Evaluation of a portable clinical analyzer for the determination of blood gas partial pressures, electrolyte concentrations, and hematocrit in venous blood samples collected from cattle, horses, and sheep

Juliana R. Peiró Department of Clinics, Surgery, and Animal Reproduction, School of Veterinary Medicine, Universidade Estadual Paulista (UNESP), Araçatuba, SP 16050-680, Brazil.

Search for other papers by Juliana R. Peiró in
Current site
Google Scholar
PubMed
Close
 DVM, PhD
,
Alexandre S. Borges Department of Veterinary Clinics, School of Veterinary Medicine and Animal Science, Universidade Estadual Paulista (UNESP), Botucatu, SP 18600-010, Brazil.

Search for other papers by Alexandre S. Borges in
Current site
Google Scholar
PubMed
Close
 DVM, PhD
,
Roberto C. Gonçalves Department of Veterinary Clinics, School of Veterinary Medicine and Animal Science, Universidade Estadual Paulista (UNESP), Botucatu, SP 18600-010, Brazil.

Search for other papers by Roberto C. Gonçalves in
Current site
Google Scholar
PubMed
Close
 DVM, PhD
, and
Luiz Claudio N. Mendes Department of Clinics, Surgery, and Animal Reproduction, School of Veterinary Medicine, Universidade Estadual Paulista (UNESP), Araçatuba, SP 16050-680, Brazil.

Search for other papers by Luiz Claudio N. Mendes in
Current site
Google Scholar
PubMed
Close
 DVM, PhD

Abstract

Objective—To compare results reported for blood gas partial pressures, electrolyte concentrations, and Hct in venous blood samples collected from cattle, horses, and sheep and analyzed by use of a portable clinical analyzer (PCA) and reference analyzer (RA).

Animals—Clinically normal animals (24 cattle, 22 horses, and 22 sheep).

Procedures—pH; Pco2; Po2; total carbon dioxide concentration; oxygen saturation; base excess; concentrations of HCO3, Na+, K+, and ionized calcium; Hct; and hemoglobin concentration were determined with a PCA. Results were compared with those obtained for the same blood sample with an RA. Bias (mean difference) and variability (95% confidence interval) were determined for all data reported. Data were also subjected to analyses by Deming regression and Pearson correlation.

Results—Analysis of Bland-Altman plots revealed good agreement between results obtained with the PCA and those obtained with the RA for pH and total carbon dioxide concentration in cattle, K+ concentration in horses and sheep, and base excess in horses. Except for Na+ concentration and Hct in horses and sheep, correlation was good or excellent for most variables reported.

Conclusions and Clinical Relevance—Data from blood gas and electrolyte analyses obtained by use of the PCA can be used to evaluate the health status of cattle, horses, and sheep. Furthermore, the handheld PCA device may have a great advantage over the RA device as a result of the ability to analyze blood samples on farms that may be located far from urban centers.

Abstract

Objective—To compare results reported for blood gas partial pressures, electrolyte concentrations, and Hct in venous blood samples collected from cattle, horses, and sheep and analyzed by use of a portable clinical analyzer (PCA) and reference analyzer (RA).

Animals—Clinically normal animals (24 cattle, 22 horses, and 22 sheep).

Procedures—pH; Pco2; Po2; total carbon dioxide concentration; oxygen saturation; base excess; concentrations of HCO3, Na+, K+, and ionized calcium; Hct; and hemoglobin concentration were determined with a PCA. Results were compared with those obtained for the same blood sample with an RA. Bias (mean difference) and variability (95% confidence interval) were determined for all data reported. Data were also subjected to analyses by Deming regression and Pearson correlation.

Results—Analysis of Bland-Altman plots revealed good agreement between results obtained with the PCA and those obtained with the RA for pH and total carbon dioxide concentration in cattle, K+ concentration in horses and sheep, and base excess in horses. Except for Na+ concentration and Hct in horses and sheep, correlation was good or excellent for most variables reported.

Conclusions and Clinical Relevance—Data from blood gas and electrolyte analyses obtained by use of the PCA can be used to evaluate the health status of cattle, horses, and sheep. Furthermore, the handheld PCA device may have a great advantage over the RA device as a result of the ability to analyze blood samples on farms that may be located far from urban centers.

Contributor Notes

Presented in part as a poster presentation at the 21st Annual American College of Veterinary Internal Medicine Forum, Charlotte, NC, June 2003.

The authors thank Lucilena Barbella Stelzer for technical assistance.

Address correspondence to Dr. Peiró (jpeiro@fmva.unesp.br).
  • 1.

    Erickson KA, Wilding P. Evaluation of a novel point-of-care system, the i-STAT portable clinical analyzer. Clin Chem 1993;39:283287.

  • 2.

    Jacobs E, Vadasdi E, Sarkozi L, et alAnalytical evaluation of i-STAT portable clinical analyzer and use by nonlaboratory health-care professionals. Clin Chem 1993;39:10691074.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 3.

    Sibbald WJ, Eberhard JA, Inman KJ, et alNew technologies, critical care, and economic realities. Crit Care Med 1993;21:17771780.

  • 4.

    Woo J, Mccabe JB, Chauncey D, et alThe evaluation of a portable clinical analyzer in the emergency department. Am J Clin Pathol 1993;100:599605.

  • 5.

    Wong RJ, Mahoney JJ, Harvey JA, et alStatPal II pH and blood gas analysis system evaluated. Clin Chem 1994;40:124129.

  • 6.

    Tortella BJ, Lavery RF, Doran JV, et alPrecision, accuracy, and managed care implications of a hand-held whole blood analyzer in the prehospital setting. Clin Chem 1996;106:124127.

