Comparison of Doppler ultrasonography and high-definition oscillometry for blood pressure measurements in healthy awake dogs

Valérie Chetboul Unité de Cardiologie d'Alfort, Unité Mixte de Recherche INSERM U955, Ecole Nationale Vétérinaire d'Alfort, 94 704 Maisons-Alfort cedex, France; and Unité Mixte de Recherche 181 INRA, ENVT, Physiopathologie et Toxicologie Expérimentales

Search for other papers by Valérie Chetboul in
Current site
Google Scholar
PubMed
Close
 DVM, PhD
,
Renaud Tissier Unité Mixte de Recherche INSERM U955, Ecole Nationale Vétérinaire d'Alfort, 94 704 Maisons-Alfort cedex, France; and Unité Mixte de Recherche 181 INRA, ENVT, Physiopathologie et Toxicologie Expérimentales, Unité de Pharmacie-Toxicologie, Ecole Nationale Vétérinaire d'Alfort, 94 704 Maisons-Alfort cedex, France; and Unité Mixte de Recherche 181 INRA, ENVT, Physiopathologie et Toxicologie Expérimentales

Search for other papers by Renaud Tissier in
Current site
Google Scholar
PubMed
Close
 DVM, PhD
,
Vassiliki Gouni Unité de Cardiologie d'Alfort, Ecole Nationale Vétérinaire d'Alfort, 94 704 Maisons-Alfort cedex, France; and Unité Mixte de Recherche 181 INRA, ENVT, Physiopathologie et Toxicologie Expérimentales

Search for other papers by Vassiliki Gouni in
Current site
Google Scholar
PubMed
Close
 DVM
,
Virginie de Almeida Unité de Pathologie Médicale, Ecole Nationale Vétérinaire d'Alfort, 94 704 Maisons-Alfort cedex, France; and Unité Mixte de Recherche 181 INRA, ENVT, Physiopathologie et Toxicologie Expérimentales

Search for other papers by Virginie de Almeida in
Current site
Google Scholar
PubMed
Close
,
Hervé P. Lefebvre Département des Sciences Cliniques, Ecole Nationale Vétérinaire de Toulouse, 23 chemin des Capelles, BP 87614, F31076 Toulouse cedex 03, France

Search for other papers by Hervé P. Lefebvre in
Current site
Google Scholar
PubMed
Close
 DVM, PhD
,
Didier Concordet Département des Sciences Cliniques, Ecole Nationale Vétérinaire de Toulouse, 23 chemin des Capelles, BP 87614, F31076 Toulouse cedex 03, France

Search for other papers by Didier Concordet in
Current site
Google Scholar
PubMed
Close
 PhD
,
Nathalie Jamet Unité de Pathologie Médicale, Ecole Nationale Vétérinaire d'Alfort, 94 704 Maisons-Alfort cedex, France; and Unité Mixte de Recherche 181 INRA, ENVT, Physiopathologie et Toxicologie Expérimentales

Search for other papers by Nathalie Jamet in
Current site
Google Scholar
PubMed
Close
,
Carolina Carlos Sampedrano Unité de Cardiologie d'Alfort, Ecole Nationale Vétérinaire d'Alfort, 94 704 Maisons-Alfort cedex, France; and Unité Mixte de Recherche 181 INRA, ENVT, Physiopathologie et Toxicologie Expérimentales

Search for other papers by Carolina Carlos Sampedrano in
Current site
Google Scholar
PubMed
Close
 DVM, PhD
,
François Serres Unité de Cardiologie d'Alfort, Ecole Nationale Vétérinaire d'Alfort, 94 704 Maisons-Alfort cedex, France; and Unité Mixte de Recherche 181 INRA, ENVT, Physiopathologie et Toxicologie Expérimentales

Search for other papers by François Serres in
Current site
Google Scholar
PubMed
Close
 DVM
, and
Jean-Louis Pouchelon Unité de Cardiologie d'Alfort, Unité Mixte de Recherche INSERM U955, Ecole Nationale Vétérinaire d'Alfort, 94 704 Maisons-Alfort cedex, France; and Unité Mixte de Recherche 181 INRA, ENVT, Physiopathologie et Toxicologie Expérimentales

Search for other papers by Jean-Louis Pouchelon in
Current site
Google Scholar
PubMed
Close
 DVM, PhD

Abstract

Objective—To determine the intra- and interobserver variability of systolic arterial pressure (SAP) and diastolic arterial pressure (DAP) measurements obtained with 2 indirect methods in awake dogs and percentage of successful measurements.

