Abstract
Objective—To determine the degree of agreement between 3 commercially available point-of-care blood glucose meters and a laboratory analyzer for measurement of blood glucose concentrations in Hispaniolan Amazon parrots (Amazona ventralis).
Animals—20 healthy adult Hispaniolan Amazon parrots.
Procedures—A 26-gauge needle and 3-mL syringe were used to obtain a blood sample (approx 0.5 mL) from a jugular vein of each parrot. Small volumes of blood (0.6 to 1.5 μL) were used to operate each of the blood glucose meters, and the remainder was placed into lithium heparin microtubes and centrifuged. Plasma was harvested and frozen at −30°C. Within 5 days after collection, plasma samples were thawed and plasma glucose concen-trations were measured by means of the laboratory analyzer. Agreement between pairs of blood glucose meters and between each blood glucose meter and the laboratory analyzer was evaluated by means of the Bland-Altman method, and limits of agreement (LOA) were calculated.
Results—None of the results of the 3 blood glucose meters agreed with results of the laboratory analyzer. Each point-of-care blood glucose meter underestimated the blood glucose concentration, and the degree of negative bias was not consistent (meter A bias, −94.9 mg/dL [LOA, −148.0 to −41.7 mg/dL]; meter B bias, −52 mg/dL [LOA, −107.5 to 3.5 mg/dL]; and meter C bias, −78.9 mg/dL [LOA, −137.2 to −20.6 mg/dL]).
Conclusions and Clinical Relevance—On the basis of these results, use of handheld blood glucose meters in the diagnosis or treatment of Hispaniolan Amazon parrots and other psittacines cannot be recommended.
Abstract
OBJECTIVE
To measure baseline plasma corticosterone levels in Hispaniolan Amazon parrots (Amazona ventralis) and assess the effects of handling and restraint on corticosterone levels over 1 hour, reflective of what parrots might experience during veterinary care.
ANIMALS
10 male and 12 female Hispaniolan Amazon parrots.
PROCEDURES
Each parrot was removed from its cage and wrapped in a towel for restraint similar to that performed in a clinical setting. An initial baseline blood sample was collected in < 3 minutes upon entrance into the parrot room, after which blood samples were taken every 15 minutes for 1 hour (a total of 5 blood samples). An enzyme-linked immunoassay was validated for Hispaniolan Amazon parrots and used to determine concentrations of plasma corticosterone.
RESULTS
On average, parrots showed a significant increase in corticosterone between baseline samples and all subsequent postrestraint time points (average baseline corticosterone ± SD: 0.51 ± 0.65 ng/mL). Females, on average, displayed significantly higher corticosterone levels than males after 30, 45, and 60 minutes of restraint (P = .016, P = .0099, and P = .015, respectively). Birds with feather-destructive behavior did not have significantly higher corticosterone levels than birds without the condition (P = .38).
CLINICAL RELEVANCE
Understanding the physiological stress response in companion psittacine birds during routine handling will allow clinicians to better evaluate how this may affect the patient’s condition and diagnostic test results. Assessing how corticosterone correlates to behavioral conditions such as feather-destructive behavior will provide clinicians with the potential to develop treatment options.
Abstract
Objective—To determine the pharmacokinetics and safety of voriconazole administered orally in single and multiple doses in Hispaniolan Amazon parrots (Amazona ventralis).
Animals—15 clinically normal adult Hispaniolan Amazon parrots.
Procedures—Single doses of voriconazole (12 or 24 mg/kg) were administered orally to 15 and 12 birds, respectively; plasma voriconazole concentrations were determined at intervals via high-pressure liquid chromatography. In a multiple-dose trial, voriconazole (18 mg/kg) or water was administered orally to 6 and 4 birds, respectively, every 8 hours for 11 days (beginning day 0); trough plasma voriconazole concentrations were evaluated on 3 days. Birds were monitored daily, and clinicopathologic variables were evaluated before and after the trial.
Results—Voriconazole elimination half-life was short (0.70 to 1.25 hours). In the single-dose experiments, higher drug doses yielded proportional increases in the maximum plasma voriconazole concentration (Cmax) and area under the curve (AUC). In the multiple-dose trial, Cmax, AUC, and plasma concentrations at 2 and 4 hours were decreased on day 10, compared with day 0 values; however, there was relatively little change in terminal half-life. With the exception of 1 voriconazole-treated parrot that developed polyuria, adverse effects were not evident.
