In vitro flow characteristics of the Ahmed and self-constructed anterior chamber shunts

D. Todd Strubbe From the Gwathmey-Adams Laboratory for Visual Science and the Department of Small Animal Clinical Sciences, College of Veterinary Medicine, University of Florida, Gainesville, FL 32610-0126.

Search for other papers by D. Todd Strubbe in
Current site
Google Scholar
PubMed
Close
 DVM
,
Kirk N. Gelatt From the Gwathmey-Adams Laboratory for Visual Science and the Department of Small Animal Clinical Sciences, College of Veterinary Medicine, University of Florida, Gainesville, FL 32610-0126.

Search for other papers by Kirk N. Gelatt in
Current site
Google Scholar
PubMed
Close
 VMD
, and
Edward O. MacKay From the Gwathmey-Adams Laboratory for Visual Science and the Department of Small Animal Clinical Sciences, College of Veterinary Medicine, University of Florida, Gainesville, FL 32610-0126.

Search for other papers by Edward O. MacKay in
Current site
Google Scholar
PubMed
Close
Free access

SUMMARY

Objective

To compare in vitro opening pressures (OP) and closing pressures (CP) of the Ahmed VS-1 and VS-2 glaucoma valves with those of several selfconstructed valve ‘prototypes,’ and to assess their ability to maintain perfusion pressures between 6 and 21 mm of Hg.

Sample Population

Ahmed VS-1 (n = 6), 2 groups of Ahmed VS-2 (group 1: n = 12; group 2: n = 14), and self-constructed valves with linear incisions in the long axis of the tube wall (n = 6) or X-shaped incisions in the tube walls (n = 2).

Procedure

Valves were perfused with deionized water, lactated Ringer's solution (LRS), Dulbecco's modified Eagle's medium (DMEM), DMEM plus 50% equine serum (ES), and 100% ES. Flow rates of 2.85, 4.2, 6.0, 9.0, and 12.0 μl/min were used for each perfusate. Valves were tested 3 times for reproducibility, and OP/CP were compared for each system.

Results

OP/CP of the VS-1, VS-2 (group 1), VS-2 (group 2), and linear 1.0-cm incisional valves with thick tubing consistently increased with increasing perfusion rate. Linear 0.5-cm (thick tubing) and 1.0-cm (thin tubing) incisional valves had increasing OP/CP with increasing perfusion rate in all but a few instances. Mean OP/CP decreased with increasing perfusate osmolarity for all perfusates except LRS, using the VS-1 and V-2 (group 2) valves. Mean OP/CP were consistently lower for VS-1 than VS-2 (group 1) valves at any given flow rate and for any given perfusate. Mean OP/CP were consistently lower for VS-2 (group 1) than VS-2 (group 2) valves at any given flow rate and for any given perfusate. The linear 0.5-cm incisional valves with thick and thin tubing induced the highest mean OP/CP, maximizing at > 30 mm of Hg.

Conclusions

Only the VS-2 (group 2) valves consistently had mean OP/CP between 6 and 21 mm of Hg for all perfusates and at all flow rates.

Clinical Relevance

Anterior chamber shunts, although imperfect, appear to offer a physiologically sound alternative for glaucoma management. (Am J Vet Res 1997;58:1332–1337)