Abstract
OBJECTIVE To compare blood glucose concentrations of black-tailed prairie dogs (Cynomys ludovicianus) measured by use of a variety of portable analyzers with results from a laboratory biochemistry analyzer.
SAMPLE Venous blood samples (3 mL) obtained from each of 16 healthy black-tailed prairie dogs.
PROCEDURES A portion of each blood sample was used to measure glucose concentrations by use of an amperometric human point-of-care glucometer and a colorimetric species-specific portable blood glucose meter designed for veterinary use with both canine (code 5) and feline (code 7) settings. The remainder of each blood sample was placed into 2 tubes (one contained lithium heparin and the other contained no anticoagulant). A portable veterinary chemistry analyzer (PVCA) and a handheld analyzer were used to measure glucose concentration in heparinized blood. Serum glucose concentration was measured in the remaining portion by use of a biochemistry analyzer. A general linear mixed models approach was used to compare glucose concentrations and measurement bias obtained with the various measurement methods.
RESULTS Measurement bias and differences in mean glucose concentrations were apparent with all measurement methods. In particular, the veterinary glucometer, whether used on the canine or feline setting, overestimated mean glucose concentrations, whereas the human glucometer, PVCA, and handheld analyzer underestimated mean glucose concentrations relative to the concentration obtained with the biochemistry analyzer.
CONCLUSIONS AND CLINICAL RELEVANCE Results indicated that none of the measurement methods provided consistently accurate blood glucose concentrations of black-tailed prairie dogs, compared with values determined with a biochemistry analyzer.
Abstract
OBJECTIVE To investigate effects of storage duration and temperature on biochemical analytes in plasma from red-eared sliders (Trachemys scripta elegans).
ANIMALS 8 red-eared sliders.
PROCEDURES Blood samples were collected. Plasma was harvested and analyzed at room temperature (approx 23°C; time = 1 hour) and then fractioned into 0.1-mL aliquots that were stored at room temperature or were refrigerated (4°C) or frozen (−20°C). Biochemical analysis of stored samples was performed at 4 (room temperature), 8 (4°C), 24 (4°C), 48 (4° and −20°C), and 72 (−20°C) hours and at 7 days (−20°C). For each time point for each storage temperature, bias was calculated by subtracting values from the value obtained at 1 hour. Bias was modeled by use of a linear mixed model.
RESULTS Storage temperature had a significant effect on several plasma biochemical analytes. In general, aspartate aminotransferase activity and uric acid, total protein, and potassium concentrations increased after storage at 4° and −20°C. Differences in values after storage were mostly within the acceptable range for allowable total error, except for calcium and potassium concentrations for samples stored at −20°C. Both storage temperatures increased variability of measurement results. Results for samples stored at room temperature for 4 hours did not differ significantly from values at 1 hour. Results differed significantly between refrigerated and frozen samples stored for 48 hours.
CONCLUSIONS AND CLINICAL RELEVANCE Short-term storage conditions influenced results for some biochemical analytes. These effects should be considered when performing biochemical analyses of plasma samples obtained from red-eared sliders.