Conclusions and Clinical Relevance—In Hispaniolan Amazon parrots, oral administration of voriconazole was associated with proportional kinetics following administration of single doses and a decrease in plasma concentration following administration of multiple doses. Oral administration of 18 mg of voriconazole/kg every 8 hours would require adjustment to maintain therapeutic concentrations during long-term treatment. Safety and efficacy of voriconazole treatment in this species require further investigation.
Abstract
Objective—To determine plasma osmolality in healthy adult Hispaniolan Amazon parrots (Amazona ventralis) and validate osmolality equations in these parrots.
Animals—20 healthy adult Hispaniolan Amazon parrots.
Procedures—A blood sample (0.5 mL) was collected from the right jugular vein of each parrot and placed into a lithium heparin microtainer tube. Samples were centrifuged, and plasma was harvested and frozen at −30°C. Samples were thawed, and plasma osmolality was measured in duplicate with a freezing-point depression osmometer. The mean value was calculated for the 2 osmolality measurements.
Results—Plasma osmolality values were normally distributed, with a mean ± SD of 326.0 ± 6.878 mOsm/kg. The equations (2 × [Na+ + K+]) + (glucose/18), which resulted in bias of 2.3333 mOsm/kg and limits of agreement of −7.0940 to 11.7606 mOsm/kg, and (2 × [Na+ + K+]) + (uric acid concentration/16.8) + (glucose concentration/18), which resulted in bias of 5.8117 mOsm/kg and limits of agreement of −14.6640 to 3.0406 mOsm/kg, yielded calculated values that were in good agreement with the measured osmolality.
Conclusions and Clinical Relevance—IV administration of large amounts of hypotonic fluids can have catastrophic consequences. Osmolality of the plasma from parrots in this study was significantly higher than that of commercially available prepackaged fluids. Therefore, such fluids should be used with caution in Hispaniolan Amazon parrots as well as other psittacines. Additional studies are needed to determine whether the estimation of osmolality has the same clinical value in psittacines as it does in other animals.
Abstract
Objective—To establish a computed tomography (CT)-angiography protocol and measure the diameters of major arteries in parrots.
Animals—13 Hispaniolan Amazon parrots (Amazona ventralis).
Procedures—16-slice CT scanning was used to measure the apparent diameter of the ascending aorta, abdominal aorta, pulmonary arteries, and brachiocephalic trunk. Before scanning, all birds underwent ECG and echocardiographic assessment and were considered free of detectable cardiovascular diseases. Each bird was anesthetized, and a precontrast helical CT scan was performed. Peak aortic enhancement was established with a test bolus technique via dynamic axial CT scan over a predetermined single slice. An additional bolus of contrast medium was then injected, and a helical CT-angiography scan was performed immediately afterward. Arterial diameter measurements were obtained by 2 observers via various windows before and after injection, and intra- and interobserver agreement was assessed.
Results—Reference limits were determined for arterial diameter measurements before and after contrast medium administration in pulmonary, mediastinal, and manual angiography windows. Ratios of vertebral body diameter to keel length were also calculated. Intraobserver agreement was high (concordance correlation coefficients ≥ 0.95); interobserver agreement was medium to high (intraclass correlation coefficients ≥ 0.65).
Conclusions and Clinical Relevance—CT-angiography was safe and is of potential diagnostic value in parrots. We recommend performing the angiography immediately after IV injection of 3 mL of iohexol/kg. Arterial diameter measurements at the described locations were reliable.
Abstract
Objective—To assess the effects of dopamine and dobutamine on the blood pressure of isoflurane-anesthetized Hispaniolan Amazon parrots (Amazona ventralis).
Animals—8 Hispaniolan Amazon parrots.
