Effects of sterilization with hydrogen peroxide gas plasma, ethylene oxide, and steam on bioadhesive properties of nylon and polyethylene lines used for stabilization of canine stifle joints

Matthieu Gatineau Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Montreal, St-Hyacinthe, QC J2S 2M2, Canada.

Search for other papers by Matthieu Gatineau in
Current site
Google Scholar
PubMed
Close
 DMV
,
Alexander O El-Warrak Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Montreal, St-Hyacinthe, QC J2S 2M2, Canada.

Search for other papers by Alexander O El-Warrak in
Current site
Google Scholar
PubMed
Close
 DMV, PhD
,
Christian Bolliger Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Montreal, St-Hyacinthe, QC J2S 2M2, Canada.

Search for other papers by Christian Bolliger in
Current site
Google Scholar
PubMed
Close
 Dr med vet
,
Michael Mourez Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Montreal, St-Hyacinthe, QC J2S 2M2, Canada.

Search for other papers by Michael Mourez in
Current site
Google Scholar
PubMed
Close
 PhD
, and
Frederic Berthiaume Department of Clinical Sciences, Faculty of Veterinary Medicine, University of Montreal, St-Hyacinthe, QC J2S 2M2, Canada.

Search for other papers by Frederic Berthiaume in
Current site
Google Scholar
PubMed
Close
 MSc

Abstract

Objective—To compare effects of sterilization with hydrogen peroxide gas plasma (HPGP), ethylene oxide, and steam on bioadhesive properties of nylon and polyethylene lines used for stabilization of canine stifle joints.

Sample—Samples of a 36.3-kg test nylon leader line, 57.8-kg test nylon fishing line, and 2-mm ultrahigh–molecular weight polyethylene (UHMWPE) were used.

Procedures—In this in vitro study, samples of nylon leader line, fishing line, and UHMWPE sterilized by use of HPGP, ethylene oxide, and steam or unsterilized samples were used. Bacterial adherence on unsterilized and sterilized samples was tested with Staphylococcus epidermidis and Escherichia coli. Five samples were examined for each line type and sterilization condition, and final colony counts were obtained.

Results—Bacterial adherence was significantly affected by method of sterilization for all 3 line types. For most of the samples, bacterial adherence was similar or lower when HPGP sterilization was used, compared with results for sterilization via ethylene oxide and steam, respectively. Bacterial adherence was significantly higher for UHMWPE, compared with adherence for the nylon line, regardless of the sterilization method used. Bacterial adherence was higher for nylon fishing line than for nylon leader line for S epidermidis after ethylene oxide sterilization and for E coli after HPGP and ethylene oxide sterilization.

Conclusions and Clinical Relevance—Effects of HPGP sterilization on bioadhesive properties of nylon and polyethylene lines compared favorably with those for ethylene oxide and steam sterilization. Also, nylon line may be a more suitable material than UHMWPE for suture prostheses on the basis of bacterial adherence properties.

Abstract

Objective—To compare effects of sterilization with hydrogen peroxide gas plasma (HPGP), ethylene oxide, and steam on bioadhesive properties of nylon and polyethylene lines used for stabilization of canine stifle joints.

Sample—Samples of a 36.3-kg test nylon leader line, 57.8-kg test nylon fishing line, and 2-mm ultrahigh–molecular weight polyethylene (UHMWPE) were used.

Procedures—In this in vitro study, samples of nylon leader line, fishing line, and UHMWPE sterilized by use of HPGP, ethylene oxide, and steam or unsterilized samples were used. Bacterial adherence on unsterilized and sterilized samples was tested with Staphylococcus epidermidis and Escherichia coli. Five samples were examined for each line type and sterilization condition, and final colony counts were obtained.

Results—Bacterial adherence was significantly affected by method of sterilization for all 3 line types. For most of the samples, bacterial adherence was similar or lower when HPGP sterilization was used, compared with results for sterilization via ethylene oxide and steam, respectively. Bacterial adherence was significantly higher for UHMWPE, compared with adherence for the nylon line, regardless of the sterilization method used. Bacterial adherence was higher for nylon fishing line than for nylon leader line for S epidermidis after ethylene oxide sterilization and for E coli after HPGP and ethylene oxide sterilization.

Conclusions and Clinical Relevance—Effects of HPGP sterilization on bioadhesive properties of nylon and polyethylene lines compared favorably with those for ethylene oxide and steam sterilization. Also, nylon line may be a more suitable material than UHMWPE for suture prostheses on the basis of bacterial adherence properties.