In vitro evaluation of a low-modulus mesh canine prosthetic hip stem

Denis J. Marcellin-Little Department of Clinical Sciences, College of Veterinary Medicine, College of Engineering, North Carolina State University, Raleigh, NC 27606.

Search for other papers by Denis J. Marcellin-Little in
Current site
Google Scholar
PubMed
Close
 DEDV
,
Omer Cansizoglu Edward P. Fitts Department of Industrial and Systems Engineering, College of Engineering, North Carolina State University, Raleigh, NC 27606.

Search for other papers by Omer Cansizoglu in
Current site
Google Scholar
PubMed
Close
 PhD
,
Ola L. A. Harrysson Edward P. Fitts Department of Industrial and Systems Engineering, College of Engineering, North Carolina State University, Raleigh, NC 27606.

Search for other papers by Ola L. A. Harrysson in
Current site
Google Scholar
PubMed
Close
 PhD
, and
Simon C. Roe Department of Clinical Sciences, College of Veterinary Medicine, College of Engineering, North Carolina State University, Raleigh, NC 27606.

Search for other papers by Simon C. Roe in
Current site
Google Scholar
PubMed
Close
 BVSc, PhD

Abstract

Objective—To compare an electron beam melting-processed (EBMP) low-modulus titanium alloy mesh stem with a commercial cobalt-chromium (CC) stem in a canine cadaver model.

Sample Population—9 pairs of cadaver femora.

Procedures—EBMP stems of 3 sizes were placed in randomly chosen sides of femora (left or right) and CC stems in opposite sides. Stem impaction distances were recorded. Five strain gauges were attached to the femoral surface to record transverse tensile (hoop) strains in the femur during axial loading. Constructs were axially loaded 4 times to 800 N and 4 times to 1,600 N in a materials testing machine. Axial stiffness of constructs and bone surface strains were compared between EBMP and CC constructs.

Results—Stems were impacted without creating femoral fissures or fractures. Stem impaction distances were larger for EBMP stems than for CC stems. Mean axial stiffness of EBMP constructs was lower than mean axial stiffness of CC constructs. Subsidence did not differ between groups. Bone strains varied among strain gauge positions and were largest at the distal aspect of the stems. At a load of 1,600 N, bones strains were higher in CC constructs than in EBMP constructs for 2 of 4 medial strain gauges.

Conclusions and Clinical Relevance—EBMP stems were successfully impacted and stable and led to a focal decrease in bone strain; this may represent an acceptable option for conventional or custom joint replacement. (Am J Vet Res 2010;71:1089–1095)

Although cemented hip stems have been used successfully as part of total hip replacements in humans, their success rate has been reportedly lower in younger patients than in older patients.1 The longterm success of hip stems is affected by aseptic implant loosening, implant wear, and stress-mediated bone resorption (stress shielding).2 Cementless hip stems were originally developed in part because polymethylmethacrylate bone cement was considered to be a contributing factor to aseptic loosening of cemented hip stems.3 A portion of a cementless stem is textured or coated with porous surfaces for bone ongrowth and ingrowth.4,5 Stem stability relies on initial press fit and long-term bone ingrowth into the porous portions of the stems. Cementless stems are large and have a high

Abstract

Objective—To compare an electron beam melting-processed (EBMP) low-modulus titanium alloy mesh stem with a commercial cobalt-chromium (CC) stem in a canine cadaver model.

Sample Population—9 pairs of cadaver femora.

Procedures—EBMP stems of 3 sizes were placed in randomly chosen sides of femora (left or right) and CC stems in opposite sides. Stem impaction distances were recorded. Five strain gauges were attached to the femoral surface to record transverse tensile (hoop) strains in the femur during axial loading. Constructs were axially loaded 4 times to 800 N and 4 times to 1,600 N in a materials testing machine. Axial stiffness of constructs and bone surface strains were compared between EBMP and CC constructs.

Results—Stems were impacted without creating femoral fissures or fractures. Stem impaction distances were larger for EBMP stems than for CC stems. Mean axial stiffness of EBMP constructs was lower than mean axial stiffness of CC constructs. Subsidence did not differ between groups. Bone strains varied among strain gauge positions and were largest at the distal aspect of the stems. At a load of 1,600 N, bones strains were higher in CC constructs than in EBMP constructs for 2 of 4 medial strain gauges.

Conclusions and Clinical Relevance—EBMP stems were successfully impacted and stable and led to a focal decrease in bone strain; this may represent an acceptable option for conventional or custom joint replacement. (Am J Vet Res 2010;71:1089–1095)

Although cemented hip stems have been used successfully as part of total hip replacements in humans, their success rate has been reportedly lower in younger patients than in older patients.1 The longterm success of hip stems is affected by aseptic implant loosening, implant wear, and stress-mediated bone resorption (stress shielding).2 Cementless hip stems were originally developed in part because polymethylmethacrylate bone cement was considered to be a contributing factor to aseptic loosening of cemented hip stems.3 A portion of a cementless stem is textured or coated with porous surfaces for bone ongrowth and ingrowth.4,5 Stem stability relies on initial press fit and long-term bone ingrowth into the porous portions of the stems. Cementless stems are large and have a high