• 1. Alpha TRAK 2 [package insert]. Abbott Park, Ill: Abbott Laboratories, 2012.

  • 2. Coldman MF, Good W. The distribution of sodium, potassium and glucose in the blood of some mammals. Comp Biochem Physiol 1967;21:201206.

    • Search Google Scholar
    • Export Citation
  • 3. Mackay EM. The distribution of glucose in human blood. J Biol Chem 1932;97:685689.

  • 4. Johnson BM, Fry MM, Flatland B, et al. Comparison of a human portable blood glucose meter, veterinary portable blood glucose meter, and automated chemistry analyzer for measurement of blood glucose concentrations in dogs. J Am Vet Med Assoc 2009;235:13091313.

    • Search Google Scholar
    • Export Citation
  • 5. Paul AE, Shiel RE, Juvet F, et al. Effect of hematocrit on accuracy of two point-of-care glucometers for use in dogs. Am J Vet Res 2011;72:12041208.

    • Search Google Scholar
    • Export Citation
  • 6. Solnica B, Skupien J, Kusnierz-Cabala B, et al. The effect of hematocrit on the results of measurements using glucose meters based on different techniques. Clin Chem Lab Med 2011;50:361365.

    • Search Google Scholar
    • Export Citation
  • 7. Jamaluddin FA, Gunavathy M, Yean CY, et al. Variability of point-of-care testing blood glucometers versus the laboratory reference method in a tertiary teaching hospital. Asian Biomed 2012;6:6774.

    • Search Google Scholar
    • Export Citation
  • 8. Tang Z, Lee JH, Louie RF, et al. Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing. Arch Pathol Lab Med 2000;124:11351140.

    • Search Google Scholar
    • Export Citation
  • 9. Barreau PB, Buttery JE. The effect of the haemoatocrit value on the determination of glucose levels by the reagent-strip methods. Med J Aust 1987;147:286288.

    • Search Google Scholar
    • Export Citation
  • 10. Pidcoke HF, Wade CE, Mann EA, et al. Anemia causes hypoglycemia in intensive care unit patients due to error in single-channel glucometers: methods of reducing patient risk. Crit Care Med 2010;38:471476.

    • Search Google Scholar
    • Export Citation
  • 11. Hussain K, Sharief N. The inaccuracy of venous and capillary blood glucose measurement using reagent strips in the newborn period and the effect of haematocrit. Early Hum Dev 2000;57:111121.

    • Search Google Scholar
    • Export Citation
  • 12. Krinsley JS, Grover A. Severe hypoglycemia in critically ill patients: risk factors and outcomes. Crit Care Med 2007;35:22622267.

  • 13. Ginsberg BH. Factors affecting blood glucose monitoring: sources of errors in measurement. J Diabetes Sci Technol 2009;3:903913.

  • 14. Lane SL, Koenig A, Brainard BM. Formulation and validation of a predictive model to correct blood glucose concentrations obtained with a veterinary point-of-care glucometer in hemodiluted and hemoconcentrated canine blood samples. J Am Vet Med Assoc 2015;246:307312.

    • Search Google Scholar
    • Export Citation
  • 15. Tonyushkina K, Nichols JH. Glucose meters: a review of technical challenges to obtaining accurate results. J Diabetes Sci Technol 2009;3:971980.

    • Search Google Scholar
    • Export Citation
  • 16. Dungan K, Chapman J, Braithwaite SS, et al. Glucose measurement: confounding issues in setting targets for inpatient management. Diabetes Care 2007;30:403409.

    • Search Google Scholar
    • Export Citation
  • 17. Tang Z, Du X, Louie RF, et al. Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer. Am J Clin Pathol 2000;113:7586.

    • Search Google Scholar
    • Export Citation
  • 18. Karon BS, Griesmann L, Scott R, et al. Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters. Diabetes Technol Ther 2008;10:111120.

    • Search Google Scholar
    • Export Citation
  • 19. Louie RF, Tang Z, Sutton DV, et al. Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design. Arch Pathol Lab Med 2000;124:257266.

    • Search Google Scholar
    • Export Citation
  • 20. Rebel A, Rice MA, Fahey BG. The accuracy of point-of-care glucose measurements. J Diabetes Sci Technol 2012;6:396411.

  • 21. D'Orazio P, Burnett RW, Fogh-Andersen N, et al. Approved IFCC recommendation on reporting results for blood glucose: International Federation of Clinical Chemistry and Laboratory Medicine Scientific Division, Working Group on Selective Electrodes and Point-of-Care Testing (IFCC-SD-WGSEPOCT). Clin Chem Lab Med 2006;44:14861490.