Animals—6 healthy conscious adult dogs.

Procedures—4 observers with different levels of training measured SAP and DAP on 4 days by use of Doppler ultrasonography (DU) and high-definition oscillometry (HDO). The examinations were randomized. Measurements for each technique were recorded 5 consecutive times, and mean values (total, 720 measurements) were used for statistical analysis.

Results—All within- and between-day coefficients of variation (CVs) for SAP were < 15% irrespective of the observer or method (HDO, 3.6% to 14.1%; DU, 4.1% to 12.4%). Conversely, half the CVs for DAP were > 15% with the highest within- and between-day CVs obtained by the least experienced observer by use of DU (19.5% and 25.9%, respectively). All attempts with HDO were successful, whereas DAP could not be measured by use of DU by the least experienced observer in 17% of attempts.

Conclusions and Clinical Relevance—SAP may be assessed in healthy dogs by use of DU and HDO with good repeatability and reproducibility after a short period of training. Conversely, the variability of DAP is higher and longer training is required to assess DAP via DU than via HDO.

Abstract

Objective—To determine the intra- and interobserver variability of systolic arterial pressure (SAP) and diastolic arterial pressure (DAP) measurements obtained with 2 indirect methods in awake dogs and percentage of successful measurements.

Animals—6 healthy conscious adult dogs.

Procedures—4 observers with different levels of training measured SAP and DAP on 4 days by use of Doppler ultrasonography (DU) and high-definition oscillometry (HDO). The examinations were randomized. Measurements for each technique were recorded 5 consecutive times, and mean values (total, 720 measurements) were used for statistical analysis.

Results—All within- and between-day coefficients of variation (CVs) for SAP were < 15% irrespective of the observer or method (HDO, 3.6% to 14.1%; DU, 4.1% to 12.4%). Conversely, half the CVs for DAP were > 15% with the highest within- and between-day CVs obtained by the least experienced observer by use of DU (19.5% and 25.9%, respectively). All attempts with HDO were successful, whereas DAP could not be measured by use of DU by the least experienced observer in 17% of attempts.

Conclusions and Clinical Relevance—SAP may be assessed in healthy dogs by use of DU and HDO with good repeatability and reproducibility after a short period of training. Conversely, the variability of DAP is higher and longer training is required to assess DAP via DU than via HDO.

Systemic arterial hypertension is defined as a sustained increase in arterial BP, which may occur in the absence of any identifiable cause or may be secondary to various conditions or pathological processes, including chronic kidney disease and several endocrine diseases (eg, hyperadrenocorticism, diabetes mellitus, primary hyperaldosteronism, pheochromocytoma, and hypothyroidism).1–6

Over the past 10 years, systemic hypertension has been increasingly recognized as a potential cause of severe irreversible organ lesions in veterinary medicine. The main target organs in cats and dogs include the kidneys (glomerular sclerosis and interstitial fibrosis3,4), eyes (retinopathy and choroidopathy),7,8 CNS (hemorrhage, edema, and infarction),9,10 and heart (myocardial hypertrophy with systolic and diastolic dysfunction11–13). An early and accurate diagnosis of systemic arterial hypertension is therefore highly recommended to limit the degree of such injuries or prevent their occurrence in patients at risk.1 A reliable diagnosis is also needed to follow-up and control the efficacy of antihypertensive treatments.1,9,14–18

Diagnosis of systemic hypertension relies on measurements of systemic BP, which may be directly obtained by arterial blood puncture. However, in clinical practice, indirect measurements made by use of devices incorporating a compressive cuff are usually preferred because of their noninvasive aspect. The most commonly used methods are therefore DU and oscillometry or HDO.1 In dogs, the accuracy of these indirect techniques has been reported in normotensive19–21 and hypertensive21 animals, either conscious19,20 or anesthetized,21 in comparison with direct intra-arterial methods. The comparative diagnostic abilities of oscillometric and DU methods in detecting systemic hypertension in dogs have also been assessed.22,23 A consensus statement of the ACV1M has recently been published, providing guidelines to veterinarians regarding their correct use in practice.1 According to the guidelines, a well-defined standard protocol involving suitably trained observers should be followed because technical errors associated with personnel inexperience are major causes of unreliable BP measurements.1 However, to the best of our knowledge, the precise definition of suitable training has not been established and the intra- and interobserver within- and between-day variability of BP measurements have not been evaluated in dogs in clinical settings.

