Use of bispectral index to monitor depth of anesthesia in isoflurane-anesthetized dogs

Daniela Campagnol Department of Veterinary Surgery and Anesthesiology, Faculdade de Medicina Veterinária e Zootecnia, Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brazil.

Search for other papers by Daniela Campagnol in
Current site
Google Scholar
PubMed
Close
 MV, MSc
,
Francisco J. Teixeira Neto Department of Veterinary Surgery and Anesthesiology, Faculdade de Medicina Veterinária e Zootecnia, Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brazil.

Search for other papers by Francisco J. Teixeira Neto in
Current site
Google Scholar
PubMed
Close
 MV, PhD
,
Eduardo R. Monteiro Department of Veterinary Surgery and Anesthesiology, Faculdade de Medicina Veterinária e Zootecnia, Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brazil.

Search for other papers by Eduardo R. Monteiro in
Current site
Google Scholar
PubMed
Close
 MV, MSc
,
Suzane L. Beier Department of Veterinary Surgery and Anesthesiology, Faculdade de Medicina Veterinária e Zootecnia, Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brazil.

Search for other papers by Suzane L. Beier in
Current site
Google Scholar
PubMed
Close
 MV, PhD
, and
Antônio J. A. Aguiar Department of Veterinary Surgery and Anesthesiology, Faculdade de Medicina Veterinária e Zootecnia, Universidade Estadual Paulista (UNESP), Botucatu, São Paulo, Brazil.

Search for other papers by Antônio J. A. Aguiar in
Current site
Google Scholar
PubMed
Close
 MV, PhD

Abstract

Objective—To evaluate the correlation between the bispectral index (BIS) and end-tidal isoflurane (ETISO) concentration and compare the use of 3 BIS sensor positions in dogs.

Animals—6 adult dogs.

Procedures—Mechanically ventilated dogs received pancuronium, and depth of anesthesia was altered by increasing ETISO concentration from 1.5% to 2.3% and 3.0%. The BIS, suppression ratio (relative percentage of isoelectric electroencephalographic waveforms), and signal quality index (SQI) were recorded at each ETISO concentration for each of 3 BIS sensor positions (frontal-occipital, bifrontal, and frontal-temporal positions).

Results—The BIS and ETISO concentration were poorly correlated; regardless of sensor positioning, mean BIS values did not change significantly as ETISO was increased. At 3% isoflurane, regardless of sensor positioning, there was an increase in suppression ratio coincident with BIS < 40 in some dogs, whereas paradoxic increases in BIS (> 60) were recorded in others. Furthermore, at 3.0% isoflurane, the SQI was significantly lower for the bifrontal sensor position (compared with values for the other positions), but low SQI values prevented recording of BIS values from the frontal-occipital sensor position in 2 dogs. Overall, BIS values derived from the 3 sensor positions did not differ.

Conclusions and Clinical Relevance—In dogs, BIS values may not reflect changes in depth of isoflurane anesthesia in the absence of noxious stimulation. Of the 3 sensor positions, frontal-temporal positioning provided better correlation with changes in depth of anesthesia induced via changes in isoflurane concentrations. However, the sensor placements yielded similar results at SQI values > 50.

Abstract

Objective—To evaluate the correlation between the bispectral index (BIS) and end-tidal isoflurane (ETISO) concentration and compare the use of 3 BIS sensor positions in dogs.

Animals—6 adult dogs.

Procedures—Mechanically ventilated dogs received pancuronium, and depth of anesthesia was altered by increasing ETISO concentration from 1.5% to 2.3% and 3.0%. The BIS, suppression ratio (relative percentage of isoelectric electroencephalographic waveforms), and signal quality index (SQI) were recorded at each ETISO concentration for each of 3 BIS sensor positions (frontal-occipital, bifrontal, and frontal-temporal positions).

Results—The BIS and ETISO concentration were poorly correlated; regardless of sensor positioning, mean BIS values did not change significantly as ETISO was increased. At 3% isoflurane, regardless of sensor positioning, there was an increase in suppression ratio coincident with BIS < 40 in some dogs, whereas paradoxic increases in BIS (> 60) were recorded in others. Furthermore, at 3.0% isoflurane, the SQI was significantly lower for the bifrontal sensor position (compared with values for the other positions), but low SQI values prevented recording of BIS values from the frontal-occipital sensor position in 2 dogs. Overall, BIS values derived from the 3 sensor positions did not differ.

Conclusions and Clinical Relevance—In dogs, BIS values may not reflect changes in depth of isoflurane anesthesia in the absence of noxious stimulation. Of the 3 sensor positions, frontal-temporal positioning provided better correlation with changes in depth of anesthesia induced via changes in isoflurane concentrations. However, the sensor placements yielded similar results at SQI values > 50.