Abstract
Objective—To compare cardiac troponin I (cTnI) concentrations determined by use of a point-of-care analyzer with values determined by use of a bench-top immunoassay in plasma samples obtained from clinically normal horses with and without experimentally induced cardiac disease, and to establish a reference range for plasma equine cTnI concentration determined by use of the point-of-care analyzer.
Animals—83 clinically normal horses, 6 of which were administered monensin to induce cardiac disease.
Procedures—A blood sample was collected from each of the 83 clinically normal horses to provide plasma for analysis by use of the point-of-care analyzer; some of the same samples were also analyzed by use of the immunoassay. All 83 samples were used to establish an analyzer-specific reference range for plasma cTnI concentration in clinically normal horses. In 6 horses, blood samples were also collected at various time points after administration of a single dose of monensin (1.0 to 1.5 mg/kg) via nasogastric intubation; plasma cTnI concentration in those samples was assessed by use of both methods.
Results—The analyzer-specific reference range for plasma cTnI concentration in clinically normal horses was 0.0 to 0.06 ng/mL. Following monensin treatment in 5 horses, increases in plasma cTnI concentration determined by use of the 2 methods were highly correlated (Pearson correlation, 0.83). Peak analyzer-determined plasma cTnI concentrations in monensin-treated horses ranged from 0.08 to 3.68 ng/mL.
Conclusions and Clinical Relevance—In horses with and without experimentally induced cardiac disease, the point-of-care analyzer and bench-top immunoassay provided similar values of plasma cTnI concentration.