The purpose of the present study was to determine the intra- and interobserver variability and percentage of successful measurements of SAP and DAP measurements by use of 2 indirect methods in awake dogs.

Materials and Methods

Dogs—Six healthy female Labrador Retrievers (mean ± SD age, 4.9 ± 0.9 years, [range, 2.8 to 7.8 years]; weight, 25.8 ± 0.7 kg, [24.8 to 27.1 kg]) from a canine cohort of the National Veterinary School of Alfort were used, and procedures were conducted in accordance with guidelines established by the National Institutes of Health Guide for the Care and Use of Laboratory Animals. The dogs were considered healthy on the basis of results of a complete clinical examination, electrocardiography, and conventional echocardiography performed before inclusion in the study. The dogs were not familiar with the investigators or the BP measurement procedure.

BP measurements—An indirect measurement of BP was carefully obtained in conscious dogs by use of a standardized method according to manufacturers’ instructionsa,b and guidelines of the ACV1M consensus statements.1 The same single DUa and HDOb devices were used throughout the study. Stress and anxiety were reduced as much as possible by performing all BP measurements in the same isolated quiet room and use of a 10-minute period of acclimatization each day before starting the measurements. For both techniques, the dogs were gently restrained in right or left lateral recumbency and the inflatable cuff was placed on the tail. 1n accordance with the ACV1M consensus statement,1 several consecutive BP measurements were performed during each session to obtain a stable set of 5 values and the mean was used for the statistical analyses. Observers could choose to discard the first value if it was considered abnormally high and anxiety-induced.1 The time required to obtain the BP measurements, from cuff placement to end of recording of the 5 consecutive conserved BP values, was recorded by another person.

DU method—An appropriate-size cuffc was placed around the tail. Coupling gel was applied between the 8-MHz probe and the skin to improve contact. The hair was not clipped before placing the probe. As described,22,23 the sound volume of the Doppler device was adjusted to obtain a clearly audible signal. The cuff was manually inflated until the pulse signal was no longer audible and then was gradually deflated. The BP (read on the manometer) at which the audible pulse signal was again detected was considered to be the SAP. The cuff was further deflated until a change in tone of the flow sound was detected, and this was recorded as the DAP. Heart rate was calculated by each observer with a stethoscope at the end of each session of BP measurements.

HDO method—As with the DU method, an appropriate-size cuffd was placed around the tail and automatically inflated. For each cuff inflation and deflation, the heart rate, SAP, DAP, and MAP were automatically calculated by the HDO device.

Observers—Four observers from the National Veterinary School of Alfort with different levels of experience were involved in the study. Observer 1 (VD) was a fourth-year student with only 1 hour of training in BP measurements (DU and HDO) before starting the study. Observer 2 (NJ) was an intern who underwent the same training as observer 1 but had also done some occasional training during a year of internship. Observers 3 (FS) and 4 (VG) were residents in veterinary cardiology and were both used to performing daily BP measurements by use of the 2 techniques during their residency program (observer 4 had 2 more years of experience than observer 3). Observers 1, 2, and 3 had been trained in use of the 2 methods (DU and HDO) by observer 4.

Assessment of within- and between-day intra- and interobserver variability—All BP measurements were randomized (for the order of examinations, dogs and investigators were randomly drawn), and each observer was masked to the BP values recorded by the 3 others. The study was performed on 4 consecutive days. Each day, each observer (n = 4) performed 3 BP measurements on 3 dogs by use of each method. Therefore, for each method, 36 BP measurements/d were performed, representing, for the whole 4-day study period, a total of 144 BP measurements (ie, 720 BP recordings because each BP value used for statistical analysis was the mean of 5 consecutive values). The same dog was not used for 2 consecutive BP measurements.

