Abstract
OBJECTIVE
To validate the use of a flow cytometric assay that uses 2‘,7‘-dichlorodihydrofluorescein diacetate (DCFH-DA) to measure reactive oxygen species in the erythrocytes of healthy dogs.
ANIMALS
50 healthy adult dogs.
PROCEDURES
Erythrocytes were incubated with DCFH-DA or a vehicle control (dimethyl sulfoxide), then incubated with (stimulated) or without (unstimulated) hydrogen peroxide. The flow cytometric assay was evaluated for specificity with increasing concentrations of DCFH-DA and hydrogen peroxide, and a polynomial regression line was applied to determine optimal concentrations. For precision, samples were analyzed 5 consecutive times for determination of intra- and interassay variability. Stability of samples stored at 4°C for up to 48 hours after blood collection was determined with flow cytometric analysis. Coefficient of variation (CV) was considered acceptable at 20%. Baseline measurements were used to determine an expected range of median fluorescence intensity for unstimulated erythrocytes incubated with DCFH-DA.
RESULTS
Erythrocytes were successfully isolated, and stimulated samples demonstrated higher median fluorescence intensity, compared with unstimulated samples. The intra-assay CV was 11.9% and 8.9% and interassay CV was 11.9% and 9.1% for unstimulated and stimulated samples, respectively. Unstimulated samples were stable for up to 24 hours, whereas stimulated samples were stable for up to 48 hours.
CONCLUSIONS AND CLINICAL RELEVANCE
Flow cytometry for the measurement of reactive oxygen species in the erythrocytes of healthy dogs by use of DCFH-DA had acceptable specificity, precision, and stability. Flow cytometry is a promising technique for evaluating intraerythrocytic oxidative stress for healthy dogs.