Agreement between directly measured blood pressure and pressures obtained with three veterinary-specific oscillometric units in cats

Mark J. Acierno Department of Veterinary Clinical Science, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA 70803.

Search for other papers by Mark J. Acierno in
Current site
Google Scholar
PubMed
Close
 MBA, DVM, DACVIM
,
Diana Seaton Department of Veterinary Clinical Science, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA 70803.

Search for other papers by Diana Seaton in
Current site
Google Scholar
PubMed
Close
 DVM
,
Mark A. Mitchell Department of Veterinary Clinical Medicine, College of Veterinary Medicine, University of Illinois, Urbana, IL 61802.

Search for other papers by Mark A. Mitchell in
Current site
Google Scholar
PubMed
Close
 DVM, PhD
, and
Anderson da Cunha Department of Veterinary Clinical Science, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA 70803.

Search for other papers by Anderson da Cunha in
Current site
Google Scholar
PubMed
Close
 DVM, MS, DACVA

Abstract

Objective—To determine whether veterinary-specific oscillometric blood pressure units yield measurements that are in good agreement with directly measured blood pressures in cats.

Design—Evaluation study.

Animals—21 cats undergoing routine spaying or neutering.

Procedures—A 24-gauge catheter was inserted in a dorsal pedal artery, and systolic, diastolic, and mean arterial pressures were directly measured with a validated pressure measurement system. Values were compared with indirect blood pressure measurements obtained with 3 veterinary-specific oscillometric blood pressure units.

Results—There was poor agreement between indirectly and directly measured blood pressures. For unit 1, bias between indirectly and directly measured values was −14.9 mm Hg (95% limits of agreement [LOA], −52.2 to 22.4 mm Hg), 4.4 mm Hg (95% LOA, −26.0 to 34.8 mm Hg), and −1.3 mm Hg (95% LOA, −26.7 to 24.1 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 2, bias was −10.3 mm Hg (95% LOA, −52.9 to 32.2 mm Hg), 13.0 mm Hg (95% LOA, −32.1 to 58.0 mm Hg), and 9.1 mm Hg (95% LOA, −32.9 to 51.2 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 3, bias was −13.4 mm Hg (95% LOA, −51.8 to 25.1 mm Hg), 8.0 mm Hg (95% LOA, −25.5 to 41.6 mm Hg), and −3.6 mm Hg (95% LOA, −31.6 to 24.5 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively.

Conclusions and Clinical Relevance—Results suggested that none of the 3 veterinary-specific oscillometric blood pressure units could be recommended for indirect measurement of blood pressure in cats.

Abstract

Objective—To determine whether veterinary-specific oscillometric blood pressure units yield measurements that are in good agreement with directly measured blood pressures in cats.

Design—Evaluation study.

Animals—21 cats undergoing routine spaying or neutering.

Procedures—A 24-gauge catheter was inserted in a dorsal pedal artery, and systolic, diastolic, and mean arterial pressures were directly measured with a validated pressure measurement system. Values were compared with indirect blood pressure measurements obtained with 3 veterinary-specific oscillometric blood pressure units.

Results—There was poor agreement between indirectly and directly measured blood pressures. For unit 1, bias between indirectly and directly measured values was −14.9 mm Hg (95% limits of agreement [LOA], −52.2 to 22.4 mm Hg), 4.4 mm Hg (95% LOA, −26.0 to 34.8 mm Hg), and −1.3 mm Hg (95% LOA, −26.7 to 24.1 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 2, bias was −10.3 mm Hg (95% LOA, −52.9 to 32.2 mm Hg), 13.0 mm Hg (95% LOA, −32.1 to 58.0 mm Hg), and 9.1 mm Hg (95% LOA, −32.9 to 51.2 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 3, bias was −13.4 mm Hg (95% LOA, −51.8 to 25.1 mm Hg), 8.0 mm Hg (95% LOA, −25.5 to 41.6 mm Hg), and −3.6 mm Hg (95% LOA, −31.6 to 24.5 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively.

Conclusions and Clinical Relevance—Results suggested that none of the 3 veterinary-specific oscillometric blood pressure units could be recommended for indirect measurement of blood pressure in cats.