Statistical analysis—Data are expressed as mean ± SD. A software programe was used to perform the statistical analysis as described.24 The following linear model was used for each observer and each BP measurement:

Yijkl = μ + dayj + dogk + (day × dog)jk + £ijkl

where is the first value measured for dog k on day j by observer i, μ is the general mean, dogk is the differential effect (considered as fixed) of dog k, (day × dog)jk is the interaction term between day and dog, and ϵ k is the model error. The SD of repeatability was determined from the residual SD of the model and the SD of reproducibility from the square root of the mean square of day.

Any interaction between dog and operator was determined by use of the following general linear model:

Yijkl = μ + dayj + Obi + dogk + (Ob × dog)ik + (day × dog)jk + £dayijkl

where Yijkl is the first value measured for dog k on day j by observer i, μ is the general mean, Ob is the differential effect (considered as fixed) of observer i, dogk is the differential effect of dog k, (Ob × dog)ik is the interaction term between observer and dog, (day × dog)jk is the interaction term between day and dog, and ϵ k is the model error. A similar general linear model was used to determine for each technique the observer effect on the time taken to measure BP. The level of significance was set at values of P < 0.05.

Results

Mean ± SD values and ranges of BP values (SAP and DAP for both techniques and MAP for HDO) obtained by the reference observer (observer 4) were determined (Table 1). The SDs and CVs for the within- and between-day intraobserver variability were determined for HDO (Table 2) and DU (Table 3).

Table 1—

Repeated measurements of SAP (mean ± SD [range]) performed by a reference observer (observer 4) in 6 healthy dogs by use of HDO and DU.

VariableHDODU
SAP (mm Hg)117 ± 12(93–149)107 ± 10(91–127)
DAP (mm Hg)56 ± 8(43–76)70 ± 13(45–91)
MAP (mm Hg)78 ± 9 (54–97)
Heart rate (beats/min)77 ± 15(53–118)65 ± 14(40–100)
Table 2—

Within- and between-day intraobserver SDs and CVs for indirect measurement of BP in 6 healthy Labradors Retrievers by use of HDO.

VariableObserver 1Observer 2Observer 3Observer 4
SAP (mm Hg)
  Within-day variability
   SD12.8 mm Hg10.0 mm Hg10.2 mm Hg13.1 mm Hg
   CV10.7%8.6%8.8%11.2%
  Between-day variability
   SD16.9 mm Hg8.5 mm Hg4.2 mm Hg11.4 mm Hg
   CV14.1%7.3%3.6%9.7%
DAP (mm Hg)
  Within-day variability
   SD9.2 mm Hg8.0 mm Hg9.6 mm Hg8.4 mm Hg
   CV15.5%13.7%16.3%14.9%
  Between-day variability
   SD6.4 mm Hg10.2 mm Hg10.5 mm Hg8.2 mm Hg
   CV10.9%17.6%18.0%14.6%
MAP (mm Hg)
  Within-day variability
   SD9.7 mm Hg7.7 mm Hg9.6 mm Hg9.6 mm Hg
   CV12.1%9.8%12.2%12.4%
  Between-day variability
   SD8.2 mm Hg11.8 mm Hg5.4 mm Hg10.3 mm Hg
   CV10.2%15.0%6.9%13.3%
Table 3—

Within- and between-day intraobserver SDs and CVs for indirect measurement of BP in 6 healthy Labradors Retrievers by use of DU.

VariableObserver 1Observer 2Observer 3Observer 4
SAP (mm Hg)
  Within-day variability
   SD11.4 mm Hg7.3 mm Hg7.0 mm Hg7.9 mm Hg
   CV10.0%6.6%6.2%7.4%
  Between-day variability
   SD14.2 mm Hg8.8 mm Hg4.6 mm Hg10.1 mm Hg
   CV12.4%7.9%4.1%9.4%
DAP (mm Hg)
  Within-day variability
   SD13.5 mm Hg10.3 mm Hg6.3 mm Hg9.1 mm Hg
   CV19.5%13.7%10.4%13.1%
  Between-day variability
   SD18.0 mm Hg7.8 mm Hg11.1 mm Hg13.8 mm Hg
   CV25.9%10.4%18.3%19.9%

For the BP measurements obtained by use of HDO and for all investigators, the lowest CVs of within- and between-day variability were obtained for SAP and MAP, with all values (n = 16) < 15% (ranges, 3.6% to 14.1% and 6.9% to 15.0%, respectively). More than half the within- and between-day CVs for SAP (5/8) were < 10%. Conversely, all within- and between-day CVs of HDO-assessed DAP were > 10% (range, 10.9% to 18.0%) and half (4/8) were > 15%.