    • Search Google Scholar
    • Export Citation
  • 7.

    Looney AL, Ludders J, Erb HN, et alUse of a handheld device for analysis of blood electrolyte concentrations and blood gas partial pressures in dogs and horses. J Am Vet Med Assoc 1998;213:526530.

    • Search Google Scholar
    • Export Citation
  • 8.

    Tschudi PR. Evaluation of the portable blood analyzer i-STAT. Schweiz Arch Tierheilkd 1998;140:507512.

  • 9.

    Larsen RS, Haulena M, Grindem CB, et alBlood values of juvenile northern elephant seals (Mirounga angustirostris) obtained using a portable clinical analyzer. Vet Clin Pathol 2002;31:106110.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 10.

    Silverman SC, Birks EK. Evaluation of i-STAT hand-held chemical analyser during treadmill and endurance exercise. Equine Vet J Suppl 2002;34:551554.

    • Search Google Scholar
    • Export Citation
  • 11.

    Verwaerde P, Malet C, Lagente M, et alThe accuracy of the i-STAT portable analyser for measuring blood gases and pH in whole-blood samples from dogs. Res Vet Sci 2002;73:7175.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 12.

    Tinkey P, Lembo T, Craig S, et alUse of i-STAT portable clinical analyzer in mice. Lab Anim (NY) 2006;35:4550.

  • 13.

    Tappin S, Rizzo F, Dodkin S, et alMeasurement of ionized calcium in canine blood samples collected in prefilled and self-filled heparinized syringes using the i-STAT point-of-care analyzer. Vet Clin Pathol 2008;37:6672.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 14.

    Steinmetz HW, Vogt R, Kästner S, et alEvaluation of the i-STAT portable clinical analyzer in chickens (Gallus gallus). J Vet Diagn Invest 2007;19:382388.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 15.

    Sediame S, Zerah-Lancner F, d'Ortho MP, et alAccuracy of the i-STAT bedside blood gas analyzer. Eur Respir J 1999;14:214217.

  • 16.

    Grosenbaugh DA, Gadawski KE, Muir WW. Evaluation of a portable clinical analyzer in a veterinary hospital setting. J Am Vet Med Assoc 1998;213:691694.

    • Search Google Scholar
    • Export Citation
  • 17.

    Cohn LA, McCaw DL, Tate DJ, et alAssessment of five portable blood glucose meters, a point-of-care analyzer, and color test strips for measuring blood glucose concentration in dogs. J Am Vet Med Assoc 2000;216:198202.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 18.

    Wenker CJ, Hunziker D, Lopez J, et alHaematology, blood chemistry and urine parameters of free-ranging plains viscachas (Lagostomus maximus) in Argentina determined by use of a portable blood analyser (i-STAT) and conventional laboratory methods. J Vet Med A Physiol Pathol Clin Med 2007;54:260264.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 19.

    Saulez MN, Cebra CK, Dailey M. Comparative biochemical analyses of venous blood and peritoneal fluid from horses with colic using a portable analyzer and an in-house analyser. Vet Rec 2005;157:217223.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 20.

    i-STAT system manual. Abbott Park, Ill: i-STAT, 2008;6.

  • 21.

    AVL Omni operator's manual. Mannheim, Germany: Roche Diagnostics GmbH, 2003;396.

  • 22.

    Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307310.

  • 23.

    Gray TE, Pratt MC, Cusick PK. Determination of agreement between laboratory instruments. Contemp Top Lab Anim Sci 1999;38:5659.

  • 24.

    Jensen AL, Bantz M. Comparing laboratory tests using the difference plot method. Vet Clin Pathol 1993;22:4648.

  • 25.

    Papasouliotis K, Cue S, Crawford E, et alComparison of white blood cell differential percentages determined by the in-house LaserCyte hematology analyzer and a manual method. Vet Clin Pathol 2006;35:295302.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 26.

    Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat 2007;17:571582.

  • 27.

    Radostitis OM, Gay CC, Hinchcliff KW, et alAppendix 2: reference laboratory values. In: Radostitis OM, Gay CC, Hinchcliff KW, et al, eds. Veterinary medicine. A textbook of the diseases of cattle, horses, sheep, pigs, and goats. 10th ed. Philadelphia: Saunders Elsevier, 2007;20472050.

    • Search Google Scholar
    • Export Citation
  • 28.

    Daurès MF, Combescure C, Cristol JP. Étude comparative de six analyseurs de gaz du sang. Ann Biol Clin (Paris) 2007;65:505518.

  • 29.

    Toffaletti J. Use of novel preparations of heparin to eliminate interference of ionized calcium measurements: have all the problems been solved? Clin Chem 1994;40:508509.

    • Crossref
    • Search Google Scholar
    • Export Citation
  • 30.

    Carlson GP. Clinical chemistry tests: table 22–2. In: Smith BP, ed. Large animal internal medicine. 3rd ed. St Louis: Mosby, 2002;393.

    • Search Google Scholar
    • Export Citation
  • 31.

    Kramer JW. Normal hematology of cattle, sheep, and goats. In: Feldman BF, Zinkl JG, Jain NC, eds. Schalm's veterinary hematology. 5th ed. Philadelphia: Lippincott Williams & Wilkins, 2000;10751084.

    • Search Google Scholar
    • Export Citation
  • 32.

    Mock T, Morrison D, Yatscoff R. Evaluation of the i-STAT system: a portable chemistry analyzer for the measurement of sodium, potassium, chloride, urea, glucose, and hematocrit. Clin Biochem 1995;28:187192.

    • Crossref
    • Search Google Scholar
    • Export Citation

Advertisement