Assessment of a commercially available point-of-care assay for the measurement of bovine cardiac troponin I concentration

Brandon C. Fraser Department of Clinical Sciences, College of Veterinary Medicine, Kansas State University, Manhattan, KS 66506.

Search for other papers by Brandon C. Fraser in
Current site
Google Scholar
PubMed
Close
 DVM
,
David E. Anderson Ruminant Models Laboratory, College of Veterinary Medicine, Kansas State University, Manhattan, KS 66506.

Search for other papers by David E. Anderson in
Current site
Google Scholar
PubMed
Close
 DVM, MS
,
Brad J. White Department of Clinical Sciences, College of Veterinary Medicine, Kansas State University, Manhattan, KS 66506.

Search for other papers by Brad J. White in
Current site
Google Scholar
PubMed
Close
 DVM, MS
,
Matt D. Miesner Department of Clinical Sciences, College of Veterinary Medicine, Kansas State University, Manhattan, KS 66506.

Search for other papers by Matt D. Miesner in
Current site
Google Scholar
PubMed
Close
 DVM, MS
,
Carrie Wheeler Department of Clinical Sciences, College of Veterinary Medicine, Kansas State University, Manhattan, KS 66506.

Search for other papers by Carrie Wheeler in
Current site
Google Scholar
PubMed
Close
 DVM
,
David Amrine Department of Clinical Sciences, College of Veterinary Medicine, Kansas State University, Manhattan, KS 66506.

Search for other papers by David Amrine in
Current site
Google Scholar
PubMed
Close
 DVM
,
Jeff Lakritz Department of Clinical Sciences, College of Veterinary Medicine, The Ohio State University, Columbus, OH 43210.

Search for other papers by Jeff Lakritz in
Current site
Google Scholar
PubMed
Close
 DVM, PhD
, and
Tom Overbay Abaxis Inc, North American Animal Health, 3240 Whipple Rd, Union City, CA 94587.

Search for other papers by Tom Overbay in
Current site
Google Scholar
PubMed
Close
 DVM

Abstract

Objective—To assess a commercially available point-of-care assay for measurement of bovine cardiac troponin I (cTnI) concentration in blood and plasma samples.

Sample—Prepared bovine plasma standard samples with known concentrations (0 to 1.0 ng/mL) of cTnI and blood and plasma samples obtained from 28 healthy 2.5-month-old Holstein calves.

Procedures—Coefficients of variation were calculated for concentrations of cTnI in prepared standards determined with the point-of-care assay, and values were compared with the known concentrations. The cTnI concentrations in blood samples obtained from calves determined with the point-of-care assay were compared with cTnI concentrations in plasma samples obtained from those animals determined with a validated immunoassay.

Results—The coefficients of variation of cTnI concentrations determined for prepared standards by use of the point-of-care assay were low (< 20%) for standards with cTnI concentrations ≥ 0.025 ng/mL. The blood cTnI concentrations determined with the point-of-care assay were not significantly different from the plasma cTnI concentrations determined with the validated immunoassay.

Conclusions and Clinical Relevance—Results of this study indicated the point-of-care assay had high precision for determination of cTnI concentrations in most evaluated prepared bovine plasma standard samples. The point-of-care assay may be useful for determination of circulating concentrations of cTnI in cattle.

Abstract

Objective—To assess a commercially available point-of-care assay for measurement of bovine cardiac troponin I (cTnI) concentration in blood and plasma samples.

Sample—Prepared bovine plasma standard samples with known concentrations (0 to 1.0 ng/mL) of cTnI and blood and plasma samples obtained from 28 healthy 2.5-month-old Holstein calves.

Procedures—Coefficients of variation were calculated for concentrations of cTnI in prepared standards determined with the point-of-care assay, and values were compared with the known concentrations. The cTnI concentrations in blood samples obtained from calves determined with the point-of-care assay were compared with cTnI concentrations in plasma samples obtained from those animals determined with a validated immunoassay.

Results—The coefficients of variation of cTnI concentrations determined for prepared standards by use of the point-of-care assay were low (< 20%) for standards with cTnI concentrations ≥ 0.025 ng/mL. The blood cTnI concentrations determined with the point-of-care assay were not significantly different from the plasma cTnI concentrations determined with the validated immunoassay.

Conclusions and Clinical Relevance—Results of this study indicated the point-of-care assay had high precision for determination of cTnI concentrations in most evaluated prepared bovine plasma standard samples. The point-of-care assay may be useful for determination of circulating concentrations of cTnI in cattle.