For the BP measurements obtained by use of DU (Table 3) and for all investigators, the lowest CVs of within- and between-day variability were obtained for SAP, with all values < 15% (range, 4.1% to 12.4%), and all except 1 (7/8) were < 10%. Conversely, all CVs of DU-assessed DAP were > 10% (range, 10.4% to 25.9%), with the highest values obtained by observer 1 (within- and between-day CVs of 19.5% and 25.9%, respectively). No observer effect and no interaction between dog and observer or between dog and day were observed with either technique.

Duration of measurements and comparison of values—No significant observer effect was observed for the time taken to obtain BP measurements by use of the HDO technique (5.7 ± 2.1 minutes [range, 3.5 to 17 minutes] for all 4 observers and 5.4 ± 1.5 minutes, 6.0 ± 2.3 minutes, 5.3 ± 1.4 minutes, and 6.2 ± 2.9 minutes for observers 1, 2, 3, and 4, respectively; Figure 1). Regarding HDO, no significant day or dog effect or any interaction between day and dog or dog and observer was observed. Conversely, the time taken to obtain BP measurements by use of DU (5.9 ± 4.0 minutes [range, 3.5 to 31 minutes]) was significantly (P = 0.001) dependant on the observers (7.7 ± 4.7 minutes, 6.1 ± 4.8 minutes, 4.3 ± 1.6 minutes, and 5.4 ± 3.6 minutes for observers 1, 2, 3, and 4, respectively; Figure 2). A significant (P = 0.024) dog effect was also found for the DU technique. However, no day effect or interactions between day and dog or dog and observer were observed.

Figure 1—
Figure 1—

Box plots representing the total durations of blood pressure measurements performed by 4 observers with different levels of experience (observers 1, 2, 3, and 4) by use of the HDO method. The horizontal line within the box represents the 50th percentile, the box represents the 25th to 75th percentiles, and the whiskers represent the 5th and 95th percentiles. Each measurement (n = 144 [36/observer]) was obtained 5 consecutive times, and the time required for the 5 readings was recorded.

Citation: American Journal of Veterinary Research 71, 7; 10.2460/ajvr.71.7.766

Figure 2—
Figure 2—

Box plots representing the total durations of blood pressure measurements performed by 4 observers with different levels of experience (observers 1, 2, 3, and 4) by use of the DU method. Each measurement (n = 144 [36/observer]) was obtained 5 consecutive times, and the time required for the 5 readings was recorded. See Figure 1 for key.

Citation: American Journal of Veterinary Research 71, 7; 10.2460/ajvr.71.7.766

Figure 3—
Figure 3—

Bland-Altman analysis illustrating the difference between SAP values assessed by use of 2 indirect methods (SAP DU-HDO) versus the mean of those 2 values. Mean difference is indicated as the middle horizontal line, and the limits of agreement (95% confidence interval) are indicated as the upper and lower horizontal lines.

Citation: American Journal of Veterinary Research 71, 7; 10.2460/ajvr.71.7.766

For the DU technique, but not for HDO, observers 1 and 2 could not obtain DAP values in 6 of 36 (17%) and 2 of 36 (6%) sessions, respectively, because they were totally unable to detect a clear change from systolic to diastolic Doppler signal tone.

No significant differences were observed between the systolic or DAP values obtained by use of DU and HDO after each examination. Further comparisons were made by examining the differences in SAP (Figure 3) and DAP (Figure 4) values obtained with the 2 methods by use of Bland-Altman representations. Bland-Altman analysis revealed that the DU method underestimated SAP, compared with the HDO method (mean difference between SAP assessed by use of DU and HDO, −5.8 mm Hg [95% confidence interval, −34.2 to 22.4 mm Hg]). Conversely, the DU technique overestimated DAP, compared with HDO (mean difference between DAP assessed by DU and HDO, +9.7 mm Hg [95% confidence interval, −23.5 to 42.8 mm Hg]).

Figure 4—
Figure 4—

Bland-Altman analysis illustrating the difference between DAP values assessed by use of 2 indirect methods (DAP DU-HDO) versus the mean of those 2 values. Mean difference is indicated as the middle horizontal line, and the limits of agreement (95% confidence interval) are indicated as the upper and lower horizontal lines.