    • Search Google Scholar
    • Export Citation
  • 22. Atkin SH, Dasmahapatra A, Jaker MA, et al. Fingerstick glucose determination in shock. Ann Intern Med 1991;114:10201024.

  • 23. Mann EA, Salinas J, Pidcoke HF, et al. Error rates resulting from anemia can be corrected in multiple commonly used point-of-care glucometers. J Trauma 2008;64:1521.

    • Search Google Scholar
    • Export Citation
  • 24. Girouard J, Forest JC, Massé J, et al. Multicenter evaluation of the Glucometer Elite XL meter, an instrument specifically designed for use with neonates. Diabetes Care 2000;23:11491153.

    • Search Google Scholar
    • Export Citation

Advertisement

Development and evaluation of a formula to correct blood glucose concentration measurements in hemodiluted and hemoconcentrated feline blood samples tested by use of a veterinary point-of-care glucometer

View More View Less
  • 1 1Department of Small Animal Medicine and Surgery, College of Veterinary Medicine, University of Georgia, Athens, GA 30602.

Abstract

OBJECTIVE

To determine the effect of PCV on blood glucose concentration measurements in feline blood samples tested with a point-of-care (POC) glucometer and to develop and evaluate a correction formula that adjusts POC glucometer–measured blood glucose concentration (POCgluc) for a given PCV.

DESIGN

Experimental and prospective study.

SAMPLE

Blood samples from 4 healthy and 16 hospitalized cats.

PROCEDURES

Heparinized blood samples from healthy cats were processed into packed RBCs and plasma. Packed RBCs were resuspended with plasma to achieve PCVs ranging from 0% to 87%. Duplicate PCV and POCgluc measurements were obtained for each suspension. Plasma glucose concentration as measured by a clinical laboratory biochemical analyzer (LABgluc) was assessed; results were compared with the POCgluc. A formula to correct POCgluc measurements for PCV was developed. Blood samples from hospitalized cats were used to evaluate the formula.

RESULTS

For each healthy cat, LABgluc values were similar for all PCVs; the mean difference between POCgluc and LABgluc at PCVs outside a range of 35% to 55% was significant. Mean differences between POCgluc and LABgluc were 24.3 and 41.5 mg/dL, whereas mean differences between corrected POCgluc and LABgluc were 3 and 25.9 mg/dL for samples from healthy and hospitalized cats, respectively. Correlation between corrected POCgluc and LABgluc was stronger than that between POCgluc and LABgluc for samples from healthy and hospitalized cats.

CONCLUSIONS AND CLINICAL RELEVANCE

The POCgluc did not reflect LABgluc in hemodiluted or hemoconcentrated feline blood samples. Use of a correction formula appeared to reduce this error. Additional studies are needed to evaluate the frequency with which this correction formula might prevent clinical errors. (J Am Vet Med Assoc 2019;254:1180–1185)

Abstract

OBJECTIVE

To determine the effect of PCV on blood glucose concentration measurements in feline blood samples tested with a point-of-care (POC) glucometer and to develop and evaluate a correction formula that adjusts POC glucometer–measured blood glucose concentration (POCgluc) for a given PCV.

DESIGN

Experimental and prospective study.

SAMPLE

Blood samples from 4 healthy and 16 hospitalized cats.

PROCEDURES

Heparinized blood samples from healthy cats were processed into packed RBCs and plasma. Packed RBCs were resuspended with plasma to achieve PCVs ranging from 0% to 87%. Duplicate PCV and POCgluc measurements were obtained for each suspension. Plasma glucose concentration as measured by a clinical laboratory biochemical analyzer (LABgluc) was assessed; results were compared with the POCgluc. A formula to correct POCgluc measurements for PCV was developed. Blood samples from hospitalized cats were used to evaluate the formula.

RESULTS

For each healthy cat, LABgluc values were similar for all PCVs; the mean difference between POCgluc and LABgluc at PCVs outside a range of 35% to 55% was significant. Mean differences between POCgluc and LABgluc were 24.3 and 41.5 mg/dL, whereas mean differences between corrected POCgluc and LABgluc were 3 and 25.9 mg/dL for samples from healthy and hospitalized cats, respectively. Correlation between corrected POCgluc and LABgluc was stronger than that between POCgluc and LABgluc for samples from healthy and hospitalized cats.

CONCLUSIONS AND CLINICAL RELEVANCE

The POCgluc did not reflect LABgluc in hemodiluted or hemoconcentrated feline blood samples. Use of a correction formula appeared to reduce this error. Additional studies are needed to evaluate the frequency with which this correction formula might prevent clinical errors. (J Am Vet Med Assoc 2019;254:1180–1185)

Contributor Notes