitro laminar vertebral stabilization and vertebral body plating of the first and second lumbar vertebrae in specimens obtained from canine cadavers. Am J Vet Res 2011; 72: 1681–1686.