Citation: American Journal of Veterinary Research 71, 7; 10.2460/ajvr.71.7.766

Discussion

Few studies23,25,26 have focused on BP variability in veterinary medicine. To the authors’ knowledge, the present report is the first in which the intra- and interobserver variability of BP measurements have been assessed in dogs in relation to the method used (HDO or DU). Such data are prerequisite to any longitudinal studies performed with repeated BP measurements to avoid misinterpretation of the results. Repeated measurements are used to assess antihypertensive drug efficacy in clinical trials and experimental studies,14–18 and the amplitude of BP change should always be interpreted in relation to observer variability. If the decrease in BP is less than the observer's SD of variability, this change should not be interpreted as a therapeutic improvement. Although results of the present study cannot be extrapolated to other conditions (ie, investigator or device), the proposed procedure can easily be applied by other investigators to evaluate their own performance.

The main goal of this study was not to document biological (ie, interdog) variability but to obtain numerous data from different investigators on the same small number of subjects (n = 6), to determine inter- and intraobserver variability. For both techniques, HDO and DU, the total number of sessions of BP measurements was high (288, representing 1,440 readings).

In this study, the 3 most experienced observers (1 intern and 2 residents) were representative of investigators who might be involved in longitudinal studies. A beginner (a student) was also deliberately involved in the protocol to evaluate whether a short training period (1 hour) was sufficient to obtain satisfactory to good repeatability and reproducibility of BP measurements in dogs.

The authors decided that, to reproduce conditions similar to those of clinical trials and also respect the recommendations for standard BP measurement procedures in the ACVIM consensus statement,1 the dogs should not have been acclimatized to the examination procedure or accustomed to the observers, the dogs should not be sedated or anesthetized, and the procedure should be performed, as in a clinical situation, near the consulting room in a quiet place. Moreover, to better compare the variability of the 2 techniques and minimize other factors of variation, the same site of cuff placement (tail) and the same dog position (lateral recumbency) were used for both techniques. The same HDO and DU devices were used by all 4 observers. Healthy dogs were selected to limit daily pathological variations in BP values during the protocol, which might have interfered with the assessment of the intra- and interobserver intrinsic variability.

In the present study, the procedure used for BP measurements, which was in accordance with the ACVIM consensus statement,1 allowed each observer to discard the first BP values if these were considered abnormally high. This could have lead to an underestimation of BP variability, and the present results are only valid under these conditions. Nevertheless, the objective of this study was to assess the variability of BP measurement as actually done by the general practitioner following published guidelines.1

Mean values of repeated BP measurements obtained in the 6 healthy dogs were similar to those described in conscious dogs but in the lower ranges for systolic and diastolic values, irrespective of the technique used.1 These results may be explained by the fact that the Labrador Retrievers were docile, as indicated by the low heart rates recorded during the BP measurement sessions (77 ± 15 beats/min and 65 ± 14 beats/min for HDO and DU, respectively). Moreover, low BP values have already been described in Labrador Retrievers, compared with other canine breeds.27

In the present study, the SAP and DAP values obtained by use of DU and HDO devices were not significantly different. Similarly, in a previous study23 performed in conscious healthy cats, no significant difference was observed between mean SAP readings when results obtained with DU and HDO devices were compared. However, the HDO machine produced significantly higher estimates of DAP in that previous study. In the study reported here, results did not indicate similar findings in dogs, thus confirming the need to obtain specific reference values for each species with each technique. However, the present study was performed in a small number of healthy dogs. Further studies involving large populations of healthy and diseased dogs are needed.

The present study revealed several advantages of the HDO device, compared with the DU device, when used in dogs. The HDO technique requires minimal skill, probably because HDO measurements are automatically done by the device after cuff placement. Thus, the technique does not rely on detection of the arterial Doppler signal or on the interpretation of flow sound variations by the observer.