Procedures—A randomized crossover study was conducted. Each bird was anesthetized (anesthesia maintained by administration of 2.5% isoflurane in oxygen) and received 3 doses of each drug during a treatment period of 20 min/dose. Treatments were constant rate infusions (CRIs) of dobutamine (5, 10, and 15 μg/kg/min) and dopamine (5, 7, and 10 μg/kg/min). Direct systolic, diastolic, and mean arterial pressure measurements, heart rate, esophageal temperature, and end-tidal partial pressure of CO2 were recorded throughout the treatment periods.
Results—Mean ± SD of the systolic, mean, and diastolic arterial blood pressures at time 0 (initiation of a CRI) were 132.9 ± 22.1 mm Hg, 116.9 ± 20.5 mm Hg, and 101.9 ± 22.0 mm Hg, respectively. Dopamine resulted in significantly higher values than did dobutamine for the measured variables, except for end-tidal partial pressure of CO2. Post hoc multiple comparisons revealed that the changes in arterial blood pressure were significantly different 4 to 7 minutes after initiation of a CRI. Overall, dopamine at rates of 7 and 10 μg/kg/min and dobutamine at a rate of 15 μg/kg/min caused the greatest increases in arterial blood pressure.
Conclusions and Clinical Relevance—Dobutamine CRI at 5, 10, and 15 μg/kg/min and dopamine CRI at 5, 7, and 10 μg/kg/min may be useful in correcting severe hypotension in Hispaniolan Amazon parrots caused by anesthesia maintained with 2.5% isoflurane.
Abstract
Objective—To determine phenol red thread test (PRTT) values in eyes of clinically normal Hispaniolan Amazon parrots before and after topical application of an ophthalmic anesthetic agent and compare findings with Schirmer tear test (STT) values.
Design—Evaluation study.
Animals—24 Amazona ventralis parrots from a research colony.
Procedures—On 4 occasions (1-week intervals), all birds underwent a thorough ophthalmic examination of both eyes, which included (in sequence) performance of a PRTT and an STT; topical ocular application of proparacaine hydrochloride; and performance of another PRTT and another STT. Correlations between PRTT and STT values recorded with and without topical anesthesia were assessed.
Results—Without topical anesthesia, mean ± SD PRTT value was 12.5 ± 5.0 mm/15 s (range, 1 to 25 mm/15 s). With topical anesthesia, the PRTT value was 12.6 ± 5.4 mm/15 s (range, 2 to 24 mm/15 s). Without topical anesthesia, mean STT value was 7.9 ± 2.6 mm/min (range, 0 to 13 mm/min). With topical anesthesia, the STT value was 5.1 ± 3.3 mm/min (range, 0 to 18 mm/min). The correlation of PRTT and STT values recorded with or without topical anesthesia was weak (r = 0.51 and r = 0.32, respectively).
Conclusions and Clinical Relevance—Results indicated that the PRTT and STT were both viable methods for measurement of tear production in Hispaniolan Amazon parrots. Topical application of an ophthalmic anesthetic agent did not have a significant effect on the PRTT values but significantly decreased the STT values.
Abstract
Objective—To determine the level of agreement between direct and indirect blood pressure measurements obtained from healthy Hispaniolan Amazon parrots (Amazona ventralis) anesthetized with isoflurane.
Design—Validation study.
Animals—16 healthy adult Hispaniolan Amazon parrots.
Procedures—Parrots were anesthetized, and a 26-gauge, 19-mm catheter was placed percutaneously in the superficial ulnar artery for direct measurement of systolic, mean, and diastolic arterial pressures. Indirect blood pressure measurements were obtained with a Doppler ultrasonic flow detector and an oscillometric unit. The Bland-Altman method was used to compare direct and indirect blood pressure values.
Results—There was substantial disagreement between direct systolic arterial blood pressure and indirect blood pressure measurements obtained with the Doppler detector from the wing (bias, 24 mm Hg; limits of agreement, −37 to 85 mm Hg) and from the leg (bias, 14 mm Hg; limits of agreement, −14 to 42 mm Hg). Attempts to obtain indirect blood pressure measurements with the oscillometric unit were unsuccessful.
Conclusions and Clinical Relevance—Results suggested that there was substantial disagreement between indirect blood pressure measurements obtained with a Doppler ultrasonic flow detector in anesthetized Hispaniolan Amazon parrots and directly measured systolic arterial blood pressure.