Use of vascular access ports in femoral veins of dogs and cats with cancer

Alane Kosanovich Cahalane Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853.

Search for other papers by Alane Kosanovich Cahalane in
Current site
Google Scholar
PubMed
Close
 DVM
,
Kenneth M. Rassnick Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853.

Search for other papers by Kenneth M. Rassnick in
Current site
Google Scholar
PubMed
Close
 DVM, DACVIM
, and
James A. Flanders Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853.

Search for other papers by James A. Flanders in
Current site
Google Scholar
PubMed
Close
 DVM, DACVS

Abstract

Objective—To evaluate long-term function of vascular access ports (VAPs) implanted in the femoral vein of dogs and cats undergoing cancer treatment.

Design—Prospective clinical study.

Animals—3 dogs and 6 cats treated via chemotherapy or radiation.

Procedures—VAPs were surgically implanted in the left femoral vein of 3 dogs and 6 cats over a 1-year period. Injection port location was alternated to either a caudal thoracic or ilial location in each patient. Duration of VAP function, ease of infusion, and ease of aspiration through the VAPs were recorded, and associated complications were assessed at each VAP use. Client satisfaction with VAP placement was evaluated by use of a questionnaire.

Results—Primary uses of the VAPs included blood sampling and delivering sedative or chemotherapeutic drugs. Median duration of successful infusion was 147 days (range, 60 to 370 days), and median duration of successful aspiration was 117 days (range, 10 to 271 days). The frequency of signs of VAP-related discomfort was low (7% of patient observations). Clients were satisfied with their decision to use VAPs. Complications included partial (n = 7) or complete (2) VAP occlusion, port migration (1), and presumptive infection (1).

Conclusions and Clinical Relevance—Results suggested that VAP implantation into the femoral vein provides an acceptable means of chronic venous access in dogs and cats undergoing cancer treatment.

Abstract

Objective—To evaluate long-term function of vascular access ports (VAPs) implanted in the femoral vein of dogs and cats undergoing cancer treatment.

Design—Prospective clinical study.

Animals—3 dogs and 6 cats treated via chemotherapy or radiation.

Procedures—VAPs were surgically implanted in the left femoral vein of 3 dogs and 6 cats over a 1-year period. Injection port location was alternated to either a caudal thoracic or ilial location in each patient. Duration of VAP function, ease of infusion, and ease of aspiration through the VAPs were recorded, and associated complications were assessed at each VAP use. Client satisfaction with VAP placement was evaluated by use of a questionnaire.

Results—Primary uses of the VAPs included blood sampling and delivering sedative or chemotherapeutic drugs. Median duration of successful infusion was 147 days (range, 60 to 370 days), and median duration of successful aspiration was 117 days (range, 10 to 271 days). The frequency of signs of VAP-related discomfort was low (7% of patient observations). Clients were satisfied with their decision to use VAPs. Complications included partial (n = 7) or complete (2) VAP occlusion, port migration (1), and presumptive infection (1).

Conclusions and Clinical Relevance—Results suggested that VAP implantation into the femoral vein provides an acceptable means of chronic venous access in dogs and cats undergoing cancer treatment.

All Time Past Year Past 30 Days
Abstract Views 263 0 0
Full Text Views 1634 1271 74
PDF Downloads 459 242 22