For the 2 most experienced observers (observers 3 and 4), the mean time taken to obtain the BP measurements by use of the DU technique was similar to that recorded in another study.26 Hsiang et al26 reported that 5 consecutive systemic BP readings were obtained via Doppler sphygmomanometry in all dogs within 6 minutes. Similarly, in the present study, the time taken to obtain BP measurements via DU was 4.3 ± 1.6 minutes and 5.4 ± 3.6 minutes for observers 3 and 4, respectively. However, the 2 less experienced observers (observers 1 and 2) took longer to do the BP measurements, suggesting that the DU technique requires a minimum degree of training. 1n contrast, all 4 observers took similar times (5.7 ± 2.1 minutes) to obtain BP measurements via HDO.

In the present study, all 4 observers obtained SAP, DAP, and MAP values in 100% of their attempts with the HDO device. Similarly, SAP was measured in all attempts by all 4 observers using the DU technique. 1n contrast, observers 1 and 2 were unable to obtain DAP values in 17% and 6% of cases, respectively, with the DU device, whereas observers 3 and 4 were able to detect a clear change from systolic to diastolic tone in 100% of their attempts. Diastolic arterial pressure is reported to be more difficult to assess in conscious dogs and cats.19,23 1n the report of Jepson et al,23 combining the data from both examiners involved in their study, with 5 attempts each for 28 cats, the DU device gave an SAP reading in all 280 attempts, whereas DAP readings were determined in only 144 of 280 (51.4%) attempts.

Results of the present study indicated that investigators may have different variabilities for BP measurements with the highest within- and between-day CVs obtained for DAP by the least experienced observer with the DU technique (19.5% and 25.9%, respectively). These data are of importance to ensure correct interpretation of BP variations in practice. For example, if a decrease in DAP of 17 mm Hg is observed by observer 1 between 2 consecutive DU measurements after the administration of an antihypertensive drug, it cannot be concluded that this change is an effect of treatment because it is lower than the between-day SD of variability. Conversely, the same decrease obtained by observer 4 might result from the hypotensive drug effect.

In conclusion the HDO machine allowed assessment of SAP, DAP, and MAP within a short period in 100% of attempts with correct to good repeatability and reproducibility even after a minimal period of training (1 hour). Conversely, differences among investigators were observed regarding speed, variability, and percentage of successful DAP measurements with the DU machine.

ABBREVIATIONS

ACVIM

American College of Veterinary Internal Medicine

BP

Blood pressure

CV

Coefficient of variation

DAP

Diastolic arterial pressure

DU

Doppler ultrasonography

HDO

High-definition oscillometry

MAP

Mean arterial pressure

SAP

Systolic arterial pressure

a.

811-BL, Parks Medical Electronics 1nc, Aloha, Ore.

b.

Memodiag HDO, S + B medVet, Babenhausen, Germany

c.

Soft-cuff, Ref 2525, 5 cm-large, Parks Medical Electronics 1nc, Aloha, Ore.

d.

Blue cuff, D2, S + B medVet, Babenhausen, Germany.

e.

Systat, version 10.0, SPSS 1nc, Chicago, 1ll.

References

  • 1.

    Brown S, Atkins C, Bagley R, et alGuidelines for the identification, evaluation, and management of systemic hypertension in dogs and cats. J Vet Intern Med 2007;21:542558.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 2.

    Mishina M, Watanabe T. Development of hypertension and effects of benazepril hydrochloride in a canine remnant kidney model of chronic renal failure. J Vet Med Sci 2008;70:455460.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 3.

    Wehner A, Hartmann K, Hirschberger J. Associations between proteinuria, systemic hypertension and glomerular filtration rate in dogs with renal and non-renal diseases. Vet Rec 2008;162:141147.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 4.

    Finco DR. Association of systemic hypertension with renal injury in dogs with induced renal failure. J Vet Intern Med 2004;18:289294.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 5.

    Syme HM, Barber PJ, Markwell PJ, et alPrevalence of systolic hypertension in cats with chronic renal failure at initial evaluation. J Am Vet Med Assoc 2002;220:17991804.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 6.

    Sennello KA, Schulman RL, Prosek R, et alSystolic blood pressure in cats with diabetes mellitus. J Am Vet Med Assoc 2003;223:198201.

  • 7.

    Sansom J, Rogers K, Wood JL. Blood pressure assessment in healthy cats and cats with hypertensive retinopathy. Am J Vet Res 2004;65:245252.

  • 8.

    Maggio F, DeFrancesco TC, Atkins CE, et alOcular lesions associated with systemic hypertension in cats: 69 cases (1985–1998). J Am Vet Med Assoc 2000;217:695702.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 9.

