Search Results
You are looking at 1 - 1 of 1 items for
- Author or Editor: Miyoung Lee x
- Refine by Access: All Content x
Abstract
Objective—To compare echocardiographic measurements of left ventricular (LV) volume obtained via a modified Simpson or Teichholz method with those obtained via dual-source CT (DSCT).
Animals—7 healthy Beagles.
Procedures—Each dog was anesthetized for DSCT; LV volume was determined from contrast-enhanced images of the LV lumen during all phases of contraction. Echocardiography was performed with dogs awake and anesthetized. End-diastolic volume (EDV), end-systolic volume (ESV), stroke volume, and ejection fraction were measured via a modified Simpson method and Teichholz method. Each dog was anesthetized twice with a 1-week interval between anesthetic sessions.
Results—Results obtained while dogs were anesthetized revealed that the modified Simpson method underestimated LV volume (mean ± SD EDV, 24.82 ± 2.38 mL; ESV, 12.24 ± 1.77 mL), compared with that estimated by the Teichholz method (EDV, 32.57 ± 2.85 mL; ESV, 14.87 ± 2.09 mL) or DSCT (EDV, 34.14 ± 1.57 mL; ESV, 16.71 ± 0.76 mL). Ejection fraction (modified Simpson method, 48.53% ± 4.24%; Teichholz method, 54.33% ± 4.26%; DSCT, 51.00% ± 2.71%) differed significantly among the 3 methods. Echocardiographic results obtained while dogs were awake revealed that EDV, ESV, and stroke volume differed significantly between the modified Simpson and Teichholz methods.
Conclusions and Clinical Relevance—LV volume determined via the Teichholz method was more similar to that determined via DSCT than was the LV volume determined via the modified Simpson method. The modified Simpson method underestimated LV volume, compared with that obtained via the Teichholz method, in both anesthetized and awake dogs.