Search Results

You are looking at 1 - 2 of 2 items for

  • Author or Editor: Danielle E. Strahl-Heldreth x
  • Refine by Access: All Content x
Clear All Modify Search

Abstract

OBJECTIVE: To determine the intracoelemic (ICe) dose of alfaxalone required to induce loss of righting reflex (LRR) in garter snakes (Thamnophis sirtalis) and to evaluate the tactile stimulus response in unanesthetized and alfaxalone-anesthetized snakes.

ANIMALS: 8 healthy mature garter snakes.

PROCEDURES: During the first of 3 phases, snakes received each of 3 doses (10, 20, and 30 mg/kg) of alfaxalone, ICe, with a 2-week washout period between treatments. Times to LRR and return of righting reflex were determined after each dose. During phase 2, unanesthetized snakes underwent tactile stimulation testing with Semmes-Weinstein monofilaments once daily for 3 consecutive days to determine the baseline tactile pressure required to elicit purposeful movement. During phase 3, snakes were anesthetized with alfaxalone (30 mg/kg, ICe), and the tactile pressure required to induce purposeful movement was assessed at predetermined times after LRR.

RESULTS: Intracoelomic administration of alfaxalone at doses of 10, 20, and 30 mg/kg induced LRR in 0, 5, and 8 snakes, respectively. For snakes with LRR, median time to LRR following the 30-mg/kg dose (3.8 minutes) was significantly shorter than that following the 20-mg/kg dose (8.3 minutes); median time to return of righting reflex did not differ between the 2 doses. Mean ± SD tactile pressure that resulted in purposeful movement in unanesthetized snakes was 16.9 ± 14.3 g. When snakes were anesthetized, the mean tactile pressure that resulted in purposeful movement was significantly increased from baseline at 10, 20, and 30 minutes after LRR.

CONCLUSIONS AND CLINICAL RELEVANCE: Results suggested ICe administration of alfaxalone might be effective for anesthetizing garter snakes.

Full access
in American Journal of Veterinary Research

Abstract

OBJECTIVE

To evaluate the repeatability and accuracy of fingertip pulse oximeters (FPO) for measurement of hemoglobin oxygen saturation in arterial blood and pulse rate (PR) in anesthetized dogs breathing 100% O2.

ANIMALS

29 healthy client-owned anesthetized dogs undergoing various surgical procedures.

PROCEDURES

In randomized order, each of 7 FPOs or a reference pulse oximeter (PO) was applied to the tongue of each intubated anesthetized dog breathing 100% O2. Duplicate measurements of oxygen saturation (Spo 2) and PR were obtained within 60 seconds of applying an FPO or PO. A nonparametric version of Bland-Altman analysis was used. Coefficient of repeatability was the interval between the 5th and 95th percentiles of the differences between duplicate measurements. Bias was the median difference, and the limits of agreement were the 5th and 95th percentiles of the differences between each FPO and the PO. Acceptable values for the coefficient of repeatability of Spo 2 were ≤ 6%. Agreements were accepted if the limits of agreement had an absolute difference of ≤ ± 3% in Spo 2 and relative difference of ≤ ± 10% in PR.

RESULTS

Coefficient of repeatability for Spo 2 was acceptable for 5 FPOs, but the limits of agreement for Spo 2 were unacceptable for all FPOs. The limits of agreement for PR were acceptable for 2 FPOs.

CONCLUSIONS AND CLINICAL RELEVANCE

Results suggested that some FPOs may be suitable for accurately monitoring PRs of healthy anesthetized dogs breathing 100% O2, but mild underestimation of Spo 2 was common.