    Mathur S, Syme H, Brown CA, et alEffects of the calcium channel antagonist amlodipine in cats with surgically induced hypertensive renal insufficiency. Am J Vet Res 2002;63:833839.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 10.

    Kyles AE, Gregory CR, Wooldridge JD, et alManagement of hypertension controls postoperative neurologic disorders after renal transplantation in cats. Vet Surg 1999;28:436441.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 11.

    Chetboul V, Lefebvre HP, Pinhas C, et alSpontaneous feline hypertension: clinical and echocardiographic abnormalities, and survival rate. J Vet Intern Med 2003;17:8995.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 12.

    Henik RA, Stepien RL, Bortnowski HB. Spectrum of M-mode echocardiographic abnormalities in 75 cats with systemic hypertension. J Am Anim Hosp Assoc 2004;40:359363.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 13.

    Carlos Sampedrano C, Chetboul V, Gouni V, et alSystolic and diastolic myocardial dysfunction in cats with hypertrophic cardiomyopathy or systemic hypertension. J Vet Intern Med 2006;20:11061115.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 14.

    Watanabe T, Mishina M. Effects of benazepril hydrochloride in cats with experimentally induced or spontaneously occurring chronic renal failure. J Vet Med Sci 2007;69:10151023.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 15.

    Jepson RE, Elliott J, Brodbelt D, et alEffect of control of systolic blood pressure on survival in cats with systemic hypertension. J Vet Intern Med 2007;21:402409.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 16.

    Helms SR. Treatment of feline hypertension with transdermal amlodipine: a pilot study. J Am Anim Hosp Assoc 2007;43:149156.

  • 17.

    Elliott J, Barber PJ, Syme HM, et alFeline hypertension: clinical findings and response to antihypertensive treatment in 30 cases. J Small Anim Pract 2001;42:122129.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 18.

    Snyder PS, Sadek D, Jones GL. Effect of amlodipine on echocardiographic variables in cats with systemic hypertension. J Vet Intern Med 2001;15:5256.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 19.

    Bodey AR, Michell AR, Bovee KC, et alComparison of direct and indirect (oscillometric) measurement of arterial blood pressure in conscious dogs. Res Vet Sci 1996;61:1721.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 20.

    Stepien RL, Rapoport GS. Clinical comparison of three methods to measure blood pressure in nonsedated dogs. J Am Vet Med Assoc 1999;215:16231628.

    • Search Google Scholar
    • Export Citation
  • 21.

    Chalifoux A, Dallaire A, Blais D, et alEvaluation of the arterial blood pressure of dogs by two noninvasive methods. Can J Comp Med 1985;49:419423.

    • Search Google Scholar
    • Export Citation
  • 22.

    Stepien RL, Rapoport GS, Henik RA, et alComparative diagnostic test characteristics of oscillometric and Doppler ultrasonographic methods in the detection of systolic hypertension in dogs. J Vet Intern Med 2003;17:6572.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 23.

    Jepson RE, Hartley V, Mendl M, et alA comparison of CAT Doppler and oscillometric Memoprint machines for non-invasive blood pressure measurement in conscious cats. J Feline Med Surg 2005;7:147152.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 24.

    Chetboul V, Athanassiadis N, Concordet D, et alObserver-dependent variability of quantitative clinical endpoints: the example of canine echocardiography. J Vet Pharmacol Ther 2004;27:4956.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 25.

    Bodey AR, Michell AR. Longitudinal studies of reproducibility and variability of indirect (oscillometric) blood pressure measurements in dogs: evidence for tracking. Res Vet Sci 1997;63:1521.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 26.

    Hsiang TY, Lien YH, Huang HP. Indirect measurement of systemic blood pressure in conscious dogs in a clinical setting. J Vet Med Sci 2008;70:449453.

  • 27.

    Littman MP, Fox PR. Systemic hypertension: recognition and treatment. In: Fox P, Sisson D, Moise NS, eds. Textbook of canine and feline cardiology. 2nd ed. Philadelphia: WB Saunders Co, 1988;795813.

    • Search Google Scholar
    • Export Citation
All Time Past Year Past 30 Days
Abstract Views 105 0 0
Full Text Views 1724 1219 140
PDF Downloads 359 147 